1.IntroductionBiological cells convert forces to intracellular biochemical signaling cascades by mechanotransduction, a process that is responsible for diverse physiological phenomena including bone and vascular wall remodeling, vascular caliber control, and embryonic development.1 In contrast, atherosclerosis, sensory dysfunction in diabetes mellitus, and hair cell damage in the inner ear leading to vertigo are examples of cellular mechanical regulatory processes gone awry. 2 Physiological forces arise from blood-flow-induced shear stresses, hydrostatic pressure, extracellular matrix deformations, intracellular contractions, and osmotic swelling. The length scale of these forces spans single molecules, cellular organelles, cells, tissues, organs, and physiological systems; the time scales span nanoseconds to days.3 The essential ingredients of a comprehensive understanding of mechanotransduction are a well-defined force, cellular mechanical properties, and a physiological readout directly related to the force. Well-defined forces arise from fluid flow,4 atomic force microscopes,5, 6 optical traps,7 and magnetic beads.8 Cellular deformations of these forces can be modeled using continuum mechanics8, 9, 10 or ultrastructural characterization of load-bearing structures.11, 12, 13 Examples of readouts of force-induced physiological responses include calcium signaling,14 ion channel activity,15 phosphorylation of proteins,16 and transcription17 of new RNA. A major challenge in mechanotransduction is to differentiate mechanosensors (cellular structures that are perturbed by physical forces), mechanotransducers (molecules that undergo biochemical changes in response to force and initiate signaling cascades), and mechanobiology (the resultant changes in cellular structure and function). Analysis of single molecules in cells subjected to force can provide the link between force and mechanotransduction. Major technical and computational advances in analysis of fluorescence fluctuations provide new methods to assess single-molecule dynamics in model and cellular systems.18, 19 For example, time-correlated single photon counting (TCSPC) instrumentation is available on a single compact PCI, (peripheral component interconnect) card that, when integrated with pulsed and continuous wave lasers, can be used to assess nanosecond- to second-scale dynamics of single fluorescent molecules. Analysis methods including fluorescence correlation spectroscopy (FCS) and fluorescence lifetime can then be used to gain information on diffusion, conformational changes, aggregation, chemical kinetics, and other important biomolecular phenomena.18, 19 In this paper, we describe the construction and calibration of a confocal molecular dynamics microscope (CMDM) capable of (1) applying cell-scale forces using micropipette aspiration, optical traps, and fluid flow; (2) assessing cellular mechanical properties using optical trap and particle-tracking microrheology; and (3) correlating molecular-scale physiological readouts from TCSPC with cellular structures imaged with multimodal microscopy. The intended application of this system is to assess force-induced changes in dynamics of molecules occurring on a time scale of nanoseconds, while addressing long-term adaptive responses of cells on the order of hours. It is anticipated that these new tools will enable a comprehensive analysis of cellular mechanobiology and lead to major clinical advances in treating or preventing diseases such as atherosclerosis, which have their origins in mechanotransduction.2 We begin with a background on the theory of FCS, optical traps, and multimodal microscopy. We then describe the detailed construction and implementation of the CMDM. Finally, we present results of calibration experiments in which molecular diffusion is analyzed in solutions and in model and cellular membranes. 2.Background2.1.FCSThe detection of single molecules in femtoliter volumes was made possible by the development of confocal optics, high-sensitivity detectors, and robust fluorophores.20 Probe molecules move into and out of a confocal volume yielding fluorescence fluctuations (Fig. 1 ). Autocorrelation analysis of these fluctuations can provide diffusion coefficients and other single-molecule information.21, 22 In the following analysis, we consider the relationship between autocorrelation of fluorescence fluctuations and molecular diffusion. We begin with the Stokes-Einstein relationship for the diffusion coefficient of a sphere with a hydrodynamics radius in a solvent of viscosity : where is the Boltzmann constant, and is absolute temperature. We define a 2-D characteristic molecular diffusion transit time across a small area of radius such thatUnder conditions of equilibrium, the fluorescence fluctuates around an average value , due to diffusion. The intensity of these fluctuations can be autocorrelated to obtain the autocorrelation function given bywhere (time) and (lag time) vary over all times of the data collection period. The fluorescence fluctuations are related to the instantaneous change in concentration in the observation volume. To relate the autocorrelation function to diffusion, we use the diffusion equation:and the relationship between fluorescence fluctuations and concentration of the fluorescent molecules in the confocally defined optical probe volume:where is the excitation intensity profile, is the collection efficiency profile, and is a proportionality constant. The confocal probe volume is created by placing a small aperture (pinhole or fiber optic) in an image plane that is conjugate to the focus of a high-numerical-aperture, infinity-corrected objective. The laser beam has a Gaussian intensity profile that fills the back aperture of the objective, leading to a Gaussian illumination profile such that , where and are the probe volume radius and half-height, respectively, and are defined as the point where the intensity falls off to of the maximum (center) intensity (Fig. 1). A single constant factor known as the structure factor is defined as the ratio of . The autocorrelation obtained from the experiment and its relation to the theoretical Gaussian confocal volume and the characteristic diffusion time is21, 22, 23 where is the average number of diffusing fluorophores in the confocal volume. The number of fluorescent molecules present in the sample volume is the inverse of the term .The relationship between the autocorrelation function and diffusion on 2-D structures (e.g., plasma membrane) is given by This measurement in two dimensions can be extended to include multicomponent diffusion of multiple noninteracting species of fluorescent molecules, according to the equationwhere is the relative proportion of the noninteracting diffusing molecules.In the case of 2-D anomalous diffusion (e.g., in a cellular membrane), the time dependence of the mean square displacements is not linear but of the form , and the autocorrelation is given by24 where is the transport coefficient, and is the temporal exponent with values between zero and one. For a more comprehensive description of FCS including analysis of other sources of fluorescence fluctuations and detailed derivations of the equations presented here, see an excellent book by Zander 232.2.Fluorescence LifetimeTime-resolved fluorescence lifetime spectroscopy enables analysis of subtle changes in photophysics of fluorescent molecules.25 When a fluorescent molecule is excited to a higher energy state using a picosecond pulse of laser light, it remains in the excited state for a finite time before it decays to the ground-level energy state. Using a high-frequency pulsed laser, histograms of photon emission times relative to excitation times can be generated and fit with a negative exponential (or multiple exponentials) with a characteristic decay time (or lifetime) (different from the characteristic diffusion time of FCS). Fluorescence lifetime depends on local molecular microenvironmental factors including ionic strength, hydration, oxygen concentration, binding to macromolecules, and the proximity to other molecules that can deplete the excited state by resonance energy transfer.25 The fluorescence lifetime and quantum yield are related to intrinsic photophysical characteristics of a fluorescent molecule such as radiative and nonradiative decay mechanisms. The fluorescence quantum yield is the ratio of the number of photons emitted to the number of photons absorbed, according to where and are the radiative and nonradiative decay rates of the molecule, respectively. Fluorescence lifetime is given byThe value for depends on the mode of the nonradiative decay, such as collisional quenching, hydration, and vibrational relaxation. Thus, any alteration of also leads to a detectable change in the value of the fluorescence lifetime.The value for fluorescence lifetime is obtained by an iterative reconvolution of an instrument response function (IRF) with the fluorescence intensity using an assumed decay law, which can be approximated by a sum of exponentials:25 where is the fraction of molecules with lifetime , normalized to unity. Fluorescence lifetimes are independent of fluorescence probe concentrations and can provide information not obtainable from intensity variations alone. When polarized light is used to excite a molecule whose excitation dipole is oriented parallel to the polarization of the pulse, it is possible to separate the parallel and perpendicular components of the emitted fluorescence signal and to extract rotational diffusion constants.252.3.Optical TrapOptical traps apply piconewton-level forces to small beads or organelles, using momentum transfer from focused laser light. Although the ability of light to apply forces to matter was discovered by 1873 by Maxwell, it is only in the last two decades that laser and optics developments permitted application of optical trapping forces to cells and molecules.26, 27 Briefly, the energy carried as momentum by a traveling photon is translated into a net force on an object hit by that photon. The integrated photonic force on a particle in a Gaussian laser beam is given by28, 29 where is a geometrical factor of order one; and are the refractive indices of the surrounding medium and the particle, respectively; is the volume of the bead; is the speed of light in a vacuum; is the intensity of the incident light; is the distance between the centers of the bead and laser beam; is a unit vector; and is the laser beam radius. In practice, when , the trap acts as a linear spring for small displacements with a spring constant determined by equating the variance of 1-D displacements with the thermal fluctuations according to the equipartition theorem:where is the mean value of , is the Boltzmann constant , is absolute temperature, and refers to the time average. Alternatively, the trap strength can be measured by quantifying the displacement of the bead due30 to applied force (e.g., from fluid flow) (i.e., .2.4.Multimodal MicroscopyTo correlate molecular-scale events with cellular structures and their locations, we have integrated TCSPC and an optical trap into a multimodal microscope consisting of differential interference contrast (DIC), total internal reflection fluorescence (TIRF), and epifluorescence microscopy followed by deconvolution using Autodeblur software (Autoquant, Albany, New York). A 3-D piezoelectric stage provides high-resolution positioning, such that molecular-scale analysis can be associated with subcellular regions of live adherent cells. Much of this hardware is commercially available, making these techniques accessible to researchers in biology and mechanobiology. The engineering challenge is to integrate these modalities on a single platform. In response to this challenge, microscope manufacturers have developed microscopes with multiple input and output ports and increased infinity space for the introduction of custom optics. To make this setup accessible to other researchers, we list all the component parts in Table 1 . For detailed description of our multimodal microscopy setup the reader is referred to Ferko 10, 31 Table 1Components used in construction of the CMDM. The CMDM is comprised of modules for single molecule spectroscopy, multimodal microscopy, and an optical trap.
3.Instrumentation and SetupIn this section we describe the details of the microscope and associated optics. All parts, part numbers, and manufacturers are listed in Table 1. 3.1.Microscope, Optics, and CameraThe CMDM is based on an Olympus IX71 and associated optics firmly secured to a vibration table. In the diascopic light path, a 100-W halogen light provides brightfield illumination for phase contrast and DIC microscopy. Our DIC setup employs a high-NA objective (LUMPlanFl NA, WD ) as a condenser to increase contrast over traditional condenser designs. We use a water-immersion objective (UPLAPO NA, WD ) or an oil-immersion objective (PLAPO NA, TIRFM-2, WD ) for DIC, FCS, fluorescence lifetime measurements, and TIRF (oil only). For TIRF, laser light is fiber-coupled to the TIRF illuminator on one end of the epiport and is focused off-center on the back aperture of the objective. Due to the high NA (1.45) of the objective, the incoming light is transmitted to the glass coverslip at an angle greater than the critical angle and is internally reflected there, resulting in an evanescent wave of light that decays exponentially with distance away from the coverslip. The fluorescent molecules at this interface within a distance of from the top of the coverslip are selectively excited. The angle of incidence of the laser beam at the interface is controlled by a micrometer on the TIRF illuminator attachment. A CCD camera is attached to the left side port of the microscope. The camera exhibits high resolution and sensitivity due to the small CCD sensor element size ( pixels; pixel size), high quantum efficiency (65%) in the visible range, and a two-stage Peltier cooling system (capable of attaining a temperature of ). The cooling fan is placed remotely and connected to the camera by flexible plastic tubing to minimize transmission of vibrations from the fan to the microscope. The camera was modified with a fast shutter to enable the collection of two successive images with an interval of . 3.2.Laser SourcesA cw krypton-argon-ion laser with stable mode laser light is used for both fluorescence correlation spectroscopy and TIRF illumination. The laser is air cooled and is placed away from the main table to reduce the propagation of the vibrations from the cooling fan to the microscope. The laser contains a prism-based micrometer adjustment to deliver wavelengths of 476, 483, 488, 496, 514, 520, 530, 568, 647, and . A fiber coupling system has four adjustable screws to align the fiber with the laser. Once one fiber is aligned properly, multiple fibers can be used interchangeably with minimal realignment. An additional red, pulsed-diode laser is used for time-resolved fluorescence lifetime and FCS experiments in the far red excitation wavelength range. The average power of the diode laser is to , with a repetition rate of and a pulse width of . For lifetime measurements on dyes in the green excitation wavelength range, we use a water-cooled , , , pulsed solid state laser ( , High-Q Laser, Hohenems, Austria). 3.3.3-D Piezostage and Nanometer-Scale PositioningFor high-precision and high-resolution 3-D positioning, we employ a piezoelectric microscope stage. The stage consists of a manual microstage and a three-axis piezo nanostage with of travel in each of three orthogonal directions. The nanostage can carry a load of up to . Custom LabView code controls the movement of the nanostage and coordinates 3-D positioning of the focused laser beam with images from the camera. 3.4.TCSPC and Single-Molecule Fluorescence SpectroscopyFluorescence spectroscopy optics are connected to the right side port of the microscope (Fig. 2 ). The laser light exits the fiber, is expanded and collimated, reflects off the dichroic mirrors, and travels through the side port to the objective, which defines the probe volume and collects emitted fluorescence. The beam diameter is matched to the objective back aperture to ensure a Gaussian excitation profile and full use of the objective NA. Alternative beam parameters are discussed in Hess and Webb.32 A laser power of measured at the objective back aperture ensures a good signal-to-noise ratio (SNR) without significant photobleaching or triplet state formation. Elimination of stray nonfluorescent light is ensured by using emission filters with attenuation optical densities (ODs) of 6 or more. High-quality emission filters are necessary to ensure good SNRs in single-molecule experiments. Collimated fluorescence light is focused with a 250-mm-focal-length lens onto a , optical fiber, which serves as the confocal pinhole. The optical fiber entrance is mounted on a three-axis manual micrometer stage to optimize the confocal probe volume such that autocorrelation curves are consistent with diffusion models. The fluorescence light emerging from the optical fiber is focused onto a photomultiplier tube (PMT) which has high gain and is optimized for single-photon counting for wavelengths of 320 to with a peak wavelength of . The quantum efficiency of the PMT is 40% at peak wavelength (manufacturer’s data) and its response time is . The photon detection event is converted into an electronic pulse, which is routed through an eight-channel router to the TCSPC module for collection and analysis. The TCSPC electronics digitally tag incoming pulses such that the photon arrival time relative to the beginning of the experiment (macrotime) and relative to the laser pulse time (microtime) are measured. A preamplifier prevents overloading of the PMT by shutting it down in response to high photon counts. The data are continuously written into a FIFO (first in, first out) buffer in the SPC-630 board. For a detailed description of the principles behind TCSPC, see the excellent monograph by Becker.33 3.5.Optical TrapThe optical trap was first modeled using ray-tracing software (Optics-Lab, Carlsbad, California) to assist in lens placement. The trap was constructed using C-mount components and is introduced into the expanded infinity space under the microscope objective (Fig. 2). Coherent 830-nm laser light can be continuous or transistor-transistor logic (TTL)-modulated using a 5-V square wave with 50% duty cycle from an analog-to-digital (A/D) board (National Instruments, Austin, Texas). A water-immersion objective is used to focus the collimated and expanded laser beam down to a diffraction-limited spot. The coordinates of the laser focal point corresponding to the image is determined by imaging the focused beam reflecting off the back of a glass coverslip. Dielectric polystyrene beads in a solution of 1% albumin and DPBS were then successfully trapped and tracked with a spatial sensitivity of using time-lapse digital imaging and particle tracking algorithms based on the methods of Gelles 34 TTL-modulation of the laser power enabled precise modulation of trap strength for rapid mechanical testing of 3% gelatin/water solution, which has elastic properties similar to endothelial cell cytoplasm.35 We recently integrated a quadrant photodiode above the condenser, which will enable 3-D tracking of trapped microbeads and photonic force microscopy.36 3.6.Chambers for Micromanipulation and FlowWe designed a temperature-controlled micromanipulation chamber (Fig. 3 ) for TCSPC measurements and imaging of cells and membranes undergoing application of forces with a micropipette or an optical trap. The position of the micropipette inside the chamber is controlled by a motorized, three-axis micromanipulator with a computer interface (MP-285, Sutter Instruments, Novato, California). The micropipette tip is shaped on a microscope-based microforge using a low-melting-temperature glass bead and a micromanipulator. For microaspiration, raising or lowering an adjustable reservoir relative to a reference reservoir produces an intake or outflow of water at the tip of the micropipette. A variable-pressure transducer (Validyne, Northridge, California) attached to both reservoirs generates a voltage proportional to the relative height difference, which is then converted to a digital signal using an A/D board (National Instruments, Austin, Texas). The transducer system is sensitive to pressure changes of ( of water). Chamber temperature is monitored using a thermocouple and maintained by using circulating water from a temperature-controlled water bath (TC500, Brookfield, Middleboro, Massachusetts). The accuracy and stability of the temperature control in the chamber is . For microscopy of live cells under continuous shear flow we use a commercial flow chamber system (FCS2 system, Bioptechs, Butler, Pennsylvania). The chamber has a thermistor-based temperature controller and an objective heater to maintain constant temperature. Physiological pH is maintained by perfusing humidified 5% gas and culture media through an oxygenator (Living Systems Inc., Brattleboro, Vermont). 4.Materials and Methods4.1.Preparation of Dye Solutions for FCS and Fluorescence Lifetime MeasurementsWe obtained 9-(2-(ethoxycarbonyl)phenyl)-3,6-bis(ethylamino)-2,7-dimethyl, chloride [rhodamine 6G (R6G)] powder from Invitrogen (Eugene, Oregon) and used it without further purification. The dye solutions were dissolved and diluted in distilled water to a final R6G concentration ranging from 1 to . Fresh dye solutions were prepared before the start of each experiment. In experiments designed to demonstrate the relationship between diffusion and solvent viscosity, water-based solutions containing of R6G and 10 to 80 % (v/v) of reagent-grade glycerol (Fisher Scientific, Hampton, New Hampshire) were prepared. The bulk viscosities of the prepared solutions were measured using a cone-and-plate viscometer (Haake Rotovisco 1, Thermo Electron, Waltham, Massachusetts). For the simultaneous FCS and lifetime measurements, a pulsed 650-nm red laser was used to excite , 4-chlorobenzenesulfonate (DiD, Probes, Eugene, Oregon) in dimethylsulfoxide (DMSO, EMD chemicals, Gibbstown, New Jersey), ethanol, methanol, or dimethyl formamide (DMF) (Sigma-Aldrich, St. Louis, Missouri). Experiments were performed at room temperature unless otherwise specified. 4.2.Preparation of Giant Unilamellar VesiclesGiant unilamellar vesicles were prepared based on the method by Dimitrov and Angelova37 with minor modifications. 1,2-Dimyristoyl- -glycero-3-phosphocholine (DMPC) (Avanti polar lipids, Alabaster, Alabama) and % perchlorate (Invitrogen, Eugene, Oregon) were dissolved in chloroform. After vacuum drying this mixture onto nitinol wires, the assembly was placed in a chamber containing 0.1-M sucrose and an alternating current was applied across the wire electrodes. A LabView-controlled A/D board (National Instruments, Austin, Texas) was used to apply to the wires 0 to in increments of at 5-min intervals, followed by a constant voltage of for . The sizes of the resulting giant unilamellar vesicles varied from 10 to . 4.3.Cell Culture and Staining ProtocolsBovine aortic endothelial cells (BAECs) (VEC Technologies, Rensselaer, New York) were subcultured between passages 3 and T-25 flasks. Culture medium consisted of MCDB-131 complete medium supplemented with FBS (fetal bovine serum), EGF (epidermal growth factor), and hydrocortisone with addition of antibiotic/antimycotic solution, heparin, and ENDO GRO (VEC Technologies, Rensselaer, New York). Cells were incubated in this medium at with 5% and grown to confluence on No. 1 glass coverslips. Cell membranes were stained using . The stock solution was prepared in DMSO and diluted with phosphate-buffered saline (PBS) (no calcium, magnesium, or albumin) to a final concentration of (for imaging) and (for FCS measurements). After rinsing the cells three times in PBS without calcium or magnesium, the cells were incubated with the staining solution for 3 to at . The cells were rinsed five times and returned to the PBS with calcium, magnesium, and 1% albumin. The cells were maintained at physiological pH and temperature in the incubators and while conducting the FCS measurements. 4.4.Curve Fitting of Autocorrelation and Lifetime DataAutocorrelation curves were fit with Eqs. 6, 7, 8, 9 as appropriate, using the Levenberg-Marquardt nonlinear least-squares regression algorithm with the aid of Origin software (Originlab, Northampton, Massachusetts). Quality of fits was assessed by minimizing residuals and values. The structure factor obtained from the fit was between 2 and all the experiments performed. For the lifetime measurements, the IRF was collected from a sample of a dilute scattering solution of aqueous dairy creamer colloidal particles using the pulsed red diode laser before the start of each experiment. The full width at half maximum (FWHM) of the IRF when the 650-nm diode laser was at a high power setting was calculated to be (see Sec. 5.4). The fluorescence lifetime data of dye in various solvents were collected subsequently. Fluorescence decay curves were extracted from histograms of photon arrival times (relative to laser pulse times) by a process of iterative reconvolution using Eq. 12 in Fluofit software (PicoQuant, Berlin, Germany). Data were fit using a biexponential decay curve with the values of the fit between 1 and 1.5. Using an additional exponential did not result in an improved fit. In some cases, diffusion coefficients and lifetimes are expressed as averages standard deviation. 5.Results5.1.Effects of Dye Concentration on FCS-Determined Diffusion CoefficientsRhodamine 6G (R6G) was used to calibrate FCS because of its known diffusion coefficient ( in water), high quantum yield , and the relatively large absorption cross section38, 39 ( at ). Using these known values for rhodamine dye diffusion, we computed the radius of the focal volume to be . This value compares favorably with the theoretically computed values of the radius of the focal volume of . This experimental radius of the confocal volume was used in all subsequent experiments on cells and vesicles to compute the respective diffusion coefficients. To test whether dye concentration would affect the resulting diffusion coefficients (for example, due to aggregation), we measured the dye diffusion coefficient at progressively increasing concentrations of R6G of 1, 2, 5, 8, and in distilled water at room temperature. For each concentration level, 10 autocorrelation curves were obtained and the resulting diffusion coefficients were averaged (Fig. 4 ). The value of the autocorrelation curve at , , is inversely proportional to the average number of molecules present in the confocal volume [Eq. 6]. Values of increased from 0.3 to 2.49, corresponding to decreasing average particle concentrations of 3.3 to 0.40, which are consistent with the number of molecules calculated from the known concentration of diffusing dye (data not shown). The values of for the five concentrations of R6G measured was approximately (inset of Fig. 4) indicating that diffusion of the R6G dye for very dilute concentrations is independent of the number of molecules in the confocal volume. The effective volume of the confocal volume measured in this study can be approximated using23 the equation , where the radius of the confocal volume is calibrated from the experiment and is the value of the axial half-height obtained from the structure factor. The confocal volume of our one-photon FCS setup calculated in this way is when averaged over all the 50 measurements of the R6G concentrations. The term “one photon” indicates that fluorescence excitation is accomplished using a one-photon mechanism rather than two photons and that confocality is achieved using a pinhole placed in a conjugate image plane. 5.2.Effects of Solvent Viscosity on FCS-Determined Diffusion CoefficientsTo assess the sensitivity of FCS-determined diffusion coefficients to solvent viscosity, fluorescence fluctuation analysis was performed on R6G dye dissolved in aqueous solutions of glycerol concentrations ranging from 10 to 80% glycerol. The solution bulk viscosities were measured in a cone-and-plate viscometer. Ten FCS measurements were collected for each solution, autocorrelation curves were computed, and diffusion coefficients were averaged [Fig. 5a ]. All the experiments were performed at room temperature. The average ranged from a value of for 10% glycerol to for 80% glycerol. Data in Fig. 5b show that FCS-determined diffusion coefficients decrease with increasing viscosity in a manner consistent with Eq. 1. In addition, values of viscosity obtained from FCS-determined diffusion coefficients using Eq. 1 differed from bulk viscosity measurements obtained by cone-and-plate viscometry by 4 to 20%, indicating a good correspondence between the two methods (data not shown). 5.3.FCS Measurements of DMPC Giant Unilamellar Vesicles and BAEC MembranesFCS measurements were performed on DMPC vesicles and endothelial cell membranes stained with dye excited with 520-nm light from the cw laser or 532-nm light from the pulsed solid state laser. Fluorescence light with wavelengths of and longer was collected for analysis. For cells and vesicles, 5-nM was found to best ensure sufficient fluorescence signal and the ease of fit of autocorrelation curves. About were collected and the number of fluorescent molecules in the probe volume at any time was between 1 and 8. The laser beam position in the and plane, relative to the imaging system, was assessed by preparing a single monolayer of on a glass coverslip and bleaching it with the laser. The coordinates of the laser spot were then recorded using the camera. During FCS measurements, the position of the focus in the axis was adjusted to correspond to the membrane by moving the microscope objective such that fluorescence intensity was maximized. For DMPC vesicles, 20 FCS measurements were taken at the giant unilamellar vesicle (GUV) apical region, and no more than 5 measurements were obtained on a single vesicle. The temperature in the chamber was maintained at , which is above the phase-transition temperature for DMPC. The results were fit using the equation for 2-D diffusion [Eq. 7] to obtain a diffusion coefficient for in DMPC vesicles of . In BAECs, moving the stage such that the confocal volume intersected the desired location on the membrane apical surface enabled FCS measurements on spatially distinct points on a cell (Fig. 6 ). FCS measurements were performed on five cells at three points in each cell, and 5 measurements were recorded at each point for a total of 75 measurements. Schwille showed that the diffusion of in a rat basophilic leukemia cell membrane can be assessed using models for 2-D, two-species fit or anomalous diffusion.24 In another recent study, Gielen reported that the diffusion on a nerve cell was well described using a 2-D, two-species fit.40 Consistent with this finding, we observed that a model for 2-D, two-species diffusion fit the autocorrelation curves better than models for either anomalous subdiffusion or 2-D single species. Thus, Eq. 8 was used to determine diffusion coefficients in BAEC membranes. Values given in Table 2 represent averages of five FCS measurements taken on the cell and point indicated. Analysis of a 2-D, two-species fit gives a fast fraction (considered to be unbound dye) and a slow fraction (which is considered to be the membrane mobile fraction). The proportion of dye in the membrane, therefore, ranged from 78 to 93%, with an average value of 83% and exhibited diffusion coefficients ranging from to . The value of the fast fraction (or unbound dye) ranged from 7 to 22% with an average value of 17% with a diffusion coefficient ranging from to . Two-sample test statistics were used to analyze the differences between diffusion coefficients at different points on a given cell with . As an indication of heterogeneity in diffusion in living cells, statistically significant differences in slow-component diffusion between at least two points on a given cell were found in every cell tested (see Table 2). Representative plots of the autocorrelation function, curve fits, and residuals from cell and model membranes are shown in Fig. 7 . Table 2Diffusion coefficients of FCS measurements obtained from BAECs.
2-D, two-species fit was performed on 75 measurements, obtained from five different cells. Three different spatial points were chosen for each individual cell (see Fig. 6). Here,
D1
and
D2
are diffusion coefficients of the fast and slow moving components of
DiI
molecules, respectively; (not shown)
1-b2
and
b2
are their respective relative fractions.Two-sample
t
test analysis was used to determine whether differences in diffusion coefficients between points were statistically significant.Slow component diffusion values ranged from
1.75×10−8
to
7.03×10−8cm2∕s
and averaged
4.38±1.62(×10−8)cm2∕s
[mean
±
standard deviation (SD)]. 5.4.Fluorescence Lifetime MeasurementsWe measured the fluorescence lifetime of DiD dissolved in the polar protic solvents, ethanol and methanol, and polar aprotic solvents, dimethyl sulfoxide (DMSO) and dimethyl formamide (DMF). In contrast to polar aprotic solvents, polar protic solvents have dissociable and can form hydrogen bonds. They also have lower dielectric constants and lower dipole moments than polar aprotic solvents. Since fluorescence lifetime depends on the electronic cloud configuration of the dye molecule, these solvents are expected to affect fluorescence lifetime by altering electron distribution and intramolecular charge transfer. In each solvent, fluorescence lifetimes were best fit using biexponential decays [Eq. 12], consistent with earlier results41 for . Each value reported in Table 3 represents the average of 10 measurements. Lifetime values for ethanol and methanol were 0.48, 1.22 and 0.40, , respectively, suggesting that fluorescence lifetime can be different even in closely related solvents. Buschmann obtained the fluorescence lifetime values of DiD in various solvents.42 Using a single exponential to fit the lifetime curve, they obtained a value of for the decay of DiD in ethanol. In our experiments, the characteristic diffusion times of DiD in DMSO and DMF were 51.2 and , respectively. However, the fluorescence lifetimes of the dye (using a biexponential decay) were 0.74, 1.57 and 1.10, , in DMSO and DMF, respectively (Table 3). Thus, molecules with similar diffusion characteristics can be readily distinguished based on lifetime. Fluorescence emission of DiD in ethanol and methanol is intrinsically weak and therefore autocorrelation curves were noisy and could not be fit with models for FCS. A representative curve showing the fluorescence decay with a fit along with the instrument response function is shown in Fig. 8 . 5.5.Optical Trap MicrorheometryThe value of the optical trap’s spring constant was determined from particle trajectories in water and the equipartition theorem using Hooke’s law. The spring constant of an optical trap was modulated by a TTL pulse delivered to the laser controller. The result was a trap stiffness exerted on polystyrene beads that varied as a square wave. Bead displacement as a function of time depended on this time-varying trap stiffness and the mechanical properties of the matrix surrounding the bead. To demonstrate the interaction of beads and a viscoelastic environment, a mixture of 3% gelatin and filtered water—intended to mimic the cell cytoplasm—was combined with beads in a solution of 1% albumin and DPBS. Trapped-bead locations were imaged and tracked with custom tracking software. When the trap was turned on, a large time-dependent displacement of the bead toward the center of the trap was observed (Fig. 9 ). The time-lag of displacement relative to the laser power indicated that the bead was in a viscoelastic environment. Thus, modulating optical trap forces and measuring the time-dependent displacements of beads provides a convenient means to measure the viscoelastic properties of endothelial cells without moving the cells or the trap location. 5.6.Multimodal MicroscopyThe optical system permits sequential imaging from DIC, epifluorescence, and TIRF imaging modalities with minimal instrument manipulation. As an example, DIC, epifluorescence, and TIRF images of -stained BAECs are shown in Fig. 10 . The great majority of the cells were stained on the membrane. The TIRF imaging is sensitive to membrane basal topography in which bright areas correspond to a membrane that is adherent to the cover glass. Table 3Fluorescence lifetimes of DiD in polar protic (ethanol and methanol) and polar aprotic (DMSO and DMF) solvents.
Pulsed laser excitation
(650nm)
and TCSPC were used to determine fluorescence lifetime. Photon arrival time histograms were fit with a biexponential decay [Eq. 12]. Diffusion of DiD in the polar aprotic solvents, DMF and DMSO, was determined using FCS and Eq. 6. Fluorescence emission of DiD in polar protic solvents, ethanol and methanol, was sufficient for lifetime analysis but too weak to generate an autocorrelation curve which could be fit.NA=not applicable. 6.DiscussionThe CMDM described in this paper was designed to address the following fundamental issues in mechanotransduction research: (1) forces are spatially distributed; (2) relationships between force application and cell perturbation require the integration of continuum and molecular-scale predictive models; and (3) determination of molecular activation requires temporal and spatial registration of single-molecule events in live, intact cells under controlled conditions. To meet these design criteria, we constructed and calibrated a multimodal microscope for cellular imaging and single-molecule fluorescence detection and analysis. First, collection volume of the fluorescence emission was reduced to using a fiber aperture placed in a plane confocal to the laser focus. Second, fluorescence was collected with high-detection efficiency PMTs. Third, noise arising from stray laser light, autofluorescence, and fluorescence from glass coverslips and impurities in the solution was systematically reduced or eliminated by using high-quality optical filters. Fourth, all systems (confocal and trap) were spatially correlated to positions on the cell through careful 3-D calibration of the focus point and imaging plane of the multimodal microscope. 6.1.Improvements in FCS InstrumentationIn contrast to majority of the more traditional FCS instruments, which use avalanche photodiodes (APDs) for photon detection, we employ single photon counting PMTs. These newer PMTs (e.g., Hamamatsu H7422-40P) have a higher quantum efficiency than older generation PMTs, making them suitable for photon correlation studies.33 In addition, PMTs are easier to align (because their active area is larger (squared millimeters versus squared micrometers), less susceptible to photon-induced degradation, and have a superior response time (tens of picoseconds) compared to APDs (hundreds of picoseconds). These properties also make them optimal for fluorescence lifetime measurements. Thus, use of PMTs enables fluorescence lifetime and FCS data to be obtained from a single set of measurements.43 We use software to generate autocorrelation curves from FIFO data obtained from TCSPC instrumentation (i.e., TCSPC-FCS). Hardware autocorrelator cards can also generate autocorrelation curves in near real time, using current fluctuations from APDs or PMTs. However, when combined with pulsed-laser excitation, TCSPC instrumentation records all photon arrival times and saves them into memory directly. This feature makes all the raw photon data available for additional computational analysis necessary for generation of photon counting histograms (PCHs), multispectral fluorescence lifetime imaging (FLIM), burst-induced fluorescence lifetimes (BIFL), and higher order correlations without the necessity for an additional, expensive, hardware correlator.33, 43 Advances in TCSPC instrumentation and modern autocorrelation software algorithms (e.g., the multitau time correlation method) ensure near real time autocorrelation curves that were previously only possible with dedicated hardware correlators.44 Recent studies support the increasing use of TCSPC for FCS and lifetime measurements.43 6.2.FCS Can Detect Subtle Changes in Dye ConcentrationsTo test whether the system could detect subtle changes in concentration of diffusing species, we performed FCS analysis of autocorrelation curves arising from diffusion of rhodamine dye, which varied in concentration from 1 to . These calibration experiments support the use of FCS to determine the absolute number of molecules in the observation volume. Absolute molecular concentrations are essential in the study of the binding kinetics of molecules present on the surfaces of cell membranes and in solution.45 In the context of mechanotransduction, it is thus possible to follow a surface signal transduction pathway by following the association and disassociation of a receptor complex in response to an externally imposed force from fluid flow. One of the main limitations of the technique of FCS is the requirement of extremely low concentrations of probe molecules. We observed degradation of correlation signal with dye concentrations greater than . While the requirement for low concentrations of molecules in FCS makes imaging impractical, it is an advantage in studies of binding kinetics in cells, where the concentrations of aggregating molecules are typically low. The low concentration also means the molecular events in cells can be studied with relatively less perturbation of the biological system compared to methods which require high concentrations of fluorescent molecules such as fluorescence recovery after photobleaching (FRAP) and any imaging-based methods such as fluorescence resonance energy transfer (FRET). 6.3.FCS-Determined Diffusion Coefficients Are Sensitive to Small Changes in ViscosityIn our calibration experiment, we show that increases in glycerol concentration resulted in right-shifted autocorrelation curves indicating longer diffusion times for increasingly viscous environments [Fig. 5a]. Thus, FCS is likely to be sufficiently sensitive to monitor subtle changes in the local microviscosity, which may be biologically significant. In our recent study, we showed that the diffusion of in cellular membranes is not uniform and varies substantially in response to applied shear over the cells.46 Combined with 3-D piezoscanning, FCS can monitor such changes occurring in cells due to imposed fluid flow or other mechanical perturbations to determine spatially resolved mechanical effects on membranes. 6.4.Differences in Diffusion in Cell and Model Membranes Determined by FCSTo test whether the CMDM could detect spatially varying diffusion of lipids in cell and model membranes we measured the fluorescence fluctuations arising from in model DMPC membranes and bovine aortic endothelial cells. Model membranes are useful in studying diffusion in lipid bilayers, phase segregation in lipids, and model protein diffusion in vesicles.47, 48 They are compositionally simple and homogeneous in nature enabling one to fit the simple 2-D equation [Eq. 7] of diffusion to the autocorrelation curves. In DMPC vesicles above the phase transition temperature, we obtained diffusion coefficient values of . More importantly, the small differences of diffusion values obtained within measurements performed on one vesicle and between different individual vesicles support the assumption that the vesicles are homogeneous in nature and that the FCS measurement is repeatable. Almeida 49 and Vaz 50 obtained values of the diffusion coefficients in DMPC/Chol bilayers at various temperatures using FRAP. At , the value of for a pure DMPC bilayer was , which agrees well with the value obtained in the current study. FRAP experiments in other model membranes of POPC (l-palmitoyl-2-oleoyl-sn-glycero-3-phosphocholine) show unrestricted translational diffusion in fluid membrane and restricted diffusion in polymer-stabilized membranes.51 Previous FCS experiments on DLPC (dilauroyl phosphatidylcholine) revealed47 a translational diffusion coefficient of . FCS analysis of fluorescence from stained endothelial cells leads to two important observations. First, the choice of diffusion models used to fit autocorrelation curves of fluorescence in cell membranes is not straightforward, consistent with observations by others.24, 40 To fit our cell lipid diffusion data, we used models for both 2-D, two-species [Eq. 8] and anomalous subdiffusion [Eq. 9]. In general, we found that fitting of autocorrelation curves with a model for 2-D, two-species fit resulted in uncorrelated residuals and smaller values. The spatially heterogeneous nature of the cell membrane necessitates the requirement for a complex fit such as that for the 2-D, two-species fit. In our study, the 2-D, two-species fit reveals the existence of a fast-moving component, whose origin is unclear but may reflect unbound dye, as has been suggested in other studies.40 It has been speculated that the fast component of the diffusion might arise from the unwashed dye present in the focal volume.40 Measurements of dye in solution away from the membrane support this hypothesis. However, the relatively high fraction (83%) of the mobile membrane component confirms that FCS is effective in measuring cell membrane lipid diffusion. The results obtained by the 2-D, two-species fit compare favorably with membrane diffusion values in the literature24, 40, 46 using either FCS or FRAP. The second important observation in this study is that translational diffusion coefficients for different measurement points on endothelial cells vary significantly (cells 1 to 5, Table 2). The values for the membrane diffusion on different points on the cells used in this study ranged from to . While all the values for membrane diffusion are within the same order of magnitude, the interpoint variation within a single cell or between individual cells used in this study is not trivial, but represents an example of the heterogeneous nature of the membrane surface. The variability in the values of the diffusion coefficient might arise from the complexity of the cellular membrane such as differences in lipid phase and density, local protein concentration and aggregation,52 and membrane interaction with the cytoskeleton and glycocalyx.53 Thus, the ability to obtain FCS data from specific locations is an important step toward assessing the nature of spatial heterogeneity of molecular dynamics in live cells. 6.5.Time-Resolved Fluorescence of DiD in Polar Protic and Aprotic SolventsWe used our pulsed, red laser to generate data for fluorescence lifetime and FCS to test the ability of the system to deflect subtle differences in how DiD interacts with its immediate surroundings. The measurement of fluorescence lifetime is independent of dye concentration and is solely determined by changes in the dye’s photophysics induced by inter- and intramolecular interactions. Data reported in Table 3 suggest that the same fluorescent molecule, DiD, has longer fluorescence lifetimes in aprotic solvents compared to protic solvents (Table 3). Fluorescence lifetime depends on nonradiative decay mechanisms. Thus, the shorter fluorescence lifetimes in protic solvents may be due to such nonradiative decay mechanisms as hydrogen bonding and intermolecular proton transfer. Newer modalities of FLIM exploit this ultrasensitivity of a fluorophore for changes in its immediate surrounding to generate spatial maps of lifetime.43 It is expected that lifetime analysis will enable rapid mapping of mechanically induced, molecular-scale perturbation of the dye-lipid-water microenvironments in intact cell membranes. 6.6.Optical Trap Microrheometry of Viscoelastic Gelatin SolutionsAn additional goal of this project was to integrate a means for mechanical testing and force application on single adherent cells. Thus, we constructed an optical trap which used a simple, well-controlled TTL waveform to modulate the trap stiffness to assess the time-dependent mechanical properties of living cells. A TTL-modulated optical trap was successful in identifying viscous and elastic properties of a gel matrix, but further analysis is required to ensure that polystyrene beads are attached to the gel matrix. Future research will also include the development of a mathematical model of viscoelasticity for a time-dependent spring constant. 7.ConclusionsIntegrated instrumentation and interdisciplinary approaches will help solve biological and medical problems originating from the interface of cellular signaling and cell mechanics, an emerging area of biology. Finite element modeling, Monte Carlo simulations, and molecular dynamics can provide predictive multiscale models of cell behavior. However, due to the level of complexity involved in the cellular signal transduction and inherent compartmentalization of cellular signaling in microenvironments, verification of models must be done at the single-molecule level on intact cells with high-precision spatial registration. Alterations in the dynamics of molecules are assumed to accompany changes in their signaling (e.g., phosphorylation and dimerization). Such changes are detectable by FCS and fluorescence lifetime analysis, making single-molecule spectroscopy an ideal tool to address the precise molecular mechanism by which forces induce changes in cellular biology. Some possible future avenues of investigation are monitoring the diffusion-related kinetics of transmembrane proteins such as G-protein-coupled receptors in response to externally applied forces due to fluid flow, micropipette aspiration, or an optical trap. One of the authors (P.J.B.) has performed earlier work46 showing changes in membrane fluidity in response to shear flow using FRAP. The study can be extended to spatially map the changes occurring on a plasma membrane with single-molecule sensitivity using the current instrumental setup. The current setup can also potentially be used to follow the differential kinetics of protein molecules at the apical and basal surfaces of the cell by combining the techniques of one photon FCS with TIR. In conclusion, we have successfully built a multimodal microscope that combines microscale monitoring of molecular dynamics at multiple time scales with prescribed macroscale force conditions. This system addresses an unmet need for integration of force application, analysis of cell mechanics, and molecular-scale mechanotransduction detection. It is now possible to forge new research directions in which cell-specific, multicomponent models of mechanotransduction are developed from 3-D live cell imaging and validated with molecular-scale biological readouts. AcknowledgmentsWe would like thank Ahmed Heikal, PhD (Penn State), and Axel Bergman, PhD (Becker-Hickl), for their helpful discussions and trouble-shooting of TCSPC. This work was supported in part by a grant to P.J.B. from the National Heart Lung and Blood Institute (R01 HL 077542-01A1), by a National Science Foundation Career Award to P.J.B. (BES 0238910), and by a seed grant to P.J.B. from the Center for Optical Technologies, Bethlehem, Pennsylvania. R.M. was supported by the Penn State Biomaterials and Bionanotechnology Summer Institute (NIBIB-NSF EEC-0234026). ReferencesA. W. Orr,
B. P. Helmke,
B. R. Blackman, and
M. A. Schwartz,
“Mechanisms of mechanotransduction,”
Dev. Cell, 10 11
–20
(2006). 1534-5807 Google Scholar
D. E. Ingber,
“Mechanobiology and diseases of mechanotransduction,”
Ann. Med., 35 564
–577
(2003). 0785-3890 Google Scholar
P. F. Davies,
“Overview: temporal and spatial relationships in shear stress-mediated endothelial signalling,”
J. Vasc. Res., 34 208
–211
(1997). 1018-1172 Google Scholar
K. A. Barbee,
T. Mundel,
R. Lal, and
P. F. Davies,
“Subcellular distribution of shear stress at the surface of flow-aligned and nonaligned endothelial monolayers,”
Am. J. Physiol., 268 H1765
–H1772
(1995). 0002-9513 Google Scholar
A. B. Mathur,
G. A. Truskey, and
W. M. Reichert,
“Atomic force and total internal reflection fluorescence microscopy for the study of force transmission in endothelial cells,”
Biophys. J., 78 1725
–1735
(2000). 0006-3495 Google Scholar
A. Trache and
G. A. Meininger,
“Atomic force-multi-optical imaging integrated microscope for monitoring molecular dynamics in live cells,”
J. Biomed. Opt., 10
(1–17), 064023
(2005). https://doi.org/10.1117/1.2146963 1083-3668 Google Scholar
Y. Wang,
E. L. Botvinick,
Y. Zhao,
M. W. Berns,
S. Usami,
R. Y. Tsien, and
S. Chien,
“Visualizing the mechanical activation of Src,”
Nature (London), 434 1040
–1045
(2005). https://doi.org/10.1038/nature03469 0028-0836 Google Scholar
H. Huang,
C. Y. Dong,
H. S. Kwon,
J. D. Sutin,
R. D. Kamm, and
P. T. So,
“Three-dimensional cellular deformation analysis with a two-photon magnetic manipulator workstation,”
Biophys. J., 82 2211
–2223
(2002). 0006-3495 Google Scholar
G. T. Charras and
M. A. Horton,
“Determination of cellular strains by combined atomic force microscopy and finite element modeling,”
Biophys. J., 83 858
–879
(2002). 0006-3495 Google Scholar
M. C. Ferko,
B. P. Patterson, and
P. J. Butler,
“High-resolution solid modeling of biological samples imaged with 3-D fluorescence microscopy,”
Microsc. Res. Tech., 69
(8), 648
–655
(2006). 1059-910X Google Scholar
E. A. Osborn,
A. Rabodzey, C. F. Dewey Jr., J. H. Hartwig,
“Endothelial actin cytoskeleton remodeling during mechanostimulation with fluid shear stress,”
Am. J. Physiol.: Cell Physiol., 290 C444
–C452
(2006). 0363-6143 Google Scholar
C. Sultan,
D. Stamenovic, and
D. E. Ingber,
“A computational tensegrity model predicts dynamic rheological behaviors in living cells,”
Ann. Biomed. Eng., 32 520
–530
(2004). https://doi.org/10.1023/B:ABME.0000019171.26711.37 0090-6964 Google Scholar
N. Wang and
Z. Suo,
“Long-distance propagation of forces in a cell,”
Biochem. Biophys. Res. Commun., 328 1133
–1138
(2005). https://doi.org/10.1016/j.bbrc.2005.01.070 0006-291X Google Scholar
R. V. Geiger,
B. C. Berk,
R. W. Alexander, and
R. M. Nerem,
“Flow-induced calcium transients in single endothelial cells: spatial and temporal analysis,”
Am. J. Physiol., 262 C1411
–C1417
(1992). 0002-9513 Google Scholar
I. Levitan,
A. E. Christian,
T. N. Tulenko, and
G. H. Rothblat,
“Membrane cholesterol content modulates activation of volume-regulated anion current in bovine endothelial cells,”
J. Genet. Physiol., 115 405
–416
(2000). Google Scholar
S. Li,
M. Kim,
Y. L. Hu,
S. Jalali,
D. D. Schlaepfer,
T. Hunter,
S. Chien, and
J. Y. Shyy,
“Fluid shear stress activation of focal adhesion kinase. Linking to mitogen-activated protein kinases,”
J. Biol. Chem., 272 30455
–30462
(1997). https://doi.org/10.1074/jbc.272.48.30455 0021-9258 Google Scholar
S. M. McCormick,
S. G. Eskin,
L. V. McIntire,
C. L. Teng,
C. M. Lu,
C. G. Russell, and
K. K. Chittur,
“DNA microarray reveals changes in gene expression of shear stressed human umbilical vein endothelial cells,”
Proc. Natl. Acad. Sci. U.S.A., 98 8955
–8960
(2001). 0027-8424 Google Scholar
S. T. Hess,
S. Huang,
A. A. Heikal, and
W. W. Webb,
“Biological and chemical applications of fluorescence correlation spectroscopy: a review,”
Biochemistry, 41 697
–705
(2002). https://doi.org/10.1021/bi0118512 0006-2960 Google Scholar
V. Vukojevic,
A. Pramanik,
T. Yakovleva,
R. Rigler,
L. Terenius, and
G. Bakalkin,
“Study of molecular events in cells by fluorescence correlation spectroscopy,”
Cell. Mol. Life Sci., 62 535
–550
(2005). 1420-682X Google Scholar
E. Haustein and
P. Schwille,
“Ultrasensitive investigations of biological systems by fluorescence correlation spectroscopy,”
Methods, 29 153
–166
(2003). 1046-2023 Google Scholar
E. Elson and
D. Magde,
“Fluorescence correlation spectroscopy. I. Conceptual basis and theory,”
Biopolymers, 13 1
–27
(1974). https://doi.org/10.1002/bip.1974.360130102 0006-3525 Google Scholar
D. Magde,
E. Elson, and
W. W. Webb,
“Fluorescence correlation spectroscopy. II. An experimental realization,”
Biopolymers, 13 29
–61
(1974). https://doi.org/10.1002/bip.1974.360130103 0006-3525 Google Scholar
C. Zander,
J. Enderlein, and
R. A. Keller, Single-Molecule Detection in Solution Methods and Applications, Wiley, Berlin (2002). Google Scholar
P. Schwille,
J. Korlach, and
W. W. Webb,
“Fluorescence correlation spectroscopy with single-molecule sensitivity on cell and model membranes,”
Cytometry, 36 176
–182
(1999). https://doi.org/10.1002/(SICI)1097-0320(19990701)36:3<176::AID-CYTO5>3.0.CO;2-F 0196-4763 Google Scholar
J. R. Lakowicz, Principles of Fluorescence Spectroscopy, 2nd ed.Springer, New York (1999). Google Scholar
K. C. Neuman and
S. M. Block,
“Optical trapping,”
Rev. Sci. Instrum., 75 2787
–2809
(2004). https://doi.org/10.1063/1.1785844 0034-6748 Google Scholar
M. Yanai,
J. P. Butler,
T. Suzuki,
H. Sasaki, and
H. Higuchi,
“Regional rheological differences in locomoting neutrophils,”
Am. J. Physiol.: Cell Physiol., 287 C603
–C611
(2004). https://doi.org/10.1152/ajpcell.00347.2003 0363-6143 Google Scholar
R. M. Simmons,
J. T. Finer,
S. Chu, and
J. A. Spudich,
“Quantitative measurements of force and displacement using an optical trap,”
Biophys. J., 70 1813
–1822
(1996). 0006-3495 Google Scholar
L. A. Hough and
H. D. Ou-Yang,
“A new probe for mechanical testing of nanostructures in soft materials,”
J. Nanopart. Res., 1 495
–499
(1999). https://doi.org/10.1023/A:1010075223642 1388-0764 Google Scholar
W. H. Guilford,
D. E. Dupuis,
G. Kennedy,
J. Wu,
J. B. Patlak, and
D. M. Warshaw,
“Smooth muscle and skeletal muscle myosins produce similar unitary forces and displacements in the laser trap,”
Biophys. J., 72 1006
–1021
(1997). 0006-3495 Google Scholar
M. C. Ferko,
M. B. Garcia, and
P. J. Butler,
“Finite element stress analysis of a muticomponent model of sheared and focally-adhered endothelial cells,”
Ann. Biomed. Eng., 35 208
–223
(2007). 0090-6964 Google Scholar
S. T. Hess and
W. W. Webb,
“Focal volume optics and experimental artifacts in confocal fluorescence correlation spectroscopy,”
Biophys. J., 83 2300
–2317
(2002). 0006-3495 Google Scholar
W. Becker, Advanced Time-Correlated Single Photon Counting Techniques, Springer-Verlag, Berlin (2006). Google Scholar
J. Gelles,
B. J. Schnapp, and
M. P. Sheetz,
“Tracking kinesin-driven movements with nanometre-scale precision,”
Nature (London), 331 450
–453
(1988). https://doi.org/10.1038/331450a0 0028-0836 Google Scholar
S. Yamada,
D. Wirtz, and
S. C. Kuo,
“Mechanics of living cells measured by laser tracking microrheology,”
Biophys. J., 78 1736
–1747
(2000). 0006-3495 Google Scholar
A. Pralle,
M. Prummer,
E. L. Florin,
E. H. K. Stelzer, and
J. K. H. Horber,
“Three-dimensional high-resolution particle tracking for optical tweezers by forward scattered light,”
Microsc. Res. Tech., 44 378
–386
(1999). https://doi.org/10.1002/(SICI)1097-0029(19990301)44:5<378::AID-JEMT10>3.0.CO;2-Z 1059-910X Google Scholar
D. S. Dimitrov and
M. I. Angelova,
“Swelling and electroswelling of lipids—theory and experiment,”
Stud. Biophys., 113 15
–20
(1986). 0081-6337 Google Scholar
J. Widengren,
U. Mets, and
R. Rigler,
“Fluorescence correlation spectroscopy of triplet states in solution: a theoretical and experimental study,”
J. Phys. Chem., 99 13368
–13379
(1995). https://doi.org/10.1021/j100036a009 0022-3654 Google Scholar
R. F. Kubin and
A. N. Fletcher,
“Fluorescence quantum yields of some rhodamine dyes,”
J. Lumin., 27 455
–462
(1982). https://doi.org/10.1016/0022-2313(82)90045-X 0022-2313 Google Scholar
E. Gielen,
J. Vercammen,
J. Sykora,
J. Humpolickova,
M. Vandeven,
A. Benda,
N. Hellings,
M. Hof,
Y. Engelborghs,
P. Steels, and
M. Ameloot,
“Diffusion of sphingomyelin and myelin oligodendrocyte glycoprotein in the membrane of OLN-93 oligodendroglial cells studied by fluorescence correlation spectroscopy,”
C. R. Biol., 328 1057
–1064
(2005). Google Scholar
B. S. Packard and
D. E. Wolf,
“Fluorescence lifetimes of carbocyanine lipid analogues in phospholipid bilayers,”
Biochemistry, 24 5176
–5181
(1985). 0006-2960 Google Scholar
V. Buschmann,
K. D. Weston, and
M. Sauer,
“Spectroscopic study and evaluation of red-absorbing fluorescent dyes,”
Bioconjugate Chem., 14 195
–204
(2003). 1043-1802 Google Scholar
W. Becker,
A. Bergmann,
E. Haustein,
Z. Petrasek,
P. Schwille,
C. Biskup,
L. Kelbauskas,
K. Benndorf,
N. Klocker,
T. Anhut,
I. Riemann, and
K. Konig,
“Fluorescence lifetime images and correlation spectra obtained by multidimensional time-correlated single photon counting,”
Microsc. Res. Tech., 69 186
–195
(2006). https://doi.org/10.1002/jemt.20251 1059-910X Google Scholar
M. Wahl,
I. Gregor,
M. Patting, and
J. Enderlein,
“Fast calculation of fluorescence correlation data with asynchronous time-correlated single-photon counting,”
Opt. Express, 11 3583
–3591
(2003). 1094-4087 Google Scholar
A. Pramanik and
R. Rigler,
“Ligand-receptor interactions in the membrane of cultured cells monitored by fluorescence correlation spectroscopy,”
Biol. Chem., 382 371
–378
(2001). 1431-6730 Google Scholar
P. J. Butler,
G. Norwich,
S. Weinbaum, and
S. Chien,
“Shear stress induces a time- and position-dependent increase in endothelial cell membrane fluidity,”
Am. J. Physiol.: Cell Physiol., 280 C962
–C969
(2001). 0363-6143 Google Scholar
J. Korlach,
P. Schwille,
W. W. Webb, and
G. W. Feigenson,
“Characterization of lipid bilayer phases by confocal microscopy and fluorescence correlation spectroscopy,”
Proc. Natl. Acad. Sci. U.S.A., 96 8461
–8466
(1999). https://doi.org/10.1073/pnas.96.15.8461 0027-8424 Google Scholar
K. Bacia,
D. Scherfeld,
N. Kahya, and
P. Schwille,
“Fluorescence correlation spectroscopy relates rafts in model and native membranes,”
Biophys. J., 87 1034
–1043
(2004). 0006-3495 Google Scholar
P. F. Almeida,
W. L. Vaz, and
T. E. Thompson,
“Lateral diffusion in the liquid phases of dimyristoylphosphatidylcholine/cholesterol lipid bilayers: a free volume analysis,”
Biochemistry, 31 6739
–6747
(1992). https://doi.org/10.1021/bi00144a013 0006-2960 Google Scholar
W. L. C. Vaz,
R. M. Clegg, and
D. Hallmann,
“Translational diffusion of lipids in liquid-crystalline phase phosphatidylcholine multibilayers—a comparison of experiment with theory,”
Biochemistry, 24 781
–786
(1985). https://doi.org/10.1021/bi00324a037 0006-2960 Google Scholar
G. J. Schutz,
H. Schindler, and
T. Schmidt,
“Single-molecule microscopy on model membranes reveals anomalous diffusion,”
Biophys. J., 73 1073
–1080
(1997). 0006-3495 Google Scholar
A. Kusumi and
K. Suzuki,
“Toward understanding the dynamics of membrane-raft-based molecular interactions,”
Biochim. Biophys. Acta, 1746 234
–251
(2005). 0006-3002 Google Scholar
J. M. Tarbell and
M. Y. Pahakis,
“Mechanotransduction and the glycocalyx,”
J. Intern Med., 259 339
–350
(2006). 0954-6820 Google Scholar
|