Paper
10 February 2006 Relative spectral responsivity determination of photometric detectors
J. C. Molina, J. C. Bermudez, J. E. Hernández-López
Author Affiliations +
Proceedings Volume 6046, Fifth Symposium Optics in Industry; 604620 (2006) https://doi.org/10.1117/12.674612
Event: Fifth Symposium Optics in Industry, 2005, Santiago De Queretaro, Mexico
Abstract
The experimental evaluation of the relative spectral response of a standard photometer is presented; the measurements were carried out by the substitution method by means a double monochromator in the range of 405 nm to 730 nm with 5 nm resolution wavelength. The experimental uncertainty and its mismatching factor f1' are evaluated as part of the routine procedure to realize the candela unit with trazability to an absolute cryogenic radiometer which is the national standard of radiant flux at CENAM (National Center of Metrology of Mexico).
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. C. Molina, J. C. Bermudez, and J. E. Hernández-López "Relative spectral responsivity determination of photometric detectors", Proc. SPIE 6046, Fifth Symposium Optics in Industry, 604620 (10 February 2006); https://doi.org/10.1117/12.674612