28 January 2017 Simultaneous measurement of refractive index and thickness of multilayer systems using Fourier domain optical coherence tomography, part 1: theory
Author Affiliations +
Abstract
We introduce a theoretical framework for simultaneous refractive index and thickness measurement of multilayer systems using the Fourier domain optical coherence tomography (FD-OCT) system without any previous information about the item under investigation. The input data to the new formalism are the FD-OCT measured optical path lengths and properly selected spectral components of the FD-OCT interference spectrum. No additional arrangement, reference reflector, or mechanical scanning is needed in this approach. Simulation results show that the accuracy of the extracted parameters depends on the index contrast of the sample while it is insensitive to the sample’s thickness profile. For transparent biological samples with smooth interfaces, when the object is in an aqueous medium and has indices <1.55, this method can extract indices and thicknesses with the absolute error 0.001.

1.

Introduction

Optical coherence tomography (OCT) is a noninvasive interferometric imaging technique that provides high-resolution morphological cross-sectional information, thus enabling the three-dimensional (3-D) reconstruction of the reflectivity profile of structures. In the last decade, OCT has become a pervasive imaging technology with particular application in ophthalmology as a standard tool for diagnosis and monitoring of several ocular diseases.1 The first generation OCT system, called time-domain OCT (TD-OCT), was based on mechanically scanning a reference mirror to measure the time-of-flight of the optical signal reflected from the sample.2 Another main category of OCT systems is Fourier-domain OCT (FD-OCT) whereby the reference arm is fixed and the detection system is replaced with a spectrometer. The internal structure of the object is then revealed by Fourier transform of the interference spectrum between the reference and sample arms fields.3 FD-OCT dramatically improves the detection sensitivity and enables significantly higher scan speeds than earlier TD-OCT.46 Current implementations of OCT imaging can provide only the optical path length between reflecting structures. In order to extract the optical properties and the true structure size of the object under investigation, separation of refractive index and thickness from the OCT-measured optical paths is needed.

The ability to noninvasively measure both refractive index and thickness of biological tissues could have broad medical applications. In optical diagnostics, monitoring changes in refractive index can help distinguish between tissue with normal or abnormal function, thus enabling earlier disease detection.7 For example, in ophthalmology, accurate knowledge of the cornea refractive index and thickness can help to improve the outcome of surgical procedures while they are indicators of cornea states of hydration and intraocular pressure.8 Also, the retina layers profile is linked to diseases like glaucoma,9 diabetes,10 and several neuroophthalmic diseases like Alzheimer’s, Parkinson’s, and multiple sclerosis.11 Therefore, accurate measurement of the retina layers true thickness and refractive index profile can help for early diagnosis and prognosis of the diseases. Other applications can be found in brain tumor,12 breast cancer,13 etc.

Several techniques have been reported for simultaneous measurement of refractive index and thickness using an OCT system in both TD and FD. These techniques suitable for single layer objects can be listed in TD as low-coherence reflectometry,1418 focus-tracking method,1929 bifocal optical coherence refractometry,30,31 digital refocusing technique,32,33 and tandem configuration of interferometers for dispersion-insensitive measurements.3436 In the FD approach, the simultaneous measurement of refractive index and thickness of a single layer object has been reported using a reference plane behind the object,37,38 providing a step on the object,39 using a dual sample probe40 or by fitting a modeled signal to the actual spectrum.41

For more than one-layer objects, the focus-tracking technique has been used to measure the refractive index and thickness of two- and three-layer systems (the cornea)42,43 and three layers of the skin.44 This technique employs low-coherence interferometry combined with the translation of the focus of an objective lens inside biological tissues. The measurement accuracy of this method is usually limited by the resolution of translation stages (moving the focus point or the sample) installed in the system. Also, the FD approach has been used for three-layer silica–crown-dental composite–silica structure.45,46 In the proposed method, the real spectral data are fitted to a modeled signal spectrum searching for index and thickness. A computer program searches for a wide range of possible index and thickness combinations, using the optical path length of each layer as constraints until the best fit is achieved. This approach is purely a fitting process without measuring index and thickness.

In the search for a new approach, in this article, we first show that under assumptions consistent with FD-OCT technology, there is a theoretical framework for the analysis of spectral OCT data that decouples the index of refraction and thickness information in stratified media without needing any previous information about the item under investigation. We use computer simulation to implement the methodology associated with the new framework. Our simulated experimental FD-OCT signal will be generated using the transfer matrix method. We then investigate the relationship between the accuracy of this framework and the built-in assumptions for a range of multilayer cases simulating real-world applications.

For simplicity and without losing the generality, we use the new framework to extract two-layer system parameters. These parameters are refractive index and thickness of the layers and the last medium index of refraction. We test the methodology for a range of sample’s thickness and index contrast, searching for any systematic dependence of the extracted parameter errors with respect to sample’s index and thickness profile.

2.

Method

We start with the basic FD-OCT geometry, depicted in Fig. 1, where the field amplitude of the beam from a short coherent light source is divided into two beams, one incident on a reference mirror and one incident on the object. We assume a stratified object with ideally flat interfaces. We also assume that the object is nonabsorbing, isotropic, and homogeneous in refractive index. The fields reflected from the object and reference mirror are recombined onto a spectrometer. We model the spectrometer signal using the simple but often used summation method, which assumes that the field components arising from multiple scattering between interfaces (multiple reflections) inside the object are negligible. This approximation is valid for biological samples.47

Fig. 1

Schematic presentation of the FD-OCT system. {ni,di} represent the index and thickness of each layer of the sample.

JBO_22_1_015002_f001.png

At the beam splitter, the field reflected from the reference mirror has traveled a round trip distance of 2ZR and can be given as follows:

Eq. (1)

UR(k)=r(R)s(k)exp(i2kZR),
where k is the wavenumber, s(k) is the source spectral amplitude, and r(R) represents the reflectivity of the reference mirror, assuming to be unity. The total field reflected from the object at the beam splitter position is the sum of the fields reflected from each interface:45

Eq. (2)

Us(k)=s(k)1N+1r(j)exp(i2kl=0jnldl),
where k is the wavenumber in free space, N is the number of layers, nl and dl are the refractive index and the physical thickness of the l’th layer (n0d0=Z0), and r(j) is the field amplitude reflected from each interface. The reflected field amplitude depends on the incident field amplitude on the interface. For low refractive index contrast, we can make the approximation that the incident signal on each interface is the same and the reflected field amplitude r(j) is the Fresnel coefficient.

The detected signal intensity by the detector after interference between the two-arm fields is as follows:

Eq. (3)

ISUM(k)=|UR+Us|2.
For a two-layer system, Eq. (3) takes the form:47

Eq. (4)

ISUM(k)=S(k)(1+r(1)2+r(2)2+r(3)2)+2S(k){r(1)r(2)cos(2kδ1)+r(1)r(3)cos[2k(δ1+δ2)]+r(2)r(3)cos(2kδ2)}+2S(k){r(1)cos(2kΔZRZ0)+r(2)cos[2k(ΔZRZ0δ1)]+r(3)cos[2k(ΔZRZ0δ1δ2)]},
where S(k) is the source power spectral density, ΔZRZ0=ZRZ00 is the mismatch in arm’s length, and δ1, δ2 are the optical path lengths of each layer. The right-hand side of Eq. (4) contains three groups of terms: the first term is a constant offset (DC), the second is an autocorrelation (AC), and the third is a cross-correlation terms. In the conventional FD-OCT system, the DC and AC terms are considered unwanted terms and can be removed by phase-shifting interferometry.48 The normalized spectral response of the detector, I(k,) after removing the DC and AC terms is as follows:

Eq. (5)

I(k)=ISUM(k)DCAC2S(k)=r(1)cos(2kΔZRZ0)+r(2)cos[2k(ΔZRZ0δ1)]+r(3)cos[2k(ΔZRZ0δ1δ2)].

In our new approach, we first obtain the optical path length of each layer (δ1,δ2), the usual output from a conventional FD-OCT obtained from the Fourier transform of the interference spectrum. When this information is fed back into I(k) in Eq. (5), the only unknowns left are the Fresnel coefficients. For a two-layer system with three interfaces, there are three Fresnel coefficients to be calculated. We may find three appropriate values of k(k1,k2,k3), such that their application in Eq. (5) gives three linearly independent equations. The Fresnel coefficients can be found from the following matrix equation:

Eq. (6)

(r(1)r(2)r(3))=(P)1.(I(k1)I(k2)I(k3)),
where

Eq. (7)

(P)=(cos(2k1ΔZRZ0)cos[2k1(ΔZRZ0δ1)]cos[2k1(ΔZRZ0δ1δ2)]cos(2k2ΔZRZ0)cos[2k2(ΔZRZ0δ1)]cos[2k2(ΔZRZ0δ1δ2)]cos(2k3ΔZRZ0)cos[2k3(ΔZRZ0δ1)]cos[2k3(ΔZRZ0δ1δ2)]).

Not all sets of spectral components can be used to solve Eq. (6) since for some sets, the matrix (P) is singular, as the chosen wavenumber components provide the same phase information from the interfaces. The description of the determinant’s roots can be found in reference.49 Once the Fresnel coefficients are found, the required refractive index may then be extracted directly from the Fresnel coefficients, if n0 is known, using

Eq. (8)

r(j)=nj1njnj1+nj,j=1,2,3,
and thickness of each layer can be extracted then from

Eq. (9)

d1,2=δ1,2n1,2.

In summary, we extract both refractive index and thickness by

  • 1. obtaining the normalized spectrum free from DC and AC terms [Eq. (5)],

  • 2. extracting the optical thickness from the Fourier transform of Eq. (5),

  • 3. selecting appropriate wavenumber components,

  • 4. constructing the matrix equation, Eq. (6),

  • 5. extracting indices from the Fresnel coefficients [Eq. (8)],

  • 6. extracting layer thicknesses from Eq. (9).

It is worth noting that Eq. (6) can be generalized to any multilayer systems. An N-layer system has N+1 interfaces; therefore, we need N+1 spectral components to calculate N+1 Fresnel coefficients, so we have

Eq. (10)

(r(1)r(2)r(N+1))=(P)1(I(k1)I(k2)I(kN+1)),
where the matrix is

Eq. (11)

P=(cos(2k1ΔZRZ0)cos[2k1(ΔZRZ0δ1)]cos[2k1(ΔZRZ0i=1Nδi)]cos(2k2ΔZRZ0)cos[2k2(ΔZRZ0δ1)]cos[2k2(ΔZRZ0i=1Nδi)]cos(2kN+1ΔZRZ0)cos[2kN+1(ΔZRZ0δ1)]cos[2kN+1(ΔZRZ0i=1Nδi)]).

3.

Results

In this section, we test the applicability of the introduced index-thickness measurement method for specific system realizations against digitally constructed test data. A realistic FD-OCT signal will be generated using the transfer matrix method to simulate real life data acquisition. We will use the new formalism to extract two-layer system parameters (i.e., n1, n2, n3, d1, d2) to validate the new framework in a perfect situation when the error in optical thickness measurement is negligible (ϵ=0). We will test the new methodology for a wide range of index contrasts and thickness of two-layer systems to determine if there is any systematic dependence of the “error” (the difference between actual parameters and extracted parameters) with respect to index contrast and thickness. Although there might be applications for the work presented in this article outside of the OCT framework, the work presented here is focused on biological applications of OCT. The more realistic situation when ϵ0 will be discussed in the next section.

3.1.

Impact of Refractive Index Profile

As discussed previously on the summation method, the main constraint on our index-thickness extraction method is that the Fresnel coefficients be small, meaning that the indices difference across an interface (index contrast) be small. As the contrast increases, multiple reflections are no longer negligible and hence, the summation method is not sufficient to describe the reflected spectrum accurately. This induces error in the sample’s extracted parameters using the matrix formulation [Eq. (6)].

To illustrate the impact of refractive index profile on the extracted parameters for biological applications, we assume that a two-layer system is immersed into an aquatic medium with index of refraction 1.337. The extracted parameters error distribution map is provided for a range of n1 and n2 from 1.337 to 1.90 in 0.11 increments while n3=1.337 but must be estimated by the program. Both thicknesses are set to 30  μm and a total of 2600 test models were thus tested. In Fig. 2, the horizontal axis corresponds to the actual n1 value and the vertical axis corresponds to the actual n2 value in the models tested, and the color map is the extracted parameter error. The irregularities appearing on the top right and the narrow band on the bottom of the figure correspond to the models that their optical thicknesses could not be determined correctly by the FD-OCT simulation program due to very large refractive index contrast ratio between the interfaces.

Fig. 2

The contour map corresponding to the absolute error distribution of the extracted (a) n1, (b) n2, and (c) n3. The horizontal axis corresponds to varying n1 and the vertical axis corresponds to varying n2. The region with the absolute error 0.001 for all indices is limited to test models with n1,21.55.

JBO_22_1_015002_f002.png

The error distribution in n1:n2 space for n1 in Fig. 2(a) shows that for n1<1.55, the absolute error is consistently less than 0.001, an area of interest as it is more desirable for biological applications. The region with the same accuracy for n2, Fig. 2(b), is roughly restricted to n1<1.75 and n2<1.55 and for n3, in Fig. 2(c), is limited to the region with n1<1.70 and n2<1.70. From Fig. 2, we may conclude that n1, n2, and n3 can be extracted with the absolute error 0.001 for those test models having both n1,21.55.

The relative error distribution of the extracted thicknesses follows almost the same pattern of the corresponding index since from Eq. (9), we have

Eq. (12)

Δd1,2d1,2=±(Δδ1,2δ1,2)2+(Δn1,2n1,2)2.

3.2.

Impact of Sample Thickness

The impact of the sample thickness on extracting the parameters is evaluated in this section. We consider a two-layer test model with n0=1.337, n1=1.345, n2=1.351, and n3=1.337. The thickness of the first layer is fixed at 30  μm and the thickness of the second layer varies from 30  μm up to 1580  μm, which is the maximum detectable range of our FD-OCT simulation system with 20  μm increments. The parameters to be extracted are the index and thickness of each layer and the last medium index of refraction. The test result is shown in Fig. 3, where the vertical axis is the error in the extracted parameters and the horizontal axis is the varying thickness. In this figure, the left panel shows the error in the refractive indices and the right panel shows the relative error in the thicknesses. It can be seen that the error of index extraction is insensitive to the sample’s thickness profile. Our tests have also shown the same behavior when d2 is fixed and d1 is swept.

Fig. 3

Impact of sample thickness (varying d2) on the accuracy of the sample’s extracted parameters. (a) The errors associated with the refractive indices are shown, and (b) the relative error associated in the thicknesses is shown.

JBO_22_1_015002_f003.png

4.

Discussion

So far, we have discussed only the ideal situation, where the actual optical path lengths are known exactly, where Eq. (6) could be solved for any wavenumber components as long as the determinant of the matrix is not zero. However, in practice, there is always an uncertainty (ϵ) associated with the measured optical path (δ), so the value input into the parameter extraction algorithm may deviate from the actual optical path (δ) so that

Eq. (13)

δ=δ±ϵ.

Since in practice, the matrix in Eq. (6) is constructed from the measured optical path lengths, there is a nonlinear phase shift between the matrix and the signal on the right-hand side of Eq. (6). Therefore, in the presence of ϵ, there will be a constraint on wavenumber selection in addition to the roots of the determinant and inappropriate selection of the wavenumber components can lead to significant errors. Given that this matrix equation presents a coupled set of transcendental equations containing index terms on both sides of the equal signs, the impact of ϵ on index determination is not transparent so the behavior of the extracted parameters is best investigated by simulation. A framework for selecting the suitable spectral components is investigated in the following section.

4.1.

Finding a Suitable Set of Wavenumber Components

We illustrate the impact of the optical path length measurement error on the efficacy of parameter extraction using the following randomly selected two-layer system with indices n0=1.337, n1=1.345, n2=1.351, and n3=1.337 and thicknesses d1=10  μm and d2=30  μm and assuming that ΔZRZ0=0. Errors of ϵ1=+3.0  nm for the first layer and ϵ2=+10.0  nm for the second layer are arbitrarily introduced, to mimic the optical path length measurement error. Although the variation of the errors with ϵ>0 and ϵ<0 is not symmetric, our tests have shown that the whole discussion with ϵ>0 is also valid for ϵ<0. In order to solve Eq. (6), three spectral components are needed (k1,k2,k3). We fix k1 and k2 at two arbitrary values [k1=6.80  (μm)1, k2=6.85  (μm)1] and sweep k3 [from 6.85 to 7.00  (μm)1] to investigate the behavior of the extraction algorithm. The errors in extracting refractive indices are shown in Fig. 4, where the left vertical axis is the error in the extracted indices, the horizontal axis is k3, and the second axis is the plot of the determinant of P.

Fig. 4

The dependence of the extracted refractive indices errors as a function of the third wavenumber. The left axis shows error, and the right axis shows the determinant value. Two of the wavenumbers are chosen at arbitrary values and the third (k3) is scanned. In this illustrative example, n0=1.337, n1=1.345, n2=1.351, n3=1.337, d1=10  μm, and d2=30  μm. The introduced error in OT is ϵ1=+3.0  nm for the first layer and ϵ2=+10.0  nm for the second layer. The chosen wavenumbers are k1=6.80  (μm)1, k2=6.85  (μm)1, and k3 swept from 6.85 to 7.00  (μm)1.

JBO_22_1_015002_f004.png

From Fig. 4, it is seen that the errors in extracting indices vary with wavenumber. The figure also shows that the errors diverge where the determinant is zero and converge around the determinant’s extrema. In fact, better convergence is observed with a larger magnitude of the determinant. This is a systematic behavior of the extracted parameter errors with respect to the wavenumber components and has been observed throughout our large number of tests. Since the determinant value solely depends on the chosen set of the spectral components, we may conclude from the observation in Fig. 4, that in order to reduce (if not minimize) the error of all the extracted parameters, we may look for a set of wavenumbers (k1,k2,k3) that provides large magnitude (if not maximizes) of the determinant.

According to Hadamard’s maximum determinant problem,50 the formal maximum of the 3×3 matrix presented in Eq. (7) with all entries on the closed unit disk is 33/2=5.196. Since the formal approach to maximize the determinant of the matrix is too complicated and out of the scope of this article, we use a heuristic search to find a special set out of many possible sets. Thereby, it may not be the optimal selection but as we will show below it provides excellent estimates of the index parameters. Using the approach described in reference,49 a set of possible wavenumber components for the example case mentioned above is k1=6.52  (μm)1, k2=8.52  (μm)1, and k3=8.64  (μm)1, which generate a determinant of 3.977. In order to show the behavior of the extracted indices errors with the new set of wavenumbers, we fix k1 and k2 at the given values above and sweep k3 from 8.52 to 8.67  (μm)1 and plot the extracted indices errors versus wavenumber (Fig. 5). The determinant of the matrix is also plotted as a reference in the figure (the right axis).

Fig. 5

The dependence of the extracted refractive index errors as a function of wavenumber selection for a two-layer system. The left axis is for refractive index error, and the right axis shows the determinant value. The wavenumbers are obtained from the algorithm that maximizes the determinant. In this illustrative example, n0=1.337, n1=1.345, n2=1.351, n3=1.337, d1=10  μm and d2=30  μm. The introduced error in OT is ϵ1=+3.0  nm for the first layer and ϵ2=+10  nm for the second layer. The chosen wavenumbers are k1=6.52  (μm)1, k2=8.52  (μm)1 and k3 swept from 8.52 to 8.67  (μm)1. The indicator line corresponds to k3=8.64  (μm)1 obtained from the algorithm.

JBO_22_1_015002_f005.png

It can be observed from Fig. 5 that the errors are reduced simultaneously around the maximum of the determinant (indicated by a dashed line on Fig. 5). The errors at the determinant’s maximum are n1actualn1extracted=3.23×104, n2actualn2extracted=2.98×104, n3actualn3extracted=2.79×104, and for thicknesses after dividing the optical path length by the corresponding index of refraction, we obtain d1actuald1extracted=4.62×103  μm and d2actuald2extracted=1.39×102  μm.

5.

Conclusion

In conclusion, we introduced in this article a new framework for simultaneous extraction of refractive index and thickness of multilayer systems, from its FD-OCT data, without any previous information about the item under investigation. No additional arrangement, reference reflector, or mechanical scanning is needed, which is the main advantage of this approach.

In this method, the accuracy of the extracted parameters depends on the refractive index contrast of the object and is insensitive to thickness profile. The index contrast limitation is due to inherent differences between the modeled and actual spectrum. Our theoretical model takes into account only single scattering events, but the actual spectrum contains multiple scattering events. We showed in Sec. 3.1 that in biological applications, when n1,21.55, two-layer systems indices can be extracted with the absolute error 0.001. This implies that the maximum tolerable index contrast is in the range of |1.551.337|=0.233. This limitation is not a problem for the biological system, where the refractive index rarely exceeds 1.45.

We showed in this article that the uncertainty in measuring the optical path lengths (ϵ) has a significant impact on the extracted parameter errors. Two main questions arising here are what is the maximum ϵ that still allows extraction of parameters and how much is the accuracy loss in each parameter as ϵ increases. The impact of large amounts of ϵ, which may occur in practice, is discussed in part 2 of this article. Also, the proposed methodology for selecting the best spectral components in Sec. 4.1 is based on empirical examples and has not been proven to be the best methodology and needs to be validated through a proper theoretical work-up.

Disclosures

The authors have no relevant financial interests in the manuscript and no other potential conflicts of interest to disclose.

Acknowledgments

The authors would like to thank members of the PTC Community, Programming and Extending Mathcad online forum for their advice on simulation and programming techniques.

References

1. 

Optical Coherence Tomography, Springer, Berlin, Heidelberg (2008). Google Scholar

2. 

D. Huang et al., “Optical coherence tomography,” Science, 254 (5035), 1178 –1181 (1991). http://dx.doi.org/10.1126/science.1957169 SCIEAS 0036-8075 Google Scholar

3. 

A. F. Fercher et al., “Measurement of intraocular distances by backscattering spectral interferometry,” Opt. Commun., 117 43 –48 (1995). http://dx.doi.org/10.1016/0030-4018(95)00119-S OPCOB8 0030-4018 Google Scholar

4. 

R. Leitgeb, C. Hitzenberger and A. Fercher, “Performance of Fourier domain vs time domain optical coherence tomography,” Opt. Express, 11 (8), 889 (2003). http://dx.doi.org/10.1364/OE.11.000889 OPEXFF 1094-4087 Google Scholar

5. 

M. Choma et al., “Sensitivity advantage of swept source and Fourier domain optical coherence tomography,” Opt. Express, 11 (18), 2183 (2003). http://dx.doi.org/10.1364/OE.11.002183 OPEXFF 1094-4087 Google Scholar

6. 

J. F. de Boer et al., “Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography,” Opt. Lett., 28 (21), 2067 (2003). http://dx.doi.org/10.1364/OL.28.002067 OPLEDP 0146-9592 Google Scholar

7. 

Z. Wang et al., “Tissue refractive index as marker of disease,” J. Biomed. Opt., 16 (11), 116017 (2011). http://dx.doi.org/10.1117/1.3656732 JBOPFO 1083-3668 Google Scholar

8. 

Y. L. Kim et al., “Variation of corneal refractive index with hydration,” Phys. Med. Biol., 49 (5), 859 –868 (2004). http://dx.doi.org/10.1088/0031-9155/49/5/015 PHMBA7 0031-9155 Google Scholar

9. 

M. A. Mayer et al., “Retinal nerve fiber layer segmentation on FD-OCT scans of normal subjects and glaucoma patients,” Biomed. Opt. Express, 1 (5), 1358 –1383 (2010). http://dx.doi.org/10.1364/BOE.1.001358 BOEICL 2156-7085 Google Scholar

10. 

S. J. Chiu et al., “Kernel regression based segmentation of optical coherence tomography images with diabetic macular edema,” Biomed. Opt. Express, 6 (4), 1172 –1194 (2015). http://dx.doi.org/10.1364/BOE.6.001172 BOEICL 2156-7085 Google Scholar

11. 

T. Garcia et al., “Optical coherence tomography in neuro-ophthalmology,” Optical Coherence Tomography, 77 –100 InTech, Vienna, Austria (2013). Google Scholar

12. 

B. Vuong et al., “Measuring the optical characteristics of medulloblastoma with optical coherence tomography,” Biomed. Opt. Express, 6 (4), 1487 (2015). http://dx.doi.org/10.1364/BOE.6.001487 BOEICL 2156-7085 Google Scholar

13. 

F. A. South et al., “Differentiation of ex vivo human breast tissue using polarization-sensitive optical coherence tomography,” Biomed. Opt. Express, 5 (10), 3417 (2014). http://dx.doi.org/10.1364/BOE.5.003417 BOEICL 2156-7085 Google Scholar

14. 

W. V. Sorin and D. F. Gray, “Simultaneous thickness and group index measurement using optical low-coherence reflectometry,” IEEE Photonics Technol. Lett., 4 (1), 105 –107 (1992). http://dx.doi.org/10.1109/68.124892 IPTLEL 1041-1135 Google Scholar

15. 

K. Watanabe, M. Ohshima and T. Nomura, “Simultaneous measurement of refractive index and thickness distributions using low-coherence digital holography and vertical scanning,” J. Opt., 16 (4), 45403 (2014). http://dx.doi.org/10.1088/2040-8978/16/4/045403 Google Scholar

16. 

H.-C. Cheng and Y.-C. Liu, “Simultaneous measurement of group refractive index and thickness of optical samples using optical coherence tomography,” Appl. Opt., 49 (5), 790 (2010). http://dx.doi.org/10.1364/AO.49.000790 APOPAI 0003-6935 Google Scholar

17. 

C.-T. Yen et al., “Simultaneously measuring the refractive index and thickness of an optical sample by using improved fiber-based optical coherence tomography,” Opt. Eng., 53 (4), 044108 (2014). http://dx.doi.org/10.1117/1.OE.53.4.044108 Google Scholar

18. 

Y.-P. Wang et al., “Reflectometry measuring refractive index and thickness of polymer samples simultaneously,” J. Mod. Opt., 53 (13), 1845 –1851 (2006). http://dx.doi.org/10.1080/09500340600654764 JMOPEW 0950-0340 Google Scholar

19. 

G. J. Tearney et al., “Determination of the refractive index of highly scattering human tissue by optical coherence tomography,” Opt. Lett., 20 (21), 2258 (1995). http://dx.doi.org/10.1364/OL.20.002258 OPLEDP 0146-9592 Google Scholar

20. 

M. Ohmi et al., “Simultaneous measurement of refractive index and thickness of transparent plates by low coherence interferometry,” Opt. Rev., 4 (4), 507 –515 (1997). http://dx.doi.org/10.1007/s10043-997-0507-1 1340-6000 Google Scholar

21. 

M. Ohmi et al., “In vitro simultaneous measurement of refractive index and thickness of biological tissue by the low coherence interferometry,” IEEE Trans. Biomed. Eng., 47 (9), 1266 –1270 (2000). http://dx.doi.org/10.1109/10.867961 IEBEAX 0018-9294 Google Scholar

22. 

M. Haruna et al., “Simultaneous measurement of the phase and group indices and the thickness of transparent plates by low-coherence interferometry,” Opt. Lett., 23 (12), 966 (1998). http://dx.doi.org/10.1364/OL.23.000966 OPLEDP 0146-9592 Google Scholar

23. 

H. Maruyama et al., “Low-coherence interferometer system for the simultaneous measurement of refractive index and thickness,” Appl. Opt., 41 (7), 1315 (2002). http://dx.doi.org/10.1364/AO.41.001315 APOPAI 0003-6935 Google Scholar

24. 

T. Fukano and I. Yamaguchi, “Separation of measurement of the refractive index and the geometrical thickness by use of a wavelength-scanning interferometer with a confocal microscope,” Appl. Opt., 38 (19), 4065 (1999). http://dx.doi.org/10.1364/AO.38.004065 APOPAI 0003-6935 Google Scholar

25. 

S. Kim et al., “Simultaneous measurement of refractive index and thickness by combining low-coherence interferometry and confocal optics,” Opt. Express, 16 (8), 5516 (2008). http://dx.doi.org/10.1364/OE.16.005516 OPEXFF 1094-4087 Google Scholar

26. 

H. Maruyama et al., “Simultaneous measurement of refractive index and thickness by low coherence interferometry considering chromatic dispersion of index,” Opt. Rev., 7 (5), 468 –472 (2000). http://dx.doi.org/10.1007/s10043-000-0468-0 1340-6000 Google Scholar

27. 

M. Ohmi et al., “High-speed simultaneous measurement of refractive index and thickness of transparent plates by low-coherence interferometry and confocal optics,” Meas. Sci. Technol., 15 (8), 1531 –1535 (2004). http://dx.doi.org/10.1088/0957-0233/15/8/017 MSTCEP 0957-0233 Google Scholar

28. 

T. Fukano and I. Yamaguchi, “Simultaneous measurement of thicknesses and refractive indices of multiple layers by a low-coherence confocal interference microscope,” Opt. Lett., 21 (23), 1942 (1996). http://dx.doi.org/10.1364/OL.21.001942 OPLEDP 0146-9592 Google Scholar

29. 

X. Wang et al., “Simultaneous refractive index and thickness measurements of bio tissue by optical coherence tomography,” J. Biomed. Opt., 7 (4), 628 –632 (2002). http://dx.doi.org/10.1117/1.1501887 JBOPFO 1083-3668 Google Scholar

30. 

S. A. Alexandrov et al., “Bifocal optical coherenc refractometry of turbid media,” Opt. Lett., 28 (2), 117 –119 (2003). http://dx.doi.org/10.1364/OL.28.000117 OPLEDP 0146-9592 Google Scholar

31. 

A. Zvyagin et al., “Refractive index tomography of turbid media by bifocal optical coherence refractometry,” Opt. Express, 11 (25), 3503 –3517 (2003). http://dx.doi.org/10.1364/OE.11.003503 OPEXFF 1094-4087 Google Scholar

32. 

G. Min, J. W. Kim and B.-H. Lee, “The refractive index measurement technique based on the defocus correction method in full-field optical coherence tomography,” Proc. SPIE, 8428 84281M (2012). http://dx.doi.org/10.1117/12.922398 Google Scholar

33. 

G. Min et al., “Refractive index measurements of multiple layers using numerical refocusing in FF-OCT,” Opt. Express, 21 (24), 29955 –29967 (2013). http://dx.doi.org/10.1364/OE.21.029955 OPEXFF 1094-4087 Google Scholar

34. 

H. Matsumoto, K. Sasaki and A. Hirai, “In situ measurement of group refractive index using tandem low-coherence interferometer,” Opt. Commun., 266 (1), 214 –217 (2006). http://dx.doi.org/10.1016/j.optcom.2006.04.082 OPCOB8 0030-4018 Google Scholar

35. 

A. Hirai and H. Matsumoto, “Low-coherence tandem interferometer for measurement of group refractive index without knowledge of the thickness of the test sample,” Opt. Lett., 28 (21), 2112 –2114 (2003). http://dx.doi.org/10.1364/OL.28.002112 OPLEDP 0146-9592 Google Scholar

36. 

D. F. Murphy and D. A. Flavin, “Dispersion-insensitive measurement of thickness and group refractive index by low-coherence interferometry,” Appl. Opt., 39 (25), 4607 (2000). http://dx.doi.org/10.1364/AO.39.004607 APOPAI 0003-6935 Google Scholar

37. 

J. Jin et al., “Thickness and refractive index measurement of a silicon wafer based on an optical comb,” Opt. Express, 18 (17), 18339 –18346 (2010). http://dx.doi.org/10.1364/OE.18.018339 OPEXFF 1094-4087 Google Scholar

38. 

J. Na et al., “Self-referenced spectral interferometry for simultaneous measurements of thickness and refractive index,” Appl. Opt., 48 (13), 2461 (2009). http://dx.doi.org/10.1364/AO.48.002461 APOPAI 0003-6935 Google Scholar

39. 

Y.-S. Ghim and S.-W. Kim, “Thin-film thickness profile and its refractive index measurements by dispersive white-light interferometry,” Opt. Express, 14 (24), 11885 (2006). http://dx.doi.org/10.1364/OE.14.011885 OPEXFF 1094-4087 Google Scholar

40. 

S. J. Park et al., “Simultaneous measurements of refractive index and thickness by spectral-domain low coherence interferometry having dual sample probes,” IEEE Photonics Technol. Lett., 23 (15), 1076 –1078 (2011). http://dx.doi.org/10.1109/LPT.2011.2155642 IPTLEL 1041-1135 Google Scholar

41. 

P. H. Tomlins and R. K. Wang, “Simultaneous analysis of refractive index and physical thickness by Fourier domain optical coherence tomography,” IEE Proc. Optoelectron., 153 (5), 222 (2006). http://dx.doi.org/10.1049/ip-opt:20050115 Google Scholar

42. 

T. Lai and S. Tang, “Cornea characterization using a combined multiphoton microscopy and optical coherence tomography system,” Biomed. Opt. Express, 5 (5), 1494 (2014). http://dx.doi.org/10.1364/BOE.5.001494 BOEICL 2156-7085 Google Scholar

43. 

Y. Zhou et al., “Characterizing refractive index and thickness of biological tissues using combined multiphoton microscopy and optical coherence tomography,” Biomed. Opt. Express, 4 (1), 38 –50 (2012). http://dx.doi.org/10.1364/BOE.4.000038 BOEICL 2156-7085 Google Scholar

44. 

A. Knüttel and M. Boehlau-Godau, “Spatially confined and temporally resolved refractive index and scattering evaluation in human skin performed with optical coherence tomography,” J. Biomed. Opt., 5 (1), 83 –92 (2000). http://dx.doi.org/10.1117/1.429972 JBOPFO 1083-3668 Google Scholar

45. 

P. H. Tomlins and R. K. Wang, “Matrix approach to quantitative refractive index analysis by Fourier domain optical coherence tomography,” J. Opt. Soc. Am. A, 23 (8), 1897 (2006). http://dx.doi.org/10.1364/JOSAA.23.001897 JOAOD6 0740-3232 Google Scholar

46. 

P. H. Tomlins and R. K. Wang, “Layer dependent refractive index measurement by Fourier domain optical coherence tomography,” Proc. SPIE, 6079 607913 (2006). http://dx.doi.org/10.1117/12.645719 Google Scholar

47. 

J. A. Izatt, M. A. Choma, “Theory of optical coherence tomography,” Optical Coherence Tomography, 47 –72 Springer, Berlin, Heidelberg (2008). Google Scholar

48. 

M. Wojtkowski et al., “Full range complex spectral optical coherence tomography technique in eye imaging,” Opt. Lett., 27 (16), 1415 (2002). http://dx.doi.org/10.1364/OL.27.001415 OPLEDP 0146-9592 Google Scholar

49. 

P. Rajai, “Measurement of refractive index and thickness of multilayer systems using Fourier domain optical coherence tomography,” University of Ottawa, (2016). Google Scholar

50. 

E. W. Weisstein, “Hadamard’s maximum determinant problem,” (2017) http://mathworld.wolfram.com/HadamardsMaximumDeterminantProblem.html November 2015). Google Scholar

Biography

Payman Rajai obtained his bachelor and master’s degrees in applied physics from Islamic Azad University, Tehran, Iran in 1995 and 2000 respectively, and his PhD in physics from the University of Ottawa in 2016. He has also spent over 10 years in the contact lens manufacturing industry.

Henry Schriemer received his PhD in physics from the University of Manitoba in 1997. He is an associate professor in electrical engineering at the University of Ottawa, where he previously held an NCIT research fellowship in photonics after spending several years in the photonics industry. Other recent activities have ranged across the photovoltaics landscape, with a present focus on energy management for smart grids. He has authored over 90 peer-reviewed publications.

Ahmad Amjadi received his master and PhD degree from Michigan State University in 1982 and 1986 respectively. Now he is associate professor in the physics department of Sharif University of technology, Tehran, Iran and head of Sharif Applied Physics Research Center since 2003. His research interest is medical physics especially application of laser in medical research.

Rejean Munger received his MSc and PhD degrees from Physics Department, the University of Waterloo, Ontario, Canada, in 1986 and 1990, respectively. He was a lecturer and research associate at the School of Optometry of the University of Waterloo from 1989 to 1994 and a research associate at the National Research Council Canada from 1994 to 1996. He then joined the University of Ottawa Eye Institute doing biomedical photonic research until 2013. He has since focused on the development of several startup biomedical photonic ventures.

© 2017 Society of Photo-Optical Instrumentation Engineers (SPIE) 1083-3668/2017/$25.00 © 2017 SPIE
Payman Rajai, Henry Schriemer, Ahmad Amjadi, and Rejean Munger "Simultaneous measurement of refractive index and thickness of multilayer systems using Fourier domain optical coherence tomography, part 1: theory," Journal of Biomedical Optics 22(1), 015002 (28 January 2017). https://doi.org/10.1117/1.JBO.22.1.015002
Received: 17 October 2016; Accepted: 3 January 2017; Published: 28 January 2017
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Refractive index

Optical coherence tomography

Interfaces

Error analysis

Optical testing

Mirrors

Refraction

Back to Top