Paper
7 December 2017 Cavity ring down spectrometry for disease diagnostics using exhaled air
Author Affiliations +
Proceedings Volume 10592, Biophotonics—Riga 2017; 1059208 (2017) https://doi.org/10.1117/12.2297126
Event: The Second International Conference "Biophotonics-Riga 2017", 2017, Riga, Latvia
Abstract
In this paper we report the current stage of the development of a cavity ring-down spectrometer (CRDS) system using exhaled human breath analysis for the diagnostics of different diseases like diabetes and later lung cancer. The portable CRDS system is made in ultraviolet spectral region using Nd:Yag laser 266 nm pulsed light. Calibration of the CRDS system was performed using generated samples by KinTek automated permeation tube system and self-prepared mixtures with known concentration of benzene and acetone in air. First experiments showed that the limits of detection for benzene and acetone are several tens of ppb.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
G. Revalde, K. Grundšteins, J. Alnis, and A. Skudra "Cavity ring down spectrometry for disease diagnostics using exhaled air", Proc. SPIE 10592, Biophotonics—Riga 2017, 1059208 (7 December 2017); https://doi.org/10.1117/12.2297126