1.IntroductionThe analysis of backscattered light from biological tissue can reveal extensive and valuable information on tissue metabolism and structure. More precisely, diffuse optical techniques enable determination of tissue oxygenation and perfusion states1–3 and are used for tissue differentiation with respect to the physical state,4 structural alignment,5 and malignancy.6 All of this information is accessible by measurement of the absorption coefficient and the reduced scattering coefficient . With the aim to also quantify microscopic structural tissue properties, there is a growing attempt for measurement of subdiffusive light scattering, which is detected at very short distances to the point of incidence.7–18 Based on the small number of scattering interactions, subdiffusive scattering is strongly influenced by the scattering phase function of the scattering particles. The phase function states the probability distribution for light to be redirected into the different scattering angles at each scattering interaction. Generally, this function is heavily dependent on the size, shape, and relative refractive index of the scattering particles. Therefore, subdiffusive reflectance contains additional information on the micro- and ultrastructure of refractive index fluctuations in tissue and may provide insights into the underlying size distribution of scattering particles.10,19 As microstructural properties are often linked to the dysplastic state of tissue,20 there is the aspiration to also use diffuse optical techniques for noninvasive and possibly inter-operative histopathologic differentiation and margin assessment. In our subsequent phantom study, we apply spatial frequency domain imaging (SFDI) to well-defined optical scattering phantoms to retrieve their microscopic scattering contrast.7–9 Similar to a previous study by Kanick et al.,7 we make use of high spatial frequency analysis to image this microscopic and subdiffusive scattering contrast on a macroscopic scale through quantification of the phase function parameter .19,21,22 In contrast to this previous study, we demonstrate subdiffusive imaging on a single heterogeneous yet well-defined tissue phantom and also reveal microscopic contrast for the case of closely matched macroscopic absorption and scattering properties. After a description of our experimental setup in Sec. 2, we present our theoretical evaluation approach in Sec. 3. In this context, we also provide useful instructions for experimental reference and correction mechanisms on spatial frequency domain (SFD) reflectance data. In Sec. 4, we present our subdiffusive imaging experiments and demonstrate both mapping and depth sensitivity of the phase function parameter . Section 5 presents our conclusions 2.Experimental SetupA sketch of our SFDI setup is given in Fig. 1. The projection unit is composed of a halogen light source, which couples into a flexible light guide for subsequent wavelength selection using a filter wheel. The wheel has a selection of eight coated bandpass interference filters for peak transmission at 500, 550, 600, 650, 700, 750, 800, and 900 nm and FWHM of 10 nm (Edmund Optics). Sinusoidal intensity modulation at spatial frequencies () is achieved by a digital micromirror device (0.7 XGA VIS, Discovery 4100 Development Kit, Vialux, Germany) and obliquely projected onto the sample at an angle of 35 deg with digital precorrection for contortion effects. Diffuse reflectance is captured from a imaging area by a cooled CCD camera (QSI 640i, Quantum Scientific Imaging) at a numerical detection aperture of 0.09. Fig. 1Sketch of the used spatial frequency domain imaging setup for oblique projection of spatial frequency patterns at up to eight different projection wavelengths. ![]() Our tissue phantoms are made of transparent epoxy resin with suspended titanium dioxide (, Larit reinweiß, Lange & Ritter, Germany) or aluminum oxide (, T530SG, Bassermann Minerals GmbH, Germany) scattering particles. A detailed description of the phantom fabrication process and the optical and physical properties of the epoxy material, as well as further analysis on the employed scattering particles, can be found in Ref. 23. To probe the subdiffusive differences of the two particle types, two pure phantoms were prepared with either or scattering particles. In order to closely match the reduced scattering value in both phantoms, the weight concentration of particles in the first phantom was times larger than that of the particles in the second phantom. The large difference in weight concentration is based on the much lower scattering efficiency and higher anisotropy of the scatterers owing to their larger diameter and lower refractive index . According to data from the distributing company, the median diameter of the particles is , while we determined the mean diameter of the particles to be using scanning electron microscopy. A similarly large difference exists for the refractive index values with 24 and .25 As is discussed further in Ref. 23, both particle types exhibit variations in shape and often come in angled and elongated nonspherical forms. Phase function modeling using the Mie theory is, therefore, inappropriate. After suspension of the scattering particles in the liquid epoxy resin and subsequent solidification, all phantoms were polished using a series of increasingly fine sandpapers (grit P160-P2500). The smoothness of the phantoms’ surfaces was required to reduce the effect of specular surface reflections. While correcting for this effect, as will be described in Sec. 3, we chose not to use crossed linear polarizers in order for the scalar light propagation model to fully agree with our measurement setting.26 3.Theory and Evaluation ProcedureThe model we apply to our experimental data is a solution to the radiative transfer equation for oblique projection of sinusoidal intensity patterns. This solution was derived by Liemert and Kienle assuming a semi-infinite scattering medium and can be found in Ref. 27. A recent improvement of the solution approach is given in Ref. 28. On the basis of computation time, we limit the computational precision to the order . The presence of two very distinct scattering particle types ( and ) with very different sizes and refractive indices renders precise yet universal phase function modeling difficult. Using forward calculations with goniometrically measured phase function data of the two scattering particles for ,23 we find that the Reynolds-McCormick scattering phase function may serve as an acceptable yet well-established model function to approximate the -reflectance characteristic of both particle types.29 The employed Reynolds-McCormick model phase function takes the two input parameters and and, allows for more variability than the related Henyey-Greenstein function, which is a special case of the former for . Owing to both the approximate nature of our phase function model and uncertainties in our experimental phase function data, our evaluation remains subject to a potential absolute error in of . This error could be determined by numerous forward calculations using both measured and model-based phase function data. The interested reader may also find a comprehensive analysis on the influence of the model phase function on the quantification of the subdiffusive parameter in our recent study in Ref. 22. To best approximate the subdiffusive scattering properties of our scattering particles, we set the input parameter in our Reynolds-McCormick model function and use the following empirical relationship for the input parameter to allow for fitting of independent of any other phase function parameter. This equation is based on a large set of forward calculations and a high accuracy fit () of the relationship.To enable rapid derivation of absorption, reduced scattering, and the subdiffusive phase function parameter , a precomputed look-up table of SFD reflectance values is used by the trust-region-reflective fit algorithm (MATLAB®, MathWorks). The error introduced by the use of the discretized look-up table is negligibly small as could be confirmed by selective evaluation without precomputed values. The four dimensional look-up table contains values for the 12 employed spatial frequencies in the range for many possible combinations of , , and . Each of these dimensions is discretized into 40 parameter steps with a logarithmic spacing of in the range and linear spacing of both and with and . The derivation of the phantoms’ optical properties is not based on raw experimental imaging data, but follows a series of computational postprocessing steps. Some of these steps are of special importance for high spatial frequency analysis. In the following, we will elaborate on three postprocessing steps that account for absolute referencing of the reflectance intensity and correction of the imaging data for system related defocusing and specular surface reflections. The first step is the common and well-known demodulation principle. In this step, the three evenly phase-shifted images , , and are turned into the actual SFD reflectance image . still represents a raw image with pixel values measured in counts. In the next step, these raw values need to be transferred into actual reflectance intensities, which are independent of both the intensity and spatial heterogeneity of the projection system and the spatial heterogeneity of the detection system. This step is often approached by measurement of a spatially homogeneous reference phantom with very well-known optical properties or reflectance intensities. In so doing, measurement inaccuracies from other imaging modalities may be transferred and the underlying scattering phase function of the reference phantom has to be known precisely.In order not to rely on other imaging modalities and possible phase function uncertainties, we follow a slightly different approach by measuring the optical properties of a very homogeneous and intensely scattering epoxy phantom (). The only assumptions we make on this phantom are the spatial homogeneity of its optical properties along with its predetermined refractive index values.23 We then determine the phantom’s optical properties (i.e., , , and ) at every imaging point using the product of light source intensity and detection sensitivity as an additional multiplicative fit variable . In a next evaluation step, the derived optical properties are averaged and assumed constant for all imaging points, which leaves the factor as a single and robust fit variable in a second reevaluation of the phantom. This fit variable quantifies the ratio between and the actual reflectance value as predicted by the theoretical model. All subsequent measurements are multiplied by the obtained map , serving as an intensity reference that quantifies both the light source intensity and the detection efficiency at the same time. Note that is independent of spatial frequency.One further postprocessing step corrects for optical defocusing in the spatial frequency projections, which is especially required for high spatial frequencies and oblique projection geometries. We achieve this by measuring the SFD reflectance from an anodized plate of aluminum. Due to the absence of volume scattering, the reflectance from the aluminum originates solely from the metal surface and the very thin anodization layer, which is typically thick. Using this reflectance measurement, the amount of defocusing at a spatial frequency can be quantified by the ratio , where and corresponds to the aluminum reflectance at spatial frequency zero, which is commonly obtained by averaging , , and at any spatial frequency. For very sharp projections and low spatial frequencies, and no correction is required. For high spatial frequencies, may become much larger based on a decrease in . We perform postcorrection for projection defocusing by multiplying of all subsequent measurements with .In a last correction step, we take into account specular surface reflection. Surface roughness related specular reflections have a potentially very strong impact on high spatial frequency analysis. The use of crossed linear polarizers can diminish this effect while complicating the theoretical modeling of the underlying light propagation.26 Through the use of a nonscattering transparent epoxy phantom with closely matched surface roughness, we were able to quantify the specular reflections and to subtract them from our actual phase measurements. In addition to this correction step, we chose a rather large projection angle in our measurement setup to further reduce the influence of specular surface reflections. At the end of the above postprocessing steps, every computed reflectance image with a pixel resolution of pixels was binned with bin sizes of pixels to reduce the overall data volume and to enhance the signal-to-noise ratio. Especially for high spatial frequency analysis, it is recommended to perform this binning step only after computation of the AC-reflectance value, as binning may otherwise add to the demodulation of the phase data. Both our experimental setup as well as our evaluation approach have been repeatedly verified by comparative studies employing various other measurement modalities, such as spatially and time resolved reflectance spectroscopy on a large set of well-defined tissue phantoms. 4.Experimental ResultsOur initial experiments relate to two pure scattering phantoms with either or as the suspended scattering particles. These two homogeneous phantoms are hardly distinguishable to the naked eye yet have some differences in their reduced scattering and absorption properties. As previously mentioned, both particle types differ strongly in their scattering phase function based on the differences in particle diameter and refractive index. The two phantoms and all further phantoms were characterized with respect to , , and using SFDI at 12 linearly spaced frequencies in the range . Figures 2(a)–2(c) show the result of our measurements on the pure scattering phantoms averaged over five phantom positions versus wavelength . The given error bars correspond to the standard deviation of the positional average. Fig. 2Wavelength-dependent (a) absorption, (b) reduced scattering, and (c) values as measured on two pure epoxy scattering phantoms with either or scattering particles. The two phantoms show very similar reduced scattering and absorption intensities for while differing strongly in . Based on the optical properties of the two pure reference phantoms, (d) a laterally structured phantom with microscopic scattering contrast could be manufactured. ![]() Figure 2(a) reveals similar absorption values in the visible to near-infrared wavelength range for both tissue phantoms. The absorption is found to be very small in both phantoms and is only slightly increased for the phantom based on the intrinsic absorption of scatterers and its higher weight concentration. Figure 2(b) shows the different wavelength dependences of for both phantoms. The two dashed curves in Fig. 2(b) are a fit of the power law to the reduced scattering values yielding and . The stronger decline of for the phantom relates to the smaller size of the scattering particles and their stronger refractive index wavelength dependence. Owing to the different wavelength dependence of the reduced scattering value of both pure scattering phantoms, an almost matched reduced scattering value can be found at the wavelength in Fig. 2(b). In addition to the overall small differences in absorption, this wavelength is well suited for our next step, in which we aim to perform subdiffusive imaging in the near absence of macroscopic absorption and scattering contrast in a combined heterogeneous phantom. Relating to the wavelength dependence of scattering, the microstructural differences of the two pure phantoms become especially apparent in Fig. 2(c), showing the obtained values for both phantoms. The larger size of particles relates to both more forward peaked and reduced high angle scattering leading to less subdiffusive backscattering and a larger value. We compare our results for (dashed curves) to those obtained by goniometric measurements on the same scattering particles in a similar refractive index medium. These comparative values are given by the two solid curves in Fig. 2(c), with the shaded areas representing their measurement errors.23,30 In spite of common features in the compared data curves, we find absolute deviations for both particle types. For , we attribute the systematic deviation of to the approximate nature of our phase function model and residual specular surface reflections, which lower the measured values. The larger deviation for the particles may have its cause in differences in the effective particle size distributions due to slow sedimentation of the bigger-sized particles during phantom curing and the surface insensitivity of the goniometric measurement approach. Using the derived optical properties of the two pure scattering phantoms from Figs. 2(a)–2(c), it is possible to create a laterally heterogeneous phantom composed of both phantom materials with only microscopic subdiffusive scattering contrast at . We achieved this by engraving the name of our institute into the back side of the phantom with a milling machine and a rounded milling head of 4 mm diameter. Using a syringe, the on average 2-mm-deep inscription was then filled with suspended epoxy resin with optical properties and particle concentration equal to that of the reference phantom [see Fig. 2(d)]. After hardening and repeated surface polishing, the lateral structuring of the so obtained phantom is hardly recognizable with the eye based on the small macroscopic scattering and absorption contrast in the visible range. Indeed, according to Fig. 2(b), the composed phantom has hardly any macroscopic scattering contrast at . We used this wavelength to perform SFDI on the heterogeneous phantom in an almost complete absence of and contrast. Figures 3(a)–3(e) give SFD reflectance images [i.e., grayscale maps of ] at various spatial frequencies for and illustrate the increasing microscopic scattering contrast with spatial frequency. These reflectance images were scaled by our intensity reference and corrected for image blurring and specular surface effects, as described previously. Fig. 3Spatial frequency domain (SFD) reflectance images after postprocessing captured at for different spatial frequencies with a camera field of view of . Spatial frequency increases from (a) to (e) with a corresponding decrease in the near maximum penetration depth : (a) , , (b) , , (c) , , (d) , , and (e) , . This causes the inscription with a maximum depth of 2.5 mm to become more and more visible. Scale bars correspond to a length of 10 mm. ![]() The increased subdiffusive backscattering of scatterers gives rise to a strong contrast at spatial frequency . Radiative transfer model based analysis finds the largest absolute derivative of with respect to near this spatial frequency and reveals that only five percent of the SFD reflectance signal is based on photons traveling deeper than 1 mm for the given phantom optical properties.22 We denote this penetration depth by and provide all corresponding depth values in the caption of Fig. 3. To best compare the achieved microscopic imaging contrast of Figs. 3(a)–3(e), the gray-scale coloring was adjusted for every image such that the ratio of pixel values corresponding to the brightest and darkest color is constant among all reflectance images. A very interesting effect can be observed in Figs. 3(c) and 3(d). In these images, the contrast of the engraved letters mostly stems from a bright contour at the inner border of the engraving. This contour is wide and explicable by a slight bias for subdiffusive light in the boundary area to reach the phantom surface on the side. By fitting of our radiative transfer model to the experimental data, optical parameter maps for , , and were obtained and are given in the top row of Fig. 4. As expected from the optical properties of the pure reference phantoms [Figs. 2(a)–2(c)], the laterally structured phantom has very little absorption and reduced scattering contrast at , as can be seen in the left and central parameter maps of Fig. 4. The apparent change in optical properties on the lower and upper left corners of the images is a boundary artifact based on the finite projection area, which extends the imaging field of view by only 10 to to each side.31 The third parameter map in Fig. 4 depicts the strong contrast relating to the different types of scattering particles and, thus, demonstrates the capability of high-frequency analysis for macroscopic imaging of quantified microscopic scattering properties. Fig. 4Derived optical property maps (top row) based on a pixel-by-pixel fit of our radiative transfer equation model to the SFD reflectance data (see Fig. 3). Both the absorption map (a) and reduced scattering map (b) show little contrast due to the similar optical properties of both phantom materials at . The map (c) clearly reveals the achieved subdiffusive scattering contrast. The plots in the bottom row illustrate the obtained optical parameter contrast along the dashed lines given in the corresponding top row parameter maps. In these line profiles, the location of the engraving is indicated by gray shading. ![]() Three line profiles in the bottom row of Fig. 4 further quantify the obtained contrast in the three parameter maps. The line profiles correspond to the single pixel lines indicated by the dashed lines in the parameter images above. The quantification of is noticeably impaired by surface roughness as can be seen by the scratches in the map. The apparent increase in at the left end of the imaging area is an artifact caused by residual specular surface reflections in connection with the oblique projection geometry. The optical properties displayed in Fig. 4 agree well with those measured for the pure reference phantoms (Fig. 2), even though our model based assumptions for a semi-infinite turbid media do not reflect the phantom’s lateral and axial structuring. The absorption map together with its line profile in Fig. 4 depict an almost complete absence of absorption contrast. This corresponds to the relatively long optical path lengths required for such absorption intensities to have a meaningful impact on the reflectance signal. The parameter sensitivity toward is mostly given by intermediate optical path lengths (i.e., intermediate spatial frequencies), which have dimensions similar to the depth and width of our engraving. Therefore, we observe the obtained contrast at wavelengths other than to be diminished (not shown) as compared to the expected contrast based on the homogeneous phantoms. This also reflects the model’s assumption of a semi-infinite and homogeneous sample geometry and its inability to correctly describe the remaining macroscopic scattering heterogeneity.1 On the contrary, we always find the obtained maps also for wavelengths other than to retain most of the contrast anticipated from our reference measurements. This is because the corresponding short photon paths are mostly limited to the inside or the outside of the engraved region. When using maps for the study and assessment of microscopic tissue structure, it is important to know the corresponding sampling depth of the obtained values. To answer this question, we took an experimental approach by adding additional scattering layers on top of the engraved phantom [see Fig. 5(a)]. These layers are made of the same epoxy phantom material and incorporate scatterers in concentrations very similar to that of the phantom engraving. By iteratively adding layers of increasing thickness onto the phantom, the contrast obtained from further reflectance measurements is more and more reduced and eventually disappears. One thereby finds the depth regime that mostly accounts for the quantification of . Fig. 5The use of an added scattering layer of thickness , which when pressed onto the engraved phantom (a) gives rise to a change in the obtained maps (b) to (e) at . The first map (b) is given for comparison and corresponds to the absence of any added layer, while the added effective layer thickness gradually increases from (c) to (e): (c) , (d) , and (e) . ![]() Figure 5 shows the obtained maps at for no added layer [Fig. 5(b)], and effective layer thicknesses of [Fig. 5(c)], [Fig. 5(d)], and [Fig. 5(e)]. Thus, refers to the reduced scattering value of the added layer and to its thickness. During measurement, the layer is squeezed firmly onto the phantom by a custom-built press and both sides of the mounted layer are coated with a microscopy immersion index-matching fluid to reduce Fresnel reflections both on the surface and between the layer and phantom. The fluid greatly reduces the noise in the maps but gives rise to a small interference effect as seen by the bow pattern in Figs. 5(b)–5(e). As expected, the contrast of the inscription diminishes with increasing effective layer thickness and is even canceled at the layer thickness of as can be seen in Fig. 5(e). When comparing the coloring (i.e., the values) within the “ilm”-inscription of Figs. 5(b) and 5(c), a slight decrease in value can be observed, which we attribute to the added interface between the cover layer and phantom. In spite of the used index-matching fluid, the added interface and the finite thickness of the oil layer have a small influence on the backscattering intensity. However, it can be observed in Figs. 5(c)–5(e) that remains almost constant in the engraved area independent of the surface layer thickness, as the surface layer is composed of the same scattering particles. By contrast, the lower subdiffusive backscattering of the outer particles gets more and more masked by the cover layer and is even hidden at the effective thickness of (Fig. 5(e)). As a conclusion from this cover layer experiment, we find the sampling depth of to lie in the range of . This finding is consistent with the basic conception for ballistic light to lose its directionality after a propagation distance of roughly . An even smaller sampling depth of images may be achieved by sole consideration of very high spatial frequencies. 5.ConclusionsThrough our phantom study, we have not only demonstrated the feasibility to fabricate laterally structured microscopic scattering contrast in tissue simulating phantoms, but also described its measurement using high spatial frequency analysis in SFDI. The use of high spatial frequency analysis allows for mapping of the phase function parameter and can, thus, provide macroscopic images of the underlying microscopic structure. In addition to the control of the phantoms’ macroscopic absorption and reduced scattering values, the use of two different scattering agents allowed us to design distinct and well-defined subdiffusive scattering properties. By measurement of , these subdiffusive scattering properties, which ultimately relate to the size and the refractive index of the scattering particles, could be quantified. The presented measurement approach has potential application for in vivo tissue analysis, where the quantification of subdiffusive scattering may provide important insights into the tissue microstructure, which is otherwise only accessible through histology. Dysplastic tissue alterations often come along with major changes in the cellular structure, and it is highly desirable to be able to detect these changes without the need for tissue biopsies. Subdiffusive reflectance imaging represents one promising approach toward such in situ measurements. In view of its potential biomedical application, the strong influence of specular surface reflections even for our almost ideally flat phantoms underscores the need to find strategies to correct for these imaging artifacts in nonideal imaging situations. In a previous study, we reported on the importance of correct phase function modeling for precise quantification.22 For this phantom study, the Reynolds-McCormick phase function was used to model the subdiffusive reflectance characteristics of the employed scattering particles. In view of similar studies on biological tissue, a fractal size distribution of Mie scatterers may provide a useful model for microscopic tissue scattering. In this case, the retrieved values could favorably relate to the fractal dimension and, thus, to the size distribution of the underlying scatterers.19,32–34 With respect to tissue histology, this information may allow for valuable insights into the microscopic tissue structure averaged over the tissue surface to a depth of about . AcknowledgmentsWe acknowledge the support by the Richard and Annemarie Wolf Foundation, Knittlingen, Germany, and by the German Academic Exchange Service through their provision of a DAAD postgraduate fellowship. Furthermore, we thank Thomas Schmid for his assistance in the manufacturing of our optical phantoms. ReferencesD. J. Cuccia et al.,
“Quantitation and mapping of tissue optical properties using modulated imaging,”
J. Biomed. Opt., 14
(2), 024012
(2009). http://dx.doi.org/10.1117/1.3088140 JBOPFO 1083-3668 Google Scholar
A. Mazhar et al.,
“Wavelength optimization for rapid chromophore mapping using spatial frequency domain imaging,”
J. Biomed. Opt., 15
(6), 061716
(2010). http://dx.doi.org/10.1117/1.3523373 JBOPFO 1083-3668 Google Scholar
A. Amelink et al.,
“In vivo measurement of the local optical properties of tissue by use of differential path-length spectroscopy,”
Opt. Lett., 29
(10), 1087
–1089
(2004). http://dx.doi.org/10.1364/OL.29.001087 OPLEDP 0146-9592 Google Scholar
A. Mazhar et al.,
“Noncontact imaging of burn depth and extent in a porcine model using spatial frequency domain imaging,”
J. Biomed. Opt., 19
(8), 086019
(2014). http://dx.doi.org/10.1117/1.JBO.19.8.086019 Google Scholar
B. Yang et al.,
“Polarized light spatial frequency domain imaging for non-destructive quantification of soft tissue fibrous structures,”
Biomed. Opt. Express, 6
(4), 1520
–1533
(2015). http://dx.doi.org/10.1364/BOE.6.001520 BOEICL 2156-7085 Google Scholar
A. M. Laughney et al.,
“Spectral discrimination of breast pathologies in situ using spatial frequency domain imaging,”
Breast Cancer Res., 15 R61
(2013). http://dx.doi.org/10.1186/bcr3455 BCTRD6 Google Scholar
S. C. Kanick et al.,
“Sub-diffusive scattering parameter maps recovered using wide-field high-frequency structured light imaging,”
Biomed. Opt. Express, 5
(10), 3376
–3390
(2014). http://dx.doi.org/10.1364/BOE.5.003376 BOEICL 2156-7085 Google Scholar
V. Krishnaswamy et al.,
“Structured light scatteroscopy,”
J. Biomed. Opt., 19
(7), 070504
(2014). http://dx.doi.org/10.1117/1.JBO.19.7.070504 JBOPFO 1083-3668 Google Scholar
III D. M. McClatchy et al.,
“High spatial frequency structured light imaging for intraoperative breast tumor margin assessment,”
Proc. SPIE, 9313 931308
(2015). http://dx.doi.org/10.1117/12.2080134 PSISDG 0277-786X Google Scholar
S. L. Jacques,
“Probing nanoscale tissue structure using light scattering,”
in Optics in the Life Sciences,
(2015). Google Scholar
R. Samatham and S. L. Jacques,
“Determine scattering coefficient and anisotropy of scattering of tissue phantoms using reflectance-mode confocal microscopy,”
Proc. SPIE, 8592 85920U
(2013). http://dx.doi.org/10.1117/12.2005072 PSISDG 0277-786X Google Scholar
A. Dunn and D. Boas,
“Transport-based image reconstruction in turbid media with small source-detector separations,”
Opt. Lett., 25
(24), 1777
–1779
(2000). http://dx.doi.org/10.1364/OL.25.001777 OPLEDP 0146-9592 Google Scholar
S. L. Jacques, J. R. Roman and K. Lee,
“Imaging superficial tissues with polarized light,”
Lasers Surg. Med., 26 119
–129
(2000). http://dx.doi.org/10.1002/(SICI)1096-9101(2000)26:2<119::AID-LSM3>3.0.CO;2-Y LSMEDI 0196-8092 Google Scholar
F. van Leeuwen-van Zaane et al.,
“In vivo quantification of the scattering properties of tissue using multi-diameter single fiber reflectance spectroscopy,”
Biomed. Opt. Express, 4
(5), 696
–708
(2013). http://dx.doi.org/10.1364/BOE.4.000696 BOEICL 2156-7085 Google Scholar
P. Thueler et al.,
“In vivo endoscopic tissue diagnostics based on spectroscopic absorption, scattering, and phase function properties,”
J. Biomed. Opt., 8
(3), 495
–503
(2003). http://dx.doi.org/10.1117/1.1578494 JBOPFO 1083-3668 Google Scholar
U. A. Gamm et al.,
“Quantification of the reduced scattering coefficient and phase-function-dependent parameter of turbid media using multidiameter single fiber reflectance spectroscopy: experimental validation,”
Opt. Lett., 37
(11), 1838
–1840
(2012). http://dx.doi.org/10.1364/OL.37.001838 OPLEDP 0146-9592 Google Scholar
I. Charvet et al.,
“A new optical method for the non-invasive detection of minimal tissue alterations,”
Phys. Med. Biol., 47 2095
–2108
(2002). http://dx.doi.org/10.1088/0031-9155/47/12/307 PHMBA7 0031-9155 Google Scholar
C. L. Hoy et al.,
“Method for rapid multidiameter single-fiber reflectance and fluorescence spectroscopy through a fiber bundle,”
J. Biomed. Opt., 18
(10), 107005
(2013). http://dx.doi.org/10.1117/1.JBO.18.10.107005 JBOPFO 1083-3668 Google Scholar
S. Chamot et al.,
“Physical interpretation of the phase function related parameter studied with a fractal distribution of spherical scatterers,”
Opt. Express, 18
(23), 23664
–23675
(2010). http://dx.doi.org/10.1364/OE.18.023664 OPEXFF 1094-4087 Google Scholar
R. Drezek et al.,
“Light scattering from cervical cells throughout neoplastic progression: influence of nuclear morphology, DNA content, and chromatin texture,”
J. Biomed. Opt., 8
(1), 7
–16
(2003). http://dx.doi.org/10.1117/1.1528950 JBOPFO 1083-3668 Google Scholar
F. Bevilacqua and C. Depeursinge,
“Monte Carlo study of diffuse reflectance at source-detector separations close to one transport mean free path,”
J. Opt. Soc. Am. A, 16
(12), 2935
–2945
(1999). http://dx.doi.org/10.1364/JOSAA.16.002935 JOAOD6 0740-3232 Google Scholar
N. Bodenschatz et al.,
“Model based analysis on the influence of spatial frequency selection in spatial frequency domain imaging,”
Appl. Opt, 54
(22), 6725
–6731
(2015). http://dx.doi.org/10.1364/AO.54.006725 Google Scholar
P. Krauter et al.,
“Optical phantoms with adjustable subdiffusive scattering parameters,”
J. Biomed. Opt, 20
(10), 105008
(2015). http://dx.doi.org/10.1117/1.JBO.20.10.105008 JBOPFO 1083-3668 Google Scholar
J. R. Devore,
“Refractive indices of rutile and sphalerite,”
J. Opt. Soc. Am., 41 416
–417
(1951). http://dx.doi.org/10.1364/JOSA.41.000416 JOSAAH 0030-3941 Google Scholar
M. J. Dodge, I. H. Malitson and A. I. Mahan,
“A special method for precise refractive index measurement of uniaxial optical media,”
Appl. Opt., 8
(8), 1703
–1706
(1969). http://dx.doi.org/10.1364/AO.8.001703 APOPAI 0003-6935 Google Scholar
J. Wiest et al.,
“Polarization influence on reflectance measurements in the spatial frequency domain,”
Phys. Med. Biol., 60 5717
–5732
(2015). http://dx.doi.org/10.1088/0031-9155/60/15/5717 PHMBA7 0031-9155 Google Scholar
A. Liemert and A. Kienle,
“Spatially modulated light source obliquely incident on a semi-infinite scattering medium,”
Opt. Lett., 37
(19), 4158
–4160
(2012). http://dx.doi.org/10.1364/OL.37.004158 OPLEDP 0146-9592 Google Scholar
A. Liemert and A. Kienle,
“Exact and efficient solution of the radiative transport equation for the semi-infinite medium,”
Sci. Rep., 3 2018
(2013). http://dx.doi.org/10.1038/srep02018 SRCEC3 2045-2322 Google Scholar
L. O. Reynolds and N. J. McCormick,
“Approximate two-parameter phase function for light scattering,”
J. Opt. Soc. Am., 70
(10), 1206
–1212
(1980). http://dx.doi.org/10.1364/JOSA.70.001206 JOSAAH 0030-3941 Google Scholar
F. Foschum and A. Kienle,
“Optimized goniometer for determination of the scattering phase function of suspended particles: simulations and measurements,”
J. Biomed. Opt., 18
(8), 085002
(2013). http://dx.doi.org/10.1117/1.JBO.18.8.085002 JBOPFO 1083-3668 Google Scholar
N. Bodenschatz et al.,
“Sources of errors in spatial frequency domain imaging of scattering media,”
J. Biomed. Opt., 19
(7), 071405
(2014). http://dx.doi.org/10.1117/1.JBO.19.7.071405 JBOPFO 1083-3668 Google Scholar
R. K. Wang,
“Modeling optical properties of soft tissue by fractal distribution of scatterers,”
J. Mod. Opt., 47
(1), 103
–120
(2000). http://dx.doi.org/10.1080/09500340008231409 JMOPEW 0950-0340 Google Scholar
Y. Pu et al.,
“Determination of optical coefficients and fractal dimensional parameters of cancerous and normal prostate tissues,”
Appl. Spectrosc., 66
(7), 828
–834
(2012). http://dx.doi.org/10.1366/11-06471 APSPA4 0003-7028 Google Scholar
M. Xu, T. T. Wu and J. Y. Qu,
“Unified Mie and fractal scattering by cells and experimental study on application in optical characterization of cellular and subcellular structures,”
J. Biomed. Opt., 13
(2), 024015
(2008). http://dx.doi.org/10.1117/1.2907790 JBOPFO 1083-3668 Google Scholar
BiographyNico Bodenschatz is a PhD student in the Materials Optics and Imaging Group under the supervision of Professor Alwin Kienle at the University of Ulm. In 2012, he received his master’s degree in physics from the University of Ulm, with thesis work at the Institute for Solid State Physics. In 2010, he completed his bachelor’s degree in physics from the University of Konstanz, with thesis work in soft matter physics at the University of Pennsylvania. Philipp Krauter received his diploma in physics from the University of Ulm in 2012 and is currently a PhD student in the Material Optics and Imaging Group of Professor Alwin Kienle. He is working on light propagation in highly scattering media. Steffen Nothelfer is a PhD student at the University of Ulm in the Materials Optics and Imaging Group under the direction of Professor Alwin Kienle and is working on modulated imaging and light scattering properties of cells. He received his master’s degree in physics from the University of Ulm in 2013, with thesis work on magnetic force microscopy. In 2011, he gained his bachelor’s degree in physics from the University of Ulm. Florian Foschum received his diploma in medical technology from the University of Applied Sciences, Ulm, in 2007. In 2011, he received his Dr. rer. nat. degree in physics from the University of Ulm. From 2011 until present, he has been working as a postdoctoral researcher in the Material Optics and Imaging Group of Professor Alwin Kienle. Florian Bergmann received his master’s degree in physics from the University of Ulm in 2014 and is currently a PhD student in the Material Optics and Imaging Group of Professor Alwin Kienle. He is working on light propagation in highly scattering and absorbing media within small volumes. André Liemert received his diploma in mechatronics from the University of Applied Sciences, Ulm, in 2007. In 2011, he received his Dr.-Ing. degree in electrical engineering from the University of Ulm. From 2011 until present, he has been working as a postdoctoral researcher in the Material Optics and Imaging Group of Professor Alwin Kienle. Alwin Kienle is the vice director (science) of the Institut für Lasertechnologien in der Medizin und Meßtechnik (ILM), Ulm, Germany, and head of the Materials Optics and Imaging Group at ILM. In addition, he is a professor at the University of Ulm. He studied physics and received his doctoral and habilitation degrees from the University of Ulm. As postdoc, he worked with research groups in Hamilton, Canada, and Lausanne, Switzerland. |