Paper
1 May 1991 Comparative study of excimer and erbium:YAG lasers for ablation of structural components of the knee
Sandor G. Vari M.D., Wei-Qiang Shi, Maurits J. van der Veen, Michael C. Fishbein M.D., J. M. Miller, Thanassis Papaioannou, Warren S. Grundfest M.D.
Author Affiliations +
Proceedings Volume 1424, Lasers in Orthopedic, Dental, and Veterinary Medicine; (1991) https://doi.org/10.1117/12.43986
Event: Optics, Electro-Optics, and Laser Applications in Science and Engineering, 1991, Los Angeles, CA, United States
Abstract
This study was designed to compare the efficiency and thermal effect of a 135 ns pulsed-stretched XeCl excimer laser (308 nm) and a free-running Erbium:YAG laser (2940 nm) with 200 microsecond(s) pulse duration for ablation of knee joint structures (hyaline and fibrous cartilage, tendon and bone). The radiant exposure used for tissue ablation ranged from 2 to 15 J/cm2 for the XeCl excimer and from 33 to 120 J/cm2 for Er:YAG. The excimer and Er:YAG lasers were operated at 4 and 5 Hz respectively. The ablative laser energy was delivered to tissue through fibers. Ablation rates of soft tissues (hyaline and fibrous cartilage, tendon) varied from 8.5 to 203 micrometers /pulse for excimer and from 8.2 to 273 micrometers /pulse for Er:YAG lasers. Ablation rates of soft tissues are linearly dependent on the radiant exposure. Within the range of parameters tested all the tissues except the bone could be rapidly ablated by both lasers. Bone ablation was much less efficient, requiring 15 J/cm2 and 110 J/cm2 radiant exposure for excimer and Er:YAG lasers to ablate 9.5 and 8.2 micrometers tissue per pulse. However, excimer laser ablation produced less thermal damage in the tissues studied compared to Er:YAG at the same laser parameters. The authors conclude that both lasers are capable of efficient knee joint tissue ablation. XeCl excimer laser requires an order of magnitude less energy than Er:YAG laser for comparable tissue ablation.
© (1991) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Sandor G. Vari M.D., Wei-Qiang Shi, Maurits J. van der Veen, Michael C. Fishbein M.D., J. M. Miller, Thanassis Papaioannou, and Warren S. Grundfest M.D. "Comparative study of excimer and erbium:YAG lasers for ablation of structural components of the knee", Proc. SPIE 1424, Lasers in Orthopedic, Dental, and Veterinary Medicine, (1 May 1991); https://doi.org/10.1117/12.43986
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Laser ablation

Excimer lasers

Cartilage

Er:YAG lasers

Tissues

Laser tissue interaction

Excimers

RELATED CONTENT

Lasers in oral surgery and implantology
Proceedings of SPIE (March 22 2016)
Excimer laser in arthroscopic surgery
Proceedings of SPIE (May 01 1991)
Effects of Er:YAG laser irradiation on human cartilage
Proceedings of SPIE (March 27 1996)
Effects of pulsed lasers on hard biological tissue
Proceedings of SPIE (July 07 1993)

Back to Top