24 October 2017 Analysis on the optical aberration effect on spectral resolution of coded aperture spectroscopy
Peng Hao, Mingbo Chi, Yihui Wu
Author Affiliations +
Proceedings Volume 10461, AOPC 2017: Optical Spectroscopy and Imaging; 104610B (2017) https://doi.org/10.1117/12.2282038
Event: Applied Optics and Photonics China (AOPC2017), 2017, Beijing, China
Abstract
The coded aperture spectrometer can achieve high throughput and high spectral resolution by replacing the traditional single slit with two-dimensional array slits manufactured by MEMS technology. However, the sampling accuracy of coding spectrum image will be distorted due to the existence of system aberrations, machining error, fixing errors and so on, resulting in the declined spectral resolution. The influence factor of the spectral resolution come from the decode error, the spectral resolution of each column, and the column spectrum offset correction. For the Czerny-Turner spectrometer, the spectral resolution of each column most depend on the astigmatism, in this coded aperture spectroscopy, the uncorrected astigmatism does result in degraded performance. Some methods must be used to reduce or remove the limiting astigmatism. The curvature of field and the spectral curvature can be result in the spectrum revision errors.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Peng Hao, Mingbo Chi, and Yihui Wu "Analysis on the optical aberration effect on spectral resolution of coded aperture spectroscopy", Proc. SPIE 10461, AOPC 2017: Optical Spectroscopy and Imaging, 104610B (24 October 2017); https://doi.org/10.1117/12.2282038
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical aberrations

Spectroscopy

Coded apertures

Optical spectroscopy

Spectral resolution

Back to Top