17 January 1990 Compact Pulsed X-Ray Generator Operated At High Repetitive Rates
Hiroshi Isobe, Eiichi Sato, Shingo Kimura, Yoshiharu Tamakawa, Toru Yanagisawa
Author Affiliations +
Abstract
A repetitive pulsed x-ray generator achieved with a compact diode utilizing a new type of cathode for biomedical radiography is described. This generator consisted of the following components: a high-voltage power supply, two ceramic condensers of about 850pF, a repetitive impulse switching system, a turbo molecular pump, and an x-ray tube. Since the high-voltage pulser employed a modified Marx circuit, this pulser produced twice the potential of the condenser charging voltage. The x-ray tube was of the demountable-diode type which was connected to the turbo molecular pump and consisted of the following components: a rod-shaped anode tip made of tungsten, a plane cathode made of aluminum and carbon, and a vacuum vessel made of glass with a diameter of 50mm. Two condensers were charged from 30 to 100kV, and the output of this pulser ranged from 50 to 180kV. The x-ray pulse widths primarily increased according to increases in the anode-cathode (A-C space) and their values ranged from 20 to 100ns. The repetitive rate was determined by the condenser capacity, the charging voltage, and the current capacity of the power supply, and its maximum value was about 100Hz. The time integrated x-ray intensities were less than 4.0pC/kg at 0.5m per pulse when the discharge capacity of about 430pF (Marx Circuit) was employed. The effective focal spot size was determined by the diameter of anode tip and its value ranged from 0.5 to 3.0mm in diameter.
© (1990) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hiroshi Isobe, Eiichi Sato, Shingo Kimura, Yoshiharu Tamakawa, and Toru Yanagisawa "Compact Pulsed X-Ray Generator Operated At High Repetitive Rates", Proc. SPIE 1155, Ultrahigh Speed and High Speed Photography, Photonics, and Videography '89: Seventh in a Series, (17 January 1990); https://doi.org/10.1117/12.962425
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

High speed photography

Photography

Power supplies

Aluminum

Carbon

Glasses

RELATED CONTENT

Biomedical Flash Radiography
Proceedings of SPIE (February 04 1988)
Generation of flash x rays using a mercury anode radiation...
Proceedings of SPIE (February 02 1993)
Repetitive Pulsed X-Ray Generator Utilizing A Triode
Proceedings of SPIE (January 17 1990)
An Oil-Cooled Flash X-Ray Tube For Biomedical Radiography
Proceedings of SPIE (February 01 1989)

Back to Top