1.IntroductionThe measurement of material deformation is necessary to assess the material mechanical properties. A wide variety of techniques have been developed to measure the deformation of materials under load, ranging from point-wise sensors, such as the resistive strain gauge1 and optical fiber Bragg gratings,2 to two-dimensional (2-D) full-field measurements, including digital image correlation (DIC),3 the grid method,4 speckle interferometry,5 and Moiré interferometry.6 For homogeneous and isotropic materials, these techniques usually provide enough information to investigate their mechanical behavior. However, for materials with more complex structures, such as biological tissues and composites, surface measurements are much less adequate to address their complete mechanical behavior since the deformation may vary significantly between the bulk and the surface. In this case, a depth-resolved three-dimensional (3-D) full-field measurement of the deformation would be highly desirable. Thanks to the development of the various tomographic techniques, digital volume correlation (DVC) has become a popular measurement technique for depth-resolved 3-D deformation fields. It is effectively a 3-D extension of DIC, widely applied to determine the surface deformation fields. DIC was first introduced in the early 1980s3 and has been widely applied in many disciplines, such as mechanical engineering, material science, medical science, etc.7–11 It basically determines the deformation field by tracking a speckle pattern imprinted on the surface of the sample between a reference and a deformed state. Based on the same principle, DVC was developed to measure the internal 3-D deformation behavior of materials by tracking internal features that resemble 3-D speckle patterns contained in the reconstructed volumes.12 It requires sufficient speckle contrast in the reconstructed volumes to ensure the correlation algorithm runs successfully. DVC has made its way into clinical and industrial applications,13–20 where it is mainly applied on x-ray computed tomography data of materials such as composites, metals, foams, and trabecular bones.13–16 In all these cases, the pattern contrast of the reconstructed volume is provided by differences in x-ray absorption of the different constituents or phases of the material. However, for soft biological tissues dominated by collagen, such as cornea, arteries, or skin, optical coherence tomography (OCT) is a more suitable technique to reconstruct the material microstructure. OCT is a noninvasive imaging technique that can acquire micrometer resolution of cross-sectional images from within semitransparent, light scattering materials. It is an interferometric technique that uses a broadband light source with short coherence length to provide depth-resolved information of the object microstructure. The contrast of the images encodes refractive index changes in the material (between fibers and a matrix into which they are embedded, for example). Indeed, this rapidly developing imaging technique has already been applied in ophthalmology, cardiology, gastroenterology, and dermatology,21–24 and many commercially available OCT systems have been developed for clinical and research purposes.25 With the aid of OCT, the volumetric data that represent the whole microstructure of the scanned sample can be reconstructed. The reconstructed data can then be utilized for structure analysis or deformation analysis using DVC. To the best of our knowledge, this is the first time that measurements of depth-resolved deformations have been undertaken combining OCT and DVC. Given a constitutive model, constitutive parameters are constants that describe the mechanical behavior of a material. Thanks to the development of full-field deformation measurement techniques, several methods have been developed to identify the material constitutive parameters. Finite element model updating (FEMU) is a widely applied approach among these methods. It compares the experimental measurements with their numerical counterparts obtained from an FE model and builds up a cost function using the difference between numerical and experimental values in terms of displacement or strain. By minimizing the cost function with respect to the sought constitutive parameters, the solution of the problem can be reached iteratively. FEMU has already been shown to be feasible to identify constitutive parameters in many cases, such as elasticity, hyperelasticity, etc.26–28 This approach, however, exhibits some shortcomings. The initial values to start the iteration procedure, the numerical model, and the quality of the cost function, for instance, can all affect the convergence time and the quality of the results. The virtual fields method (VFM) is an alternative option for solving the identification problem,29 which is based on the principle of virtual work and retrieves the constitutive parameters by utilizing the full-field deformation measurements. This method is more effective than FEMU in terms of computation time since for the latter, an FE model needs to be created and updated iteratively. So far, the VFM has successfully been applied to the identification of constitutive parameters for linear elastic materials such as composites,30–33 elasto-plasticity for metals34–36 as well as hyperelasticity for soft and biological materials such as artery walls.37,38 The aim of this study is to develop an effective methodology to investigate the internal deformation behavior of silicone rubber phantoms under load and to identify their elastic stiffness parameters. This methodology will eventually be applied to study the mechanical properties of biological tissues, such as the vertebrate eye cornea.39 First, a brief description of the material, the experimental setup, the experimental methods as well as the identification theory is presented. Then, the performance of DVC coupled with OCT is evaluated through stationary and rigid body translation tests. After that, DVC results of the tensile tests are examined and critically discussed. Finally, the depth-resolved full-field deformation measurement results are used to identify the material elastic stiffness parameters using VFM and the identification results are also discussed. 2.Materials and Methods2.1.MaterialsIn the present study, two rectangular flat phantom strips were fabricated using silicone gel (MM240-TV), one of them with a notch. The material comprises two parts, a rubber base and a hardener. They were mixed to a ratio of . The nominal Young’s modulus of the silicone rubber is 1.88 MPa, which can change with the proportion of hardener. Since the silicone gel does not present suitable speckle contrast for the application of DVC, copper particles (with an average particle size of ) were seeded into the silicone gel mixture to provide the speckle contrast. The mixture was then put into molds and left to cure at room temperature for . The flat strips were cut from a larger piece using a scalpel to (). Figure 1 shows the rectangular and the notched phantom strips and their loading and observation configurations. 2.2.Experimental Setup and Image AcquisitionThe experimental setup consists of a tensile test fixture and a swept-source OCT system (SS-OCT, Thorlabs OCS1300SS, Ely, United Kingdom, lateral resolution 25 μm, depth resolution 12 μm in air). For the test, the phantom strip was mounted with one end fixed to the fixture and the other end loaded by a dead weight. At first, a dead weight of 50 g was applied as a preload to insure the phantom strip was taut and vertical. This is a necessary step since the silicone rubber is rather compliant. The preload state was taken as the reference state. A 10 g dead weight was then added to the preload and considered as the deformed state, referred to as load step 1 hereafter. For both reference and deformed states, a 3-D volume image sequence of the specimen was acquired using the SS-OCT system. It uses a rapidly tuned narrowband light source with central wavelength of 1325 nm and spectral bandwidth of 100 nm and records the information with a single photodetector. The frequency of the interference signal is proportional to the optical path difference between the sample and reference arms of an interferometer. Depth profiles (A-scans) of the sample are obtained by evaluating the Fourier transform of the signal for each wavelength scan. The 5- to 6-mm coherence length of the laser enables depth measurement. Adjacent A-scans are then synthesized to create an image in the plane. Multiple adjacent 2-D images in the direction then form the reconstructed 3-D volume. In the present work, a region with dimensions of was scanned, corresponding to , , and directions, respectively. A 3-D data volume of was obtained. The acquisition time for each 3-D volume is . It should be pointed out that each time before acquiring the volume images, the specimen was left for 10 min under load to accommodate significant short-term viscoelastic creep. Along the lateral scanning directions and , the image voxel size was determined by dividing the 11 mm dimension by the number of corresponding 1024 voxels, which is equal to 10.7 μm. Regarding the through-thickness direction, which corresponds to the optical path, the voxel size depends on the refractive index of the medium. For the silicone rubber phantom, the voxel size inside the medium along the direction is 4.1 μm, which is determined by dividing its thickness, 1.4 mm, by the number of corresponding voxels, here 345. The reconstructed volume and a typical central transverse -slice () of the specimen are given in Figs. 2(a) and 2(b), respectively. Due to light saturation, the voxels at the top surface exhibit very high intensity values, as can be seen from the white line at the top of the phantom in Fig. 2(b). It should be pointed out that all the regions outside the phantom strip and the saturated voxels at the top surface were masked out in Fig. 2(a) for a better visualization of the 3-D reconstructed volume. The reconstructed volumes for the reference and deformed states were recorded and DVC was then used to compute the 3-D displacement and strain fields. 2.3.Digital Volume CorrelationDVC is the 3-D extension of the widely applied DIC used to measure surface deformations. During the DVC procedure, the correlated volumes are first divided into subvolumes. The displacement vector of each subvolume is determined by tracking and matching the voxels of the subvolumes in the reference and deformed states. This is performed by maximizing the correlation coefficient, which measures the degree of similarity of the gray-level distributions in the subvolumes in the reference and deformed states. The best prediction of the displacement leads to the highest degree of similarity of the gray-level distributions and thus the maximal correlation coefficient (a correlation coefficient value close to 1 indicates a perfect match). In the present study, the displacement fields were calculated using the DaVis® (LaVision) software package based on a fast Fourier transform (FFT) algorithm. It evaluates a normalized cross-correlation coefficient (NCC) defined as where represents the gray level at a voxel in the subvolume of the reference state, while represents the gray level at a point in the subvolume of the deformed state. The coordinates and stand for the same material point in the reference and deformed states, respectively, and are related by the 3-D affine transformation in the form of rigid body motion combined with displacement gradients. where , , and are the rigid body displacement components of the subset center in the , , and directions, respectively. , , and represent the distance between the subset center and the point . A double-pass approach was used whereby large subvolumes were initially used to capture large displacements. Subsequent to this, these initially calculated displacements were used to displace smaller subvolumes and, thus, ensure the pattern was followed and signal-to-noise ratio increased. Gaussian curve-fitting of the correlation function peak was used to detect the position of the displacement with subvoxel resolution. The strains were then determined from the centered finite difference of the calculated displacement fields, without any additional smoothing.2.4.Evaluation of Measurement PerformanceThe DVC algorithm requires sufficient contrast in a subvolume in order to determine a displacement vector. The size of the subvolumes influences the value of the correlation coefficient and, thus, the displacement and strain uncertainties. In the present study, four different subvolume sizes were selected and compared to determine an optimal size, considering the displacement and strain spatial variation as well as the spatial resolution. A double-pass approach used initial subvolume sizes of , , , and in the first pass, followed by 123, , , and in the second pass, and each has 50% overlap with the six adjacent neighbors. Thus, the distance between each subvolume center and its immediate neighbors is 6, 12, 18, and 24 voxels, respectively. Prior to the tensile tests, it is necessary to evaluate the errors caused by all sources of noise and reconstruction uncertainties. This can be done by performing DVC on two subsequent reconstructed volumes of the stationary phantom strip. Since the stationary specimen was not subjected to any applied force, the correlation results should show zero displacement and strain fields over the whole field of view. This is not the case in practice due to the influence of electronic noise in the SS-OCT system, environmental vibration, the volume reconstruction algorithm, etc. Therefore, any nonzero results should be attributed to the contribution of noise and other reconstruction uncertainties. The standard deviations of the spurious strains were calculated to evaluate the resolution (uncertainty) of the strain measurement. Then, two reconstructed volumes were recorded after introducing a rigid body translation of 40 μm () in the direction between the two volumes. This not only tests the effect of all sources of noise, but also tests the performance of DVC subvoxel interpolation (tri-linear in the present study) and accuracy of the correlation algorithm in determining the displacement fields for a translated specimen. The same procedure of data processing as for the stationary test was applied to the rigid body translation test and the strain measurement resolution was computed. The above tests give an overall idea of the resolution of the whole setup so that the significance of the tensile test results can be better analyzed. 2.5.Virtual Fields MethodVFM, which is based on the principle of virtual work, was adopted to identify the constitutive parameters of the silicone rubber phantom. This equation, provided below, is the integral form of the equilibrium equation for the standard deformable continuous solid model. In this equilibrium equation, is the actual stress tensor, is the virtual strain tensor, is the virtual displacement vector, is the traction vector acting on the boundary of the solid volume , is the volume force vector, is the acceleration vector, and is the mass per unit volume; “·” indicates vector dot product and “:” is the contracted product for second-order tensors or the matrix dot product. In this study, the phantom strip was loaded statically and the body forces can be neglected. Therefore, their contribution to the virtual work can be canceled out. Thus, the equilibrium equation becomes This equilibrium equation is valid for any continuous and differentiable virtual displacement field. The actual strain field and the load information are provided by the experiment. The stress components can be expressed through the material constitutive parameters and the strain components through an appropriate constitutive equation. Here, the stress and strain components are written in columns as where is the transpose operator.For the silicone rubber phantom, isotropic elasticity was assumed. Thus, the stress-strain relation can be expressed as where and are the two stiffness components to identify, relating to the elastic modulus and the Poisson’s ratio throughAfter introducing Eqs. (5) and (6) into Eq. (4), the equilibrium equation becomes where is the virtual strain component deduced from the virtual displacement. In the present study, the material is macroscopically homogeneous. Therefore, and can be taken out of the integrals and the equilibrium equation [Eq. (8)] becomesAny new virtual field in the equilibrium equation leads to an equation involving the stiffness components. Therefore, a proper choice of the virtual displacement fields enables the identification of the unknown stiffness components and . There are different methods for choosing virtual fields. In the present work, two manually defined polynomial virtual fields were employed for the two sought constitutive parameters. More details about other choices of virtual fields, such as virtual fields minimizing noise effects, piecewise virtual fields, etc., can be found in Ref. 29. The first virtual displacement field and the corresponding virtual strain field are For the second virtual field, where is the length (dimension along the direction) of the phantom strips.After introducing the above two virtual fields [Eqs. (10) and (11)] into the equilibrium equation [Eq. (9)], a linear equation system [Eq. (12)] can be formed to directly determine the sought stiffness components and . where is the tension load at the position , represents the virtual displacement at the same position calculated from Eq. (10) for the first virtual field, and represents the external virtual work done by the tension load.It should be pointed out that due to the discrete nature of the deformation measurement, the integrals above such as in Eq. (9) must be approximated by discrete sums. For instance, an integral can be approximated by , where is the number of data points, is the actual strain at data point , is the virtual strain at data point , and is the volume of each measurement point. Due to the uniaxial tension configuration, the strain field should be dominated by low spatial frequencies. In addition, a relatively small subvolume size was chosen to enable a high strain spatial resolution. Therefore, the error involved due to the discretization is small and considered as acceptable for the current tests. Thus, matrix in the linear equation system [Eq. (12)] becomes For these tests, of each data point is constant because of the constant subvolume size, which is equal to . Therefore, can be taken out of the sum and a sum in matrix , for instance, becomes where denotes the arithmetic spatial average of and is the total volume of the processed data. Finally, the linear equation system [Eq. (12)] to solve can be written as follows:3.Results and Discussions3.1.Strain Noise AnalysisA fundamental condition in image correlation techniques is that the changes in the intensity pattern are in one-to-one correspondence with the displacements of the surface. DVC results from OCT volume reconstructions of the reference and deformed states are likely to be affected by a variety of factors such as (1) electronic noise of the detectors; (2) light source stability; (3) reconstruction algorithms; (4) contrast reduction through the sample thickness due to material absorption, scattering, dispersion, defocusing, and spectral roll-off; and (5) strain-induced speckle decorrelation. We studied the combined effect of (1) to (4) by performing a stationary test and a rigid translation of the phantoms. The effect of strain in speckle decorrelation was explored through a numerical simulation described below. 3.1.1.Stationary testFrom the DVC results of the stationary test, the influence of subvolume size was analyzed quantitatively by comparing the standard deviations of the strain components for a central -slice for the four final subvolume sizes. All six strain components were derived from the centered finite differences of the spurious displacements as follows, without any smoothing: where the commas stand for the partial derivatives. To study the effect of subvolume size, the standard deviations over the whole field of view are compared in Fig. 3(a) for , , and only, for the sake of legibility. The standard deviations of the strain components drop from to when increasing the subvolume size from to . When further increasing the subvolume size to and [see Fig. 3(b)], the strain fluctuations are further reduced to . This is so because the larger number of features contained in the subvolume assists the convergence of the volume correlation algorithm and enables more accurate tracking of the subvolume deformation. A smaller number of features in the subvolume leads to bigger tracking errors. In the case of thin specimens such as corneas,39 or the phantoms studied in this work, subvolumes were found to be a good compromise in terms of strain and spatial resolution. Depending on the OCT spatial sampling rate, the speckle field may be undersampled, leading to aliasing and interpolation errors.40 This issue will be discussed later on in the paper.Based on the volume strain fields of the stationary test, the standard deviation of each strain component was calculated for each -slice and the results are plotted in Fig. 4(a). It can be observed that the standard deviations of all the strain components generally remain stable along the different -slices, with a slight increase toward the ends. This is expected as the DVC results are usually noisier near edges due to the lack of data. Although with fluctuation, all the standard deviations are generally between and without any smoothing, which is considered as satisfactory compared with the strain levels in the tensile tests later on. 3.1.2.Translation testSlightly larger values of strain standard deviation, between and , were obtained for the rigid body translation test, see Fig. 4(b). This is so as extra sources of error add to those present in the stationary test, mainly subvoxel interpolation error as the DVC algorithm tracks the subvolumes between the reference and displaced states. It was observed that lateral translations (in the plane) lead to strain standard deviation values between those obtained for the stationary and the axial translation tests. These levels are still low compared to the strain levels of in the tensile tests and were thus considered satisfactory. 3.1.3.Strain-induced speckle decorrelationDue to the backscatter illumination/observation configuration, the complex 3-D point spread function (PSF) of the OCT system has fringes across it along the axial direction [ratio between the depth resolution, 8.3 μm, and the half wavelength of the light source in the medium of refractive index 1.45, i.e., . The magnitude of the measured OCT signal corresponds to the convolution between the 3-D PSF and the scattering particles within the phantom. This magnitude, which determines the brightness of the 3-D speckle grain at any particular position within the sample, does not change with rigid body motion of the sample as relative phase differences between scatterers within the 3-D PSF remain constant. In case of strain, however, there is a limit within which the magnitude of the speckle remains nearly unchanged and beyond which an incremental DVC approach would be required. In order to estimate the level of OCT speckle decorrelation due to strain, we performed a 2-D (on the plane) numerical simulation involving the following steps:
Figure 5 shows that NCC drops to for and to for . This latter strain corresponds to a total maximum through-thickness relative displacement of the scatterers in the PSF of . This is equivalent to th of the fringe period inside the PSF and is inversely proportional to the Poisson’s ratio. NCC drops to for , which corresponds to using . This strain level is probably a good estimate to the maximum strain that we can measure with OCT and DVC without using an incremental approach. Above this level, the correlation coefficient is too low to guarantee a good estimate of displacements. When the changes in the magnitude of the interference of light scattered by particles within the PSF are large, the speckle is said to boil, i.e., its structure changes while only the average speckle size is preserved. Even though these results correspond to a 2-D case (B-scan), a 3-D simulation is expected to render results similar to those reported here. Zaitsev et al.43 have found similar results performing a numerical simulation and evaluating the zero mean normalized cross-correlation coefficient as a function of strain. They report that speckle boiling fully decorrelates the speckle for axial strain levels of , which compares well with our figure of mentioned above. 3.2.DVC Results for Tensile Tests3.2.1.Correlation coefficient mapsThe 3-D deformation field was measured under tension after performing DVC on the OCT reconstructed volume data for the rectangular and notched phantom strips. As DVC was performed using the subvolume size of voxel and 50% overlap, the reconstructed volume of interest thus contained measurement points, corresponding to the dimensions of . The reliability of the deformation measurements was assessed in terms of the 3-D correlation coefficient maps shown in Figs. 6(a) and 6(b) for the rectangular and notched phantom strips, respectively. In order to see the correlation coefficients within the specimens, subvolumes of the whole fields are represented here and are obtained by cutting the volumes in the plane at -slice 25. For -slice 18 at the top of the rectangular specimen, the mean value of the correlation coefficient is 0.95, while it is 0.92 for -slice 1 at the bottom. Similar results were obtained for the notched specimen: 0.95 and 0.84, respectively. This decrease in correlation coefficient through the thickness can be attributed to a depth-dependent speckle contrast reduction as a result of signal attenuation due to light scattering, defocusing, and spectral roll-off. In our experiments, the maximum was (see results for load step 2 in Sec. 3.2.2) and therefore speckle boiling was not an issue. Without further analysis it would seem prudent to consider NCC values as inappropriate to measure strains with OCT and DVC. 3.2.2.DVC results for rectangular phantom stripdisplacement maps for the rectangular phantom strip, which denote the displacement along the tension direction, are shown in Fig. 7. In Fig. 7, one can see the evolution of the displacement in cross-sections cut at different -slices. The absolute value of increases continuously along the direction from the fixed side to the other, ranging from 0 to 0.07 mm, as expected from the loading configuration. Figure 8(a) shows that the mean value of evolves linearly along the direction. Nevertheless, a sinusoidal oscillation is apparent when the difference between the actual values and a linear fit is plotted as a function of in Fig. 8(b). An analysis of these displacement oscillations is provided below. The strain maps were derived using the same procedure as for the noise analysis (centered finite difference of the displacement data, without any smoothing). All six strain components for central -slice 25 and central -slice 10 are shown in Figs. 9 and 10, respectively. As expected from the loading configuration, the strain maps for both - and - slices show positive values around 0.007. Strain maps and show negative values, indicating a Poisson’s contraction along the corresponding directions. It is interesting to note that is very small (close to zero) at the right-hand side, where the grip prevents Poisson’s contraction in the direction. Regarding , it is not zero at the right-hand side because the constraint from the grip acts only at the surface. The reason why is actually larger in magnitude may be from some material nonlinearity due to the compression in the grip. Figure 11 shows that the measured strain components are significantly larger than the corresponding strain noise obtained for the stationary test. Since this is a pure tensile test for the rectangular strip, all the shear strain components should be close to noise level, as confirmed in Figs. 9 and 10. Some irregularities, however, can be seen from these strain maps. The fringes observed in the strain component in Figs. 9 and 10 could be due to either a spatial variation of the elastic modulus or interpolation bias.40 In the case of the FFT-based DIC algorithm used here, the period of the oscillation due to interpolation bias corresponds to a displacement equivalent to 2 voxels.44–46 This is consistent with 2.5 fringes observed in Fig. 8(b) for a total deformation corresponding to 5 voxels. A bias in displacement directly leads to bias in strain, proportional to the slope of the displacement bias. When aliasing arises due to spatial undersampling by the SS-OCT system, the displacement values obtained when comparing subvolumes between reference and deformed states are likely to suffer from larger interpolation errors, which are further amplified when strain is calculated. It has been shown that aliasing typically shows up as a Moiré-like fringe pattern in the displacement and strain fields and that it is more obvious in the latter. In order to confirm the nature of the observed fringes, a second load step was performed on the same rectangular phantom strip with an extra 10 g dead weight (we refer to this case as load step 2). The strain maps showed twice as many fringes in Fig. 12, confirming that these are due to interpolation bias in this elastic material. One way of reducing this bias is to perform presmoothing on both the reference and deformed volume images using a Gaussian low-pass filter before correlation to reduce high-spatial frequency components47,48 or to increase the sampling density of the OCT reconstruction. 3.2.3.Results for notched phantom stripdisplacement maps for the notched phantom strip are shown in Fig. 13. One shows the internal displacement in a cross-section cut at central -slice 25, while the other shows the whole volume. It can be seen that in each -slice, the absolute value of increases continuously along the direction from the fixed side to the other, as expected from the loading configuration and consistent with the displacement maps for the rectangular phantom strip as shown in Fig. 7. In Fig. 13, a slight bending of the strip can be observed from the larger displacement values at the top half of the strip compared to those at the bottom half. This is probably due to the slight geometrical asymmetry between the two notches. The geometrical asymmetry was induced during the manufacturing process when cutting the strip to a notched shape from a larger piece. For the notched phantom strip, all six strain components for central -slice 25 and central -slice 10 are shown in Figs. 14 and 15, respectively. The strain map shows positive values, while the and strain maps show negative values, as expected. Larger deformation is expected in the notched region, which can be observed in the maps of the normal strain components in Figs. 14 and 15. In addition, strain concentration is observed near the top notch tip of the strip for the normal strain components in Fig. 15. This is consistent with the larger displacement found near the top notch in Fig. 13. The explanation for this local strain concentration is the geometrical asymmetry of the notched strip, which has already been stated earlier. In Fig. 15, the strain map shows an antisymmetric shear strain distribution. Regarding the other two shear strain components, they are close to zero. These results are consistent for this type of test. It should be pointed out, however, that these results also suffer from the interpolation errors due to aliasing, especially evident in . 3.2.4.Bias reduction using Gaussian presmoothingPresmoothing using a Gaussian filter has proved effective in reducing interpolation bias.47,48 A Gaussian filter with a kernel size of and a standard deviation of 1.5 voxel was applied to the OCT volume reconstructions of the tensile tests prior to correlation. In Fig. 16, for the rectangular phantom strip, the plots of the difference between the measured and the fitted mean displacement of each -slice as a function of position with and without presmoothing are compared. The difference is substantially reduced after presmoothing, which is expected according to Refs. 47 and 48. Gaussian smoothing was also applied to the volume images of the stationary and rigid body translation tests in order to check its influence in the strain resolution. The results show an increase in the strain standard deviations with presmoothing, generally ranging from to , as shown in Figs. 17(a) and 17(b) for the stationary and rigid body translation tests, respectively. The reason for this increase has been explained in Ref. 48, which states that the sum of squares of subset intensity gradient (SSSIG) is reduced after smoothing, and the standard deviation error is inversely proportional to the SSSIG value. In any case, these noise levels may still be considered as acceptable compared with the strain levels of the tensile tests (about one order of magnitude). Figure 18 shows the strain maps with and without presmoothing for the rectangular and notched phantom strips. The fringes due to interpolation bias are eliminated after presmoothing. 3.3.Identification ResultsUsing the experimental strain results in the linear equation system [Eq. (15)], the material stiffness components and can then be directly determined for the two tensile tests. The identification results with and without presmoothing are listed in Tables 1 and 2, respectively. From these results, Young’s modulus and Poisson’s ratio can be calculated through the relations stated in Eq. (7). In order to provide a reference for the stiffness parameters obtained through the VFM, Young’s modulus and Poisson’s ratio were also calculated for the rectangular strip based on the assumption of constant uniaxial stress through the relation. where uniform stress in the cross-section of the rectangular phantom strip was determined through the equation . is the tension load and is the cross-sectional area. and are the average values of the corresponding strain components over the whole field of view. Thus, both Young’s modulus and Poisson’s ratio can be derived and the results are listed in Tables 1 and 2. In both tensile tests, the results obtained from the VFM are consistent with each other and also with those calculated from the constant uniaxial stress assumption. This indicates that the 3-D VFM is an effective tool to identify the constitutive parameters of materials with nonuniform stress/strain states when the constant uniaxial stress assumption is no longer applicable. Moreover, the identification results with presmoothing are in agreement with those without presmoothing when comparing Table 1 to Table 2.Table 1Identified material elastic stiffness parameters without presmoothing.
Table 2Identified material elastic stiffness parameters with presmoothing.
It is interesting to note that when a larger preload of 115 g was used in the tensile test, both parameters increased ( from 1.44 to 1.82 MPa and from 0.42 to 0.48). In both cases, a load of 10 g was used. 4.ConclusionsWe have shown that DVC can provide, by means of a single channel OCT system, multicomponent displacement fields from which all the strain components required by inversion methods such as the 3-D VFM can be evaluated. OCT+DVC has low displacement sensitivity compared to phase-sensitive OCT elastography and seems appropriate for strain as large as (in the axial direction) at which point an incremental approach should be used to avoid speckle decorrelation. A strain uncertainty in the order of to was observed when Gaussian presmoothing is used to reduce bias noise. Strain below this uncertainty level would require an alternative approach, such as phase-sensitive OCT, capable of detecting subwavelength displacements with low noise. In the cases studied in this work, one fringe across a 10 mm field of view would correspond to a relative axial displacement equal to 0.45 μm and an average strain of , an order of magnitude better than the uncertainty we report for OCT+DVC. Even though most phase-sensitive OCT elastography systems have so far focused on phase measurements with only axial sensitivity, a new system with sensitivity to all displacement components has been recently developed based on a wavelength scanning OCT system using multiple illumination directions and a single observation direction.49 In this work, strain was evaluated with a centered finite difference operator applied to the displacement field. No displacement smoothing was used before strain calculation in order to achieve maximal spatial resolution with a view to further studies on thin biological tissues such as the vertebrate eye cornea. In cases where strain accuracy is paramount, a weighted-least squares strain estimator would be more appropriate.50 Uniform and nonuniform 3-D strain fields measured with OCT+DVC were used to identify the elastic stiffness components of rectangular and notched silicone rubber phantoms using the VFM with 3-D manually defined virtual fields. The material moduli extracted from this approach are consistent with those calculated from the constant uniaxial stress approach. In order to test the proposed identification methodology, simple uniaxial tensile tests and isotropic materials were used. This is the first time that volume strain data are derived by performing DVC on OCT reconstructed volumes. Future work will be aimed at applying this methodology to measuring the internal 3-D full-field deformation of biological tissues under more complex loading conditions and also identifying spatial distributions of constitutive parameters. AcknowledgmentsThe authors are thankful to the China Scholarship Council and the Wolfson School of Mechanical Engineering, Loughborough University, for their financial support. They are also thankful to the reviewers for their constructive comments and insight. ReferencesH. E. Couleset al.,
“Residual strain measurement for arc welding and localised high-pressure rolling using resistance strain gauges and neutron diffraction,”
J. Strain Anal. Eng. Des., 47
(8), 576
–586
(2012). http://dx.doi.org/10.1177/0309324712453408 JSADDZ 0309-3247 Google Scholar
C. Fernández-ValdivielsoI. R. MatıasF. J. Arregui,
“Simultaneous measurement of strain and temperature using a fiber Bragg grating and a thermochromic material,”
Sens. Actuators A Phys., 101
(1–2), 107
–116
(2002). http://dx.doi.org/10.1016/S0924-4247(02)00188-7 SAAPEB 0924-4247 Google Scholar
T. C. Chuet al.,
“Applications of digital-image-correlation techniques to experimental mechanics,”
Exp. Mech., 25
(3), 232
–244
(1985). http://dx.doi.org/10.1007/BF02325092 EXMCAZ 0014-4851 Google Scholar
R. Moulartet al.,
“On the realization of microscopic grids for local strain measurement by direct interferometric photolithography,”
Opt. Laser Eng., 45
(12), 1131
–1147
(2007). http://dx.doi.org/10.1016/j.optlaseng.2007.06.009 OLENDN 0143-8166 Google Scholar
W. AnT. E. Carlsson,
“Speckle interferometry for measurement of continuous deformations,”
Opt. Laser Eng., 40
(5–6), 529
–541
(2003). http://dx.doi.org/10.1016/S0143-8166(02)00085-4 OLENDN 0143-8166 Google Scholar
D. PostW. A. Baracat,
“High-sensitivity Moiré interferometry—a simplified approach,”
Exp. Mech., 21
(3), 100
–104
(1981). http://dx.doi.org/10.1007/BF02326365 EXMCAZ 0014-4851 Google Scholar
B. L. Boyceet al.,
“Full-field deformation of bovine cornea under constrained inflation conditions,”
Biomaterials, 29
(28), 3896
–3904
(2008). http://dx.doi.org/10.1016/j.biomaterials.2008.06.011 BIMADU 0142-9612 Google Scholar
V. LibertiauxF. PasconS. Cescotto,
“Experimental verification of brain tissue incompressibility using digital image correlation,”
J. Mech. Behav. Biomed., 4
(7), 1177
–1185
(2011). http://dx.doi.org/10.1016/j.jmbbm.2011.03.028 JMBBCP 1878-0180 Google Scholar
Q. Z. Fanget al.,
“Rate-dependent large deformation behavior of PC/ABS,”
Polymer, 50
(1), 296
–304
(2009). http://dx.doi.org/10.1016/j.polymer.2008.10.042 POLMAG 0032-3861 Google Scholar
Y. WangA. M. Cuitino,
“Full-field measurements of heterogeneous deformation patterns on polymeric foams using digital image correlation,”
Int. J. Solids Struct., 39
(13–14), 3777
–3796
(2002). http://dx.doi.org/10.1016/S0020-7683(02)00176-2 IJSOAD 0020-7683 Google Scholar
E. M. Parsonset al.,
“Three-dimensional large-strain tensile deformation of neat and calcium carbonate-filled high-density polyethylene,”
Polymer, 46
(7), 2257
–2265
(2005). http://dx.doi.org/10.1016/j.polymer.2005.01.045 POLMAG 0032-3861 Google Scholar
B. K. Bayet al.,
“Digital volume correlation: three-dimensional strain mapping using x-ray tomography,”
Exp. Mech., 39
(3), 217
–226
(1999). http://dx.doi.org/10.1007/BF02323555 EXMCAZ 0014-4851 Google Scholar
F. Forsberget al.,
“3-D micro-scale deformations of wood in bending: synchrotron radiation μCT data analyzed with digital volume correlation,”
J. Struct. Biol., 164
(3), 255
–262
(2008). http://dx.doi.org/10.1016/j.jsb.2008.08.004 JSBIEM 1047-8477 Google Scholar
I. JandejsekO. JirousekD. Vavrik,
“Precise strain measurement in complex materials using digital volumetric correlation and time lapse micro-CT data,”
Procedia Eng., 10 1730
–1735
(2011). http://dx.doi.org/10.1016/j.proeng.2011.04.288 PERNBE 1877-7058 Google Scholar
F. ForsbergC. R. Siviour,
“3-D deformation and strain analysis in compacted sugar using x-ray microtomography and digital volume correlation,”
Meas. Sci. Technol., 20
(9), 1
–8
(2009). http://dx.doi.org/10.1088/0957-0233/20/9/095703 MSTCEP 0957-0233 Google Scholar
N. Limodinet al.,
“Influence of closure on the 3-D propagation of fatigue cracks in a nodular cast iron investigated by x-ray tomography and 3-D volume correlation,”
Acta Mater., 58
(8), 2957
–2967
(2010). http://dx.doi.org/10.1016/j.actamat.2010.01.024 ACMAFD 1359-6454 Google Scholar
A. Benoitet al.,
“3-D analysis from micro-MRI during in situ compression on cancellous bone,”
J. Biomech., 42
(14), 2381
–2386
(2009). http://dx.doi.org/10.1016/j.jbiomech.2009.06.034 JBMCB5 0021-9290 Google Scholar
C. Francket al.,
“Three-dimensional full-field measurements of large deformations in soft materials using confocal microscopy and digital volume correlation,”
Exp. Mech., 47
(3), 427
–438
(2007). http://dx.doi.org/10.1007/s11340-007-9037-9 EXMCAZ 0014-4851 Google Scholar
A. Germaneauet al.,
“3-D mechanical analysis of aeronautical plain bearings: validation of a finite element model from measurement of displacement fields by digital volume correlation and optical scanning tomography,”
Opt. Laser Eng., 48
(6), 676
–683
(2010). http://dx.doi.org/10.1016/j.optlaseng.2010.01.010 OLENDN 0143-8166 Google Scholar
S. A. Maskarinecet al.,
“Quantifying cellular traction forces in three dimensions,”
Proc. Natl. Acad. Sci. U. S. A., 106
(52), 22108
–22113
(2009). http://dx.doi.org/10.1073/pnas.0904565106 PNASA6 0027-8424 Google Scholar
T. GambichlerV. JaedickeS. Terras,
“Optical coherence tomography in dermatology: technical and clinical aspects,”
Arch. Dermatol. Res., 303
(7), 457
–473
(2011). http://dx.doi.org/10.1007/s00403-011-1152-x ADREDL 0340-3696 Google Scholar
B. J. Kaluzyet al.,
“Spectral optical coherence tomography: a novel technique for cornea imaging,”
Cornea, 25
(8), 960
–965
(2006). http://dx.doi.org/10.1097/01.ico.0000224644.81719.59 CORNDB 0277-3740 Google Scholar
N. Hutchingset al.,
“Swelling of the human cornea revealed by high-speed, ultrahigh-resolution optical coherence tomography,”
Invest. Ophthalmol. Vis. Sci., 51
(9), 4579
–4584
(2010). http://dx.doi.org/10.1167/iovs.09-4676 IOVSDA 0146-0404 Google Scholar
A. F. Fercher,
“Optical coherence tomography—development, principles, applications,”
Med. Phys., 20
(4), 251
–276
(2010). http://dx.doi.org/10.1016/j.zemedi.2009.11.002 MPHYA6 0094-2405 Google Scholar
D. F. KiernanW. F. MielerS. M. Hariprasad,
“Spectral-domain optical coherence tomography: a comparison of modern high-resolution retinal imaging systems,”
Am. J. Ophthalmol., 149
(1), 18
–31
(2010). http://dx.doi.org/10.1016/j.ajo.2009.08.037 AJOPAA 0002-9394 Google Scholar
M. R. BryantP. J. McDonnell,
“Constitutive laws for biomechanical modeling of refractive surgery,”
J. Biomech. Eng., 118
(4), 473
–481
(1996). http://dx.doi.org/10.1115/1.2796033 JBENDY 0148-0731 Google Scholar
K. AndersonA. E. SheikhT. Newson,
“Application of structural analysis to the mechanical behaviour of the cornea,”
J. R. Soc. Interface, 1
(1), 3
–15
(2004). http://dx.doi.org/10.1098/rsif.2004.0002 1742-5689 Google Scholar
T. D. NguyenB. L. Boyce,
“An inverse finite element method for determining the anisotropic properties of the cornea,”
Biomech. Model Mechanobiol., 10
(3), 323
–337
(2011). http://dx.doi.org/10.1007/s10237-010-0237-3 1617-7959 Google Scholar
F. PierronM. Grédiac, The Virtual Fields Method, Springer, New York
(2012). Google Scholar
F. PierronS. ZhavoronokM. Grédiac,
“Identification of the through-thickness properties of thick laminated tubes using the virtual fields method,”
Int. J. Solids Struct., 37
(32), 4437
–4453
(2000). http://dx.doi.org/10.1016/S0020-7683(99)00149-3 IJSOAD 0020-7683 Google Scholar
M. GrédiacF. PierronY. Surrel,
“Novel procedure for complete in-plane composite characterization using a single T-shaped specimen,”
Exp. Mech., 39
(2), 142
–149
(1999). http://dx.doi.org/10.1007/BF02331118 EXMCAZ 0014-4851 Google Scholar
M. GrédiacF. PierronA. Vautrin,
“The Iosipescu in-plane shear test applied to composites: a new approach based on displacement field processing,”
Compos. Sci. Technol., 51
(3), 409
–417
(1994). http://dx.doi.org/10.1016/0266-3538(94)90109-0 CSTCEH 0266-3538 Google Scholar
M. Grédiac,
“On the direct determination of invariant parameters governing the bending of anisotropic plates,”
Int. J. Solids Struct., 33
(27), 3969
–3982
(1996). http://dx.doi.org/10.1016/0020-7683(95)00220-0 IJSOAD 0020-7683 Google Scholar
M. GrédiacF. Pierron,
“Applying the virtual fields method to the identification of elasto-plastic constitutive parameters,”
Int. J. Plasticity, 22
(4), 602
–627
(2006). http://dx.doi.org/10.1016/j.ijplas.2005.04.007 IJPLER 0749-6419 Google Scholar
Y. Pannieret al.,
“Identification of elasto-plastic constitutive parameters from statically undetermined tests using the virtual fields method,”
Exp. Mech., 46
(6), 735
–755
(2006). http://dx.doi.org/10.1007/s11340-006-9822-x EXMCAZ 0014-4851 Google Scholar
S. Avrilet al.,
“Stress reconstruction and constitutive parameter identification in plane-stress elasto-plasticity problems using surface measurements of deformation fields,”
Exp. Mech., 48
(4), 403
–419
(2008). http://dx.doi.org/10.1007/s11340-007-9084-2 EXMCAZ 0014-4851 Google Scholar
S. AvrilP. BadelA. Duprey,
“Anisotropic and hyperelastic identification of in vitro human arteries from full-field optical measurements,”
J. Biomech., 43
(15), 2978
–2985
(2010). http://dx.doi.org/10.1016/j.jbiomech.2010.07.004 JBMCB5 0021-9290 Google Scholar
J. H. Kimet al.,
“Experimental characterization of rupture in human aortic aneurysms using a full-field measurement technique,”
Biomech. Model Mechanobiol., 11
(6), 841
–853
(2012). http://dx.doi.org/10.1007/s10237-011-0356-5 BMMICD 1617-7959 Google Scholar
J. Fuet al.,
“Assessment of corneal deformation using optical coherence tomography and digital volume correlation,”
in Conf. Proc. SEM,
155
–160
(2013). Google Scholar
M. A. SuttonJ. J. OrteuH. W. Schreier, Image Correlation for Shape, Motion and Deformation Measurements, Springer, New York
(2009). Google Scholar
J. CouplandJ. Lobera,
“Optical tomography and digital holography,”
Meas. Sci. Technol., 19
(7), 070101
(2008). http://dx.doi.org/10.1088/0957-0233/19/7/070101 MSTCEP 0957-0233 Google Scholar
P. D. RuizJ. M. HuntleyJ. M. Coupland,
“Depth-resolved imaging and displacement measurement techniques viewed as linear filtering operations,”
Exp. Mech., 51
(4), 453
–465
(2011). http://dx.doi.org/10.1007/s11340-010-9393-8 EXMCAZ 0014-4851 Google Scholar
V. Y. Zaitsevet al.,
“Elastographic mapping in optical coherence tomography using an unconventional approach based on correlation stability,”
J. Biomed. Opt., 19
(2), 021107
(2014). http://dx.doi.org/10.1117/1.JBO.19.2.021107 JBOPFO 1083-3668 Google Scholar
K. Madiet al.,
“Computation of full-field displacements in a scaffold implant using digital volume correlation and finite element analysis,”
Med. Eng. Phys., 35
(9), 1298
–1312
(2013). http://dx.doi.org/10.1016/j.medengphy.2013.02.001 MEPHEO 1350-4533 Google Scholar
T. AstaritaG. Cardone,
“Analysis of interpolation schemes for image deformation methods in PIV,”
Exp. Fluids, 38
(2), 233
–243
(2005). http://dx.doi.org/10.1007/s00348-004-0902-3 EXFLDU 0723-4864 Google Scholar
G. BesnardF. HildS. Roux,
“Finite-element displacement fields analysis from digital images: application to Portevin-Le Châtelier bands,”
Exp. Mech., 46
(6), 789
–803
(2006). http://dx.doi.org/10.1007/s11340-006-9824-8 EXMCAZ 0014-4851 Google Scholar
P. Lavaet al.,
“Assessment of measuring errors in DIC using deformation fields generated by plastic FEA,”
Opt. Laser Eng., 47
(7–8), 747
–753
(2009). http://dx.doi.org/10.1016/j.optlaseng.2009.03.007 OLENDN 0143-8166 Google Scholar
B. Pan,
“Bias error reduction of digital image correlation using Gaussian pre-filtering,”
Opt. Laser Eng., 51
(10), 1161
–1167
(2013). http://dx.doi.org/10.1016/j.optlaseng.2013.04.009 OLENDN 0143-8166 Google Scholar
S. ChakrabortyP. D. Ruiz,
“Measurement of all orthogonal components of displacement in the volume of scattering materials using wavelength scanning interferometry,”
J. Opt. Soc. Am. A, 29
(9), 1776
–1785
(2012). http://dx.doi.org/10.1364/JOSAA.29.001776 JOAOD6 0740-3232 Google Scholar
B. F. Kennedyet al.,
“Strain estimation in phase-sensitive optical coherence elastography,”
Biomed. Opt. Express, 3
(8), 1865
–1879
(2012). http://dx.doi.org/10.1364/BOE.3.001865 BOEICL 2156-7085 Google Scholar
|