1 August 1991 Technique for improving the calibration of large-area sphere sources
James Walker, Christopher L. Cromer, James T. McLean
Author Affiliations +
Abstract
A new technique for improving the accuracy of radiance calibrations for large-area integrating-sphere sources has been investigated. Such sources are used to calibrate numerous aircraft and spacecraft remote sensing instruments. Recent measurements performed at NIST and NASA Goddard Space Flight Center have demonstrated that the uncertainty of sphere-source radiance measurements can be improved from the present 5 to 10 percent level to a 1 to 2 percent level. Silicon detectors with bandpass filters mounted in front of them and calibrated for absolute spectral responsivity can be used to confirm and to monitor the absolute radiance of a sphere source.
© (1991) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
James Walker, Christopher L. Cromer, and James T. McLean "Technique for improving the calibration of large-area sphere sources", Proc. SPIE 1493, Calibration of Passive Remote Observing Optical and Microwave Instrumentation, (1 August 1991); https://doi.org/10.1117/12.46707