Presentation + Paper
7 June 2024 Automated SWIR, MWIR, and LWIR camera test results compared to models
Author Affiliations +
Abstract
System level modeling and testing of SWIR, MWIR and LWIR infrared cameras are presented. Automated camera testing is performed by Santa Barbara Infrared's IRWindows software. Some of the standard tests are NETD, MTF, 3D noise, and Ensquared Energy. An Excel spreadsheet has been developed to model camera system performance which allows relatively easy addition of new camera types. Each camera type includes information on the ROIC (number and format of pixels, pixel readout rate, pixel size, well size, read noise), sensor material (spectral QE, fill factor, dark current vs temperature), lens, window and warm filter spectral transmission, cold filter temperature and spectral transmission, and cold aperture configuration (diameter and distance from sensor). Atmospheric temperature, turbulence, spectral absorption and spectral emission are included in the Excel model.
Conference Presentation
(2024) Published by SPIE. Downloading of the abstract is permitted for personal use only.
Arnold Adams "Automated SWIR, MWIR, and LWIR camera test results compared to models", Proc. SPIE 13045, Infrared Imaging Systems: Design, Analysis, Modeling, and Testing XXXV, 130450S (7 June 2024); https://doi.org/10.1117/12.3013875
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Cameras

Staring arrays

Modulation transfer functions

Optical filters

Photons

Sensors

Tunable filters

RELATED CONTENT


Back to Top