Paper
4 November 2004 Fabrication of collimators for gamma-ray imaging
Olga V. Makarova, Guohua Yang, Cha-Mei Tang, Derrick C. Mancini, Ralu Divan, Judith Yaeger
Author Affiliations +
Abstract
Collimators capable of higher resolution and optimized for greater sensitivity can significantly improve the imaging quality of gamma-cameras for single-photon-emission computed tomography of small animals. We are applying deep x-ray lithography and gold electroforming techniques to fabricate high-resolution collimators with continuous, smooth, and thin septa. Negative SU-8 photoresist was used for mold fabrication. To be efficient, collimators for gamma-cameras designed to image 140 keV gamma-rays should be over 1.5 cm tall. The height of the collimator can be achieved by stacking the appropriate number of layers.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Olga V. Makarova, Guohua Yang, Cha-Mei Tang, Derrick C. Mancini, Ralu Divan, and Judith Yaeger "Fabrication of collimators for gamma-ray imaging", Proc. SPIE 5539, Design and Microfabrication of Novel X-Ray Optics II, (4 November 2004); https://doi.org/10.1117/12.563748
Lens.org Logo
CITATIONS
Cited by 9 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Collimators

Gold

X-rays

Imaging systems

Gamma radiation

Gamma ray imaging

Single photon emission computed tomography

RELATED CONTENT

Testing of a gamma ray imaging system at the High...
Proceedings of SPIE (September 23 2014)
Fabrication of the pinhole aperture for AdaptiSPECT
Proceedings of SPIE (September 12 2014)
Evolution of the neutron imaging aperture
Proceedings of SPIE (September 18 2018)
Issues in the repair of x-ray masks
Proceedings of SPIE (March 01 1991)
Imager onboard INTEGRAL
Proceedings of SPIE (October 18 1996)

Back to Top