1 July 2003 Characterizing point spread functions of two-photon fluorescence microscopy in turbid medium
Author Affiliations +

1.

Introduction

Since the experimental realization of two-photon imaging by Webb’s group in 1990,1 this technique has found broad applications in biology and medicine. Compared with conventional, single-photon excitation microscopic techniques, a number of significant advantages popularized two-photon microscopy. To begin with, the nonlinear nature of two-photon excitation limits fluorescence generation to the focal volume. When an objective with a high numerical aperture is used, the dimension of the focal spot is about 0.3 μm (radial) and 0.7 μm (axial), corresponding to a localized volume on the order of 0.1 fl.2 Microscopic imaging by raster scanning a pointlike, two-photon spot results in an axial sectioning effect similar to that found in confocal microscopy. The few fluorescent photons generated in the off-focal region interfere little with focal imaging. As a result, image quality is greatly improved and photodamage is limited. Discussions of confocal and two-photon microscopy can be found in the literature (for example, see Ref. 3). Furthermore, the near-infrared (IR) photons used for two-photon excitation of visible fluorescent probes are scattered and absorbed less by biological specimens than the UV/visible photons used in one-photon absorption.3 4 5 Therefore, greater depth penetration can be achieved by two-photon microscopy. Finally, since the near-IR wavelength used for sample excitation is spectrally well separated from typical fluorescence emissions, the complete fluorescence emission spectrum can be more easily obtained by two-photon excitation. These unique characteristics make two-photon excitation a very attractive experimental technique for biomedical imaging, especially in optically opaque samples. For example, human and mouse skin has been extensively imaged by two-photon microscopy.6 7 Corneal structure has also been examined.8 9 Two-photon microscopy has also found significant applications in neurobiology and embryology. Neurons and dendrites have been extensively investigated10 11 12 13 and processes such as sperm–egg fusion and embryo development have also been studied.14 15 In fact, a comparative study between confocal and two-photon microscopy shows that in a number of biological specimens, two-photon microscopy can penetrate at least twice as deep as confocal microscopy.16 With further development, two-photon fluorescence microscopy can become a powerful technique for in vivo diagnosis of epithelial diseases.

2.

In-depth Characterization of the Two-Photon Point Spread Function

Optical imaging of optically thick biological specimens can be hampered by photon interaction with tissue constituents. For example, tissue absorption and scattering can limit the useful range of depth penetration. Both the absorption and scattering coefficients of many tissue types have been characterized. It was found that in the red/near-IR range, the absorption, scattering, and reduced scattering coefficients are around 0.01 to 1 mm −1, 1 to 100 mm −1, and 0.1 to 5 mm −1, respectively.17 18 19 In the literature, a number of groups have reported investigation of the imaging characteristics of two-photon microscopy in a turbid medium.16 20 21 22 23 24 25 In particular, some groups reported negligible effects on the point spread function (PSF), owing to scattering.16 25 However, there has also been at least one report in which it was demonstrated that the two-photon image resolution degrades in the presence of scattering microspheres.26 Nonetheless, to the best of our knowledge, a complete characterization of the PSF in media with different scattering properties with both oil and water immersion objectives is absent in the literature.

In this paper we address this issue by characterizing the two-photon PSF in a number of turbid media. Agarose gel samples imbedded with 0.1-μm fluorescent beads in different lipid emulsions were made and imaged. For a thorough study, lipid emulsion concentrations at 0, 0.25, 0.5, 1, and 2 were made. At these concentrations, the reduced scattering coefficients around 620 nm are approximately 0, 0.38, 0.75, 1.5, and 3 mm −1 for the respective concentrations.27 A recent study has shown that μs and μs for comparable 2 Intraplipid to be 6 and 2.1 mm −1, respectively (at 780 nm).25 These values demonstrate that the scattering properties of the samples we prepared match well with those of many tissue types and that they are fair representations of tissuelike phantoms.18

3.

Experimental Apparatus

The two-photon fluorescence microscope used in this study is shown in Fig. 1. An argon ion (Innova 300, Coherent Inc., Santa Clara, California)-pumped, femtosecond titanium:sapphire (Ti:S) laser (Mira 900, Coherent Inc.) is the source of two-photon excitation. The broad lasing range (approximately 700 to 1000 nm) of the Ti:S laser can be used for the excitation of a wide range of visible fluorescent molecules. For the imaging of our 0.1-μm spheres, a 780-nm wavelength was used. Before entering the modified fluorescent microscope used in this study (Axiovert 100TV, Carl Zeiss Inc., Thornwood, New York), the laser beam passes through an x-y scanner (model 6350, Cambridge Technology, Cambridge, Massachusetts). The scanning mirror-driven, angular deviations of the laser beam are translated into linear positions of the focal spot at the sample. To ensure overfilling of the objective’s back aperture, a beam expander is positioned between the x-y scanner and the objective.

Figure 1

A two-photon fluorescence microscope using a femtosecond Ti:S laser.

029303j.1.jpg

The expanded laser beam is reflected into the back aperture of the objective by a short-pass dichroic mirror (Chroma Technology, Brattleboro, Vermont). Fluorescence generated by the two-photon spot is collected by the microscope objective in the epi-illuminated geometry. Our near-IR reflecting dichroic mirror then transmits the shorter fluorescence wavelengths. In addition to the dichroic mirror, the fluorescence also passes additional optical filters prior to reaching the photomultiplier tube (PMT). For this study, the PSF was measured using two high numerical aperture (NA) Zeiss objectives: the water immersion, 63× C-Apochromat (NA 1.2 Korr; working distance of 0.24 mm) lens and the oil immersion, 63× Plan-Neofluar (NA 1.25; working distance of 0.1 mm) objective. For each objective, the maximum imaging depths were determined by the objectives’ working distances. With each objective, we imaged the microspheres to depths nearing the objectives’ working distances (100 μm for the oil objective and 200 μm for the water objective).

To achieve three-dimensional scanning at different depths, the objective is first manually moved into the appropriate axial position using the manual dial of the microscope’s objective positioner. Then three-dimensional scanning at each sample depth is accomplished by a combination of x-y scanner and a piezoelectric-driven objective positioner in the axial direction (P-721 PIFOC, Physik Instrumente, Germany). The x-y scan of the sample is composed of 256×256 pixels at a resolution of 0.121 μm per pixel step, and the axial scan step is set at either 0.132 or 0.144 μm, depending on the piezoelectric positioner used. Depending on the sample being imaged, 51 or 76 axial scans are acquired to form a 3-D stack. This corresponds to the total axial thickness imaged to be about 6.6 or 10.9 μm, respectively. Note that the axial positions represent the mechanical movement of the stage systems in our microscope. In our data acquisition mode of single-photon counting, photons collected at each pixel are processed by a discriminator unit and recorded by the data acquisition computer. The computer then stores the 3-D image for processing.

The sample prepared consisted of 0.1-μm yellow-green fluorescent microspheres (Molecular Probes Inc., Eugene, Oregon) imbedded in low-melting-point agarose gel (LMP, GIBCO BRL, Grand Island, New York) with different amounts of scatter (Liposyn III @20, Abbott Laboratories, Abbott Park, Illinois). To prepare the sample, 2 LMP gel is dissolved and mixed with Liposyn III to achieve a final scattering concentration of 0, 0.25, 0.5, 1, and 2. While in the molten state, a small amount of the 0.1-μm beads are mixed into the gel. The molten sample mixture is then mounted onto a well slide and covered with a glass cover glass (thickness no. 1). After the gel sets, it becomes a static 3-D structure containing fluorescent microspheres and different amounts of scatter. The 63 × C-Apochromat water immersion objective is equipped with a correction collar for the cover glass thickness. To obtain the correct collar setting, a uniform fluorescence sample is used to observe the focal spot right below the cover glass. We found that Zeiss’s correction collar marks correspond well with the cover glass thickness used. For the PSF measurements, we adjusted the microsphere concentration so that sufficient microspheres were present within the scanning field. However, the microsphere concentration is sufficiently dilute that most of the spheres are well separated and overlapping of the spheres’ PSFs occurs only minimally. In fact, visual examination of the specimen reveals that most spheres are separated by at least 2 to 3 μm.

4.

Results and Discussion

A representative three-dimensional scan of the 0.1-μm microsphere sample (0 Liposyn III) taken with the 63× Plan-Neofluar, oil immersion objective is shown in Fig. 2. The data were visually examined and the approximate centroids of five microspheres were identified. To further process the data, the approximate centroids were used as marker coordinates from which the axial PSF data were extracted from the image stack by identifying the maximum pixel intensity of the microspheres at each axial position. The extracted axial distribution is then fitted to a Gaussian model I(z):

Eq. (1)

I(z)=Ioz exp[(zzo)2/σz2]+C,
where Ioz is the peak fluorescence intensity, zo is the axial centroid, σz is the axial intensity distribution width, and C is the background. The fitted centroid zo can then be used to identify the focal plane and the lateral PSF data extracted and fitted to a Gaussian model. The PSF data and the corresponding fit for one representative microsphere are shown in Fig. 3. The two-photon theoretical intensity distribution given by

Eq. (2)

I2γ(u,v)=|201Jo(vρ)exp(iuρ2/2)ρdρ|4
is also plotted in the same figure for comparison u=k(NA)2z and v=k(NA)r are the respective normalized axial and radial coordinates for wave number k. 2 28 Qualitatively, Fig. 3 shows that our experimental results agree well with the theoretical prediction.

Figure 2

A representative three-dimensional scan of multiple microspheres (0.1 μm). Image taken near the surface using the 63× oil objective volume data of the microspheres.

029303j.2.jpg

Figure 3

The radial and axial PSF plots for the sphere chosen in Fig. 2. Experimental data, results of the Gaussian fit, and theoretical distributions are shown.

029303j.3.jpg

For a detailed characterization of the PSF widths (FWHM), image data from five beads are processed and the widths of the radial and axial PSFs are averaged. The average radial and axial widths (FWHM) of the PSF for the oil objective are plotted in Fig. 4 and the results for the water objective are shown in Fig. 5. A numerical representation of these FWHMs is shown in Table 1. For ease of viewing, the standard deviations of the fitted widths are only shown in the table.

Figure 4

Averaged widths of the radial and axial PSFs from five spheres at two depths and five Liposyn III concentrations (0, 0.25, 0.5, 1, and 2), taken with 63×, oil immersion, Plan-Neofluar objective (NA 1.25).

029303j.4.jpg

Figure 5

Averaged widths of the radial and axial PSF from five spheres at four depths and five Liposyn III concentrations (0, 0.25, 0.5, 1, and 2), taken with 63×, water immersion, C-Apochromat objective (NA 1.2). μs×d is the product of the scattering coefficient μs and imaging depth d.

029303j.5.jpg

Table 1

PSF widths (FWHM) for 63× oil (NA 1.25) and 63× water immersion (NA 1.2) objectives at different starting depths (in microns).
63× oil 63× water
Radial
FWMH
SD Axial
FWHM
SD Radial
FWMH
SD Axial
FWHM
SD
Theory2200.230 0.636 0.240 0.691
0
0 μm
μs×d
=0
0.25 0.03 0.73 0.04 0.28 0.03 0.86 0.04
25 μm 0 0.28 0.04 1.30 0.20
50 μm 0 0.30 0.02 1.40 0.06 0.30 0.04 0.79 0.05
75 μm 0 0.31 0.03 1.88 0.13
100 μm 0 0.30 0.03 1.86 0.16 0.28 0.02 0.82 0.06
150 μm 0 0.28 0.02 0.91 0.05
200 μm 0 0.28 0.02 0.84 0.07
0.25
0 μm
0 0.24 0.01 0.75 0.09 0.27 0.01 0.91 0.09
25 μm 0.01875 0.28 0.06 1.26 0.12
50 μm 0.0375 0.25 0.06 1.23 0.15 0.29 0.04 0.87 0.10
75 μm 0.05625 0.30 0.06 1.76 0.10
100 μm 0.075 0.33 0.08 2.08 0.28 0.26 0.05 0.92 0.07
150 μm 0.1125 0.29 0.06 0.89 0.06
200 μm 0.15 0.27 0.04 0.86 0.07
0.5
0 μm
0 0.26 0.01 0.75 0.02 0.29 0.03 0.89 0.09
25 μm 0.0375 0.28 0.05 1.39 0.10
50 μm 0.075 0.29 0.01 1.37 0.12 0.28 0.03 0.88 0.06
75 μm 0.1125 0.27 0.03 1.90 0.10
100 μm 0.15 0.28 0.06 1.88 0.40 0.29 0.03 0.95 0.09
150 μm 0.225 0.28 0.02 0.92 0.11
200 μm 0.30 0.27 0.02 0.84 0.07
1
0 μm
0 0.27 0.03 0.77 0.08 0.25 0.03 0.87 0.06
25 μm 0.075 0.28 0.05 1.39 0.04
50 μm 0.15 0.31 0.01 1.30 0.05 0.27 0.03 0.91 0.13
75 μm 0.225 0.31 0.04 1.86 0.23
100 μm 0.30 0.26 0.05 1.73 0.25 0.27 0.02 0.84 0.06
150 μm 0.45 0.29 0.06 0.89 0.07
200 μm 0.60 0.29 0.06 0.86 0.10
2
0 μm
0 0.24 0.04 0.77 0.06 0.26 0.04 0.91 0.07
25 μm 0.15 0.26 0.02 1.25 0.23
50 μm 0.30 0.27 0.06 1.16 0.13 0.27 0.03 0.78 0.11
75 μm 0.45 0.30 0.05 1.71 0.28
100 μm 0.60 0.24 0.06 1.92 0.47 0.27 0.07 0.85 0.06
150 μm 0.90 0.23 0.04 0.95 0.05
200 μm 1.20 0.25 0.05 0.86 0.13
Note: The results are obtained from averaging five 0.1-μm microspheres. Theoretical values are also calculated.228μs×d is the product of scattering coefficient μs and imaging depth. μs is estimated from the comparable Intralipid solution.25

Figure 4, Fig. 5, and Table 1 show that the experimental PSF widths are very different for the oil and water objectives. In the case of the oil objective, our lateral PSF widths vary from 0.24 to 0.31 μm. These values are slightly larger than the theoretical value of 0.230 μm. As Fig. 4 shows, there is a tendency for the lateral width to increase with imaging depth. However, there is no indication that increasing scatterer concentration correlates with an increase in lateral widths. For example, at a depth of 100 μm, the lateral PSF width at the surface is 0.30 μm for the 0 Liposyn III sample. At the same depth, the lateral width fluctuates as the Liposyn III concentration is increased from 0.25 to 2. In the case of the 2 sample, the lateral width is 0.24 μm, a value 20 less than the surface value. The same fluctuating trend is observed with datasets at other depths, but the fact remains that up to 2 Liposyn III, scattering has no measurable effect on lateral PSF widths. The axial FWHM data demonstrate the same independence of the PSF widths from the scatterer concentration present. However, as imaging depths increase, there is a corresponding increase in the axial PSF widths. For example, in the case of the 0 sample, the axial FWHM increased from 0.73 (at surface) to 1.86 μm (at 100 μm). The degradation of the lateral and axial widths has been reported for oil immersion objectives and is most likely due to spherical aberration.22 23 29

In the case of the water immersion objective, the PSF data are slightly larger than the theoretical value. For the 0 Liposyn III sample, our surface values of the lateral and axial widths are 0.28 and 0.86 μm, respectively. These values are slightly larger than the corresponding theoretical values of 0.240 (by about 17) and 0.691 μm (by about 24). An examination of Fig. 4, Fig. 5, and Table 1 shows that the water objective results are very different from those of the oil lens. In the case of the water objective, we also plotted the PSF FWHMs as a function of the scattering parameter μs×d. For the water lens, there is no evidence of the degradation of the PSF width with imaging depths (up to 200 μm) and Liposyn III concentration (up to 2). Regardless of the dataset examined, the PSF widths show no obvious correlation with sample depth and scatter concentration. All PSF width values fluctuate around 0.28 μm (lateral) and 0.9 μm (axial). The spherical aberration-induced degradation of the PSF observed with the oil objective is absent in the water objective data. These results support the fact that the refractive index of the gel closely matches that of water.

In addition to the PSF width analysis, we also calculated the laterally integrated axial intensity profile (LAIP) from the microsphere images. This is accomplished by summing the intensity in a 13×13 pixel region around the approximate centroid of the microsphere. The 169-pixel region corresponds to an area of 1.45×1.45 μm 2. Since the lateral widths do not appreciably broaden for both objectives (Fig. 4, Fig. 5, and Table 1), the area chosen covers the lateral extent of the PSF and accurately represents the total fluorescence generated at the chosen axial plane. Therefore, the LAIPs can be used to test the extent of image sectioning in our specimen. Our motivation for performing this analysis is the increased sensitivity in detecting PSF degradation. In regions away from the PSF maximum, there may not be sufficient intensity for the PSF to be accurately measured. However, the LAIPs in those regions will have a stronger value and allow a more sensitive analysis of the scattering effect on the PSF. The LAIP for the same 0.1-μm microsphere used for the PSF determination in Fig. 3 is shown in Fig. 6. Comparing Fig. 3 and Fig. 6, one can see that the PSF data reach the background level at about 0.75 μm from the intensity maximum. However, the LAIP did not reach background level until about 1.5 μm from the peak intensity. It is this increase in sensitivity that we hope will allow a more thorough characterization of the PSF in scattering media. For the LAIPs, the Gaussian model is again used to fit the data for their FWHMs. For comparison, the theoretical result given in Refs. 2 and 28:

Eq. (3)

Int2γ(u)=0|201J0(υρ)exp(iuρ2/2)ρdρ|4υdυ/0|201J0(υρ)ρdρ|4υdυ
is also plotted in Fig. 6.

Figure 6

Representative radially integrated intensity distribution for 63× oil immersion objective of a 0.1 μm fluorescent microsphere.

029303j.6.jpg

To thoroughly quantify the effect of imaging depths and scattering concentration on the LAIPs, we fitted the distributions for all the microsphere data to the Gaussian model and plotted the LAIP FWHMs in Fig. 7. And in the case of the water immersion objective, we plotted the LAIP FWHMs as a function of the scattering parameter μs×d. The results are also given in Table 2. Our analysis shows the same results as the PSF analysis: up to 2, the amount of scattering in Liposyn III has no discernable effects on the widths of the LAIPs. For example, in the case of the 63× water objective, the LAIP FWHM at a depth of 200 μm is 1.16 μm (0 Liposyn III sample). At the same imaging depth, results from other scattering samples (1.19 μm for 0.25, 1.20 μm for 0.5, 1.26 μm for 1, and 1.19 μm for 2 Liposyn III) show no sign of PSF degradation caused by increased scattering. Data acquired at other depths also show no correlation of scattering on the LAIP FWHMs. On the other hand, the LAIP results from the oil immersion objective show the effects of degradation induced by spherical aberration. In this case, the LAIP FWHMs increased from approximately 1 μm at the surface to about 2.5 μm at a depth of 100 μm. Therefore, the additional sensitivity of the LAIPs reveals no scattering-induced degradation of the PSF.

Figure 7

The FWHMs for radially integrated intensity at different depths for both the 63× oil and water immersion objectives. μs×d is the product of the scattering coefficient μs and imaging depth d.

029303j.7.jpg

Table 2

FWHMs (in microns) of laterally integrated axial intensity profiles for 63× oil (NA 1.25) and 63× water immersion (NA 1.2) objectives at different starting depths.
63× oil 63× water
FWHM SD FWHM SD
Theory0.676 0.734
0
0 μm
μs×d
=0
0.96 0.04 1.15 0.04
25 μm 0 1.63 0.07
50 μm 0 1.86 0.07 1.16 0.11
75 μm 0 2.40 0.06
100 μm 0 2.39 0.10 1.10 0.06
150 μm 0 1.13 0.06
200 μm 0 1.16 0.09
0.25
0 μm
0 1.00 0.06 1.28 0.09
25 μm 0.01875 1.67 0.07
50 μm 0.0375 1.68 0.08 1.20 0.13
75 μm 0.05625 2.35 0.25
100 μm 0.075 2.75 0.36 1.21 0.09
150 μm 0.1125 1.25 0.14
200 μm 0.15 1.19 0.13
0.5
0 μm
0 1.02 0.05 1.17 0.12
25 μm 0.0375 1.92 0.22
50 μm 0.075 1.90 0.22 1.24 0.10
75 μm 0.1125 2.60 0.24
100 μm 0.15 2.56 0.56 1.23 0.06
150 μm 0.225 1.32 0.17
200 μm 0.30 1.20 0.12
1
0 μm
0 1.01 0.08 1.34 0.05
25 μm 0.075 1.80 0.20
50 μm 0.15 1.81 0.18 1.33 0.21
75 μm 0.225 2.37 0.36
100 μm 0.30 2.26 0.27 1.20 0.10
150 μm 0.45 1.28 0.13
200 μm 0.60 1.26 0.20
2
0 μm
0 1.06 0.07 1.23 0.13
25 μm 0.15 2.02 0.20
50 μm 0.30 1.54 0.12 1.11 0.08
75 μm 0.45 2.44 0.89
100 μm 0.60 2.94 0.68 1.20 0.17
150 μm 0.90 1.24 0.10
200 μm 1.20 1.19 0.22
Note: The results are obtained from averaging five 0.1-μm microspheres.228μs×d is the product of scattering coefficient μs and imaging depth. μs is estimated from the comparable Intralipid solution.25

In fact, when we fitted the data for radial widths, axial widths, and laterally integrated intensity widths at each axial position as a function of Liposyn III concentration to a linear model, we found that the slope of the line stayed very close to 0. The magnitude of the slope for most fits is less than 0.1, indicating no correlation between that scattering and the radial PSF, axial PSF, and LAIP widths. Our results demonstrated conclusively that for two-photon imaging down to a depth of 200 μm, scattering due to Liposyn III samples (up to 2) produces no measurable effects on the two-photon PSF.

5.

Conclusion

In this study, we thoroughly characterized the two-photon PSF in turbid medium using oil and water immersion objectives. Tissuelike scattering samples containing Liposyn III (0, 0.25, 0.5, 1, and 2) and 0.1-μm fluorescent microspheres imbedded in agarose gel were imaged, and the results were analyzed for the radial PSF, axial PSF, and LAIP widths. Our analysis showed that at imaging depths up to 100 μm (oil objective) and 200 μm (water objective), neither the radial nor the axial PSF showed any scattering-induced degradation. The axial sectioning capabilities determined from the LAIPs also appear not to be affected by the scattering concentrations used. The only degradation in the PSF was observed for the oil immersion objective at increased depths, an effect that is most likely due to spherical aberration. Our results support the generally accepted view that the presence of scatterers in tissuelike phantoms does not affect the PSF, and the spatial resolution of two-photon imaging in a turbid medium remains close to the diffraction limits.16 25

REFERENCES

1. 

W. Denk , J. H. Strickler , and W. W. Webb , “Two-photon laser scanning fluorescence microscopy,” Science (Washington, DC, U.S.) , 248 73 –76 (1990). Google Scholar

2. 

C. J. R. Sheppard and M. Gu , “Image formation in two-photon fluorescence microscopy,” Optik (Jena) , 86 (3), 104 –106 (1990). Google Scholar

3. 

Handbook of Biological Confocal Microscopy, J. B. Pawley, Ed., Plenum, New York (1995).

4. 

J. D. Jackson, Classical Electrodynamics, Wiley, New York (1975).

5. 

I. Itzkan and J. A. Izatt , Medical Use of Laser Encyclopedia of Applied Physics , 10 33 –59 (1994). Google Scholar

6. 

B. R. Masters , P. T. C. So , and E. Gratton , “Multiphoton excitation fluorescence microscopy and spectroscopy of in vivo human skin,” Biophys. J. , 72 2405 –2412 (1997). Google Scholar

7. 

P. T. C. So , H. Kim , and I. E. Kochevar , “Two-photon deep tissue ex vivo imaging of mouse dermal and subcutaneous structures,” Opt. Express , 3 (9), 339 –350 (1998). Google Scholar

8. 

D. W. Piston , B. R. Masters , and W. W. Webb , “Three-dimensionally resolved NAD(P)H cellular metabolic redox imaging of the in situ cornea with two-photon excitation laser scanning microscopy,” J. Microsc. , 178 20 –27 (1995). Google Scholar

9. 

C. Buehler , K. H. Kim , C. Y. Dong , B. R. Masters , and P. T. C. So , “Innovations in two-photon deep tissue microscopy,” IEEE Eng. Med. Biol. Mag. , 18 23 –30 (1999). Google Scholar

10. 

W. Denk , K. R. Delaney , A. Gelperin , D. Kleinfeld , B. W. Strobridge , D. W. Tank , and R. Yuste , “Anatomical and functional imaging of neurons using 2-photon laser scanning microscopy,” J. Neurosci. Methods , 54 151 –62 (1994). Google Scholar

11. 

W. Denk , M. Sugimori , and R. Llinas , “Two types of calcium response limited to single spines in cerebellar Purkinje cells,” Proc. Natl. Acad. Sci. U.S.A. , 92 8279 –8282 (1995). Google Scholar

12. 

K. Svoboda , W. Denk , D. Kleinfeld , and D. W. Tank , “In vivo dendritic calcium dynamics in neocortical pyramidal neurons,” Nature (London) , 385 161 –165 (1997). Google Scholar

13. 

R. Yuste and W. Denk , “Dendritic spines as basic functional units of neuronal integration,” Nature (London) , 375 682 –684 (1995). Google Scholar

14. 

K. T. Jones , C. Soeller , and M. B. Cannel , “The passage of Ca2+ and fluorescent markers between the sperm and egg after fusion in the mouse,” Development , 125 4627 –4635 (1998). Google Scholar

15. 

J. M. Squirrell , D. L. Wokosin , J. G. White , and B. D. Bavister , “Long-term two-photon fluorescence imaging of mammalian embryos without compromising viability,” Nat. Biotechnol. , 17 763 –767 (1999). Google Scholar

16. 

V. E. Centonze and J. G. White , “Multiphoton excitation provides optical sections from deeper with scattering specimens than confocal imaging,” Biophys. J. , 75 2015 –2024 (1998). Google Scholar

17. 

S. T. Flock , B. C. Wilson , and M. S. Patterson , “Total attenuation coefficients and scattering phase functions of tissues and phantom materials at 633 nm,” Med. Phys. , 14 (5), 835 –841 (1987). Google Scholar

18. 

W.-F. Cheong , S. A. Prahl , and A. J. Welch , “A review of the optical properties of biological tissues,” IEEE J. Quantum Electron. , 26 (12), 2166 –2185 (1990). Google Scholar

19. 

B. C. Wilson and S. L. Jacques , “Optical reflectance and transmittance of tissues: principles and applications,” IEEE J. Quantum Electron. , 26 (12), 2186 –2199 (1990). Google Scholar

20. 

V. Daria , C. M. Blanca , O. Nakamura , S. Kawata , and C. Saloma , “Image contrast enhancement for two-photon fluorescence microscopy in a turbid medium,” Appl. Opt. , 37 (34), 7960 –7967 (1998). Google Scholar

21. 

C. M. Blanca and C. Saloma , “Monte Carlo analysis of two-photon fluorescence imaging through a scattering medium,” Appl. Opt. , 37 (34), 8092 –8102 (1998). Google Scholar

22. 

C. J. de Grauw , J. M. Vroom , H. T. M. van der Voort , and H. C. Gerritsen , “Imaging properties of two-photon excitation microscopy and effects of refractive-index mismatch in thick specimens,” Appl. Opt. , 38 (28), 5995 –6003 (1999). Google Scholar

23. 

H. C. Gerritsen and C. J. de Grauw , “Imaging of optically thick specimen using two-photon excitation microscopy,” Microsc. Res. Tech. , 47 206 –209 (1999). Google Scholar

24. 

J. Ying , F. Liu , and R. R. Alfano , “Spatial distribution of two-photon-excited fluorescence in scattering media,” Appl. Opt. , 38 (1), 224 –229 (1999). Google Scholar

25. 

A. K. Dunn , V. P. Wallace , M. Coleno , M. W. Berns , and B. J. Tromberg , “Influence of optical properties on two-photon fluorescence imaging in turbid samples,” Appl. Opt. , 39 (7), 1194 –1201 (2000). Google Scholar

26. 

S. P. Schilders and M. Gu , “Limiting factors on image quality in imaging through turbid media under single-photon and two-photon excitation,” Microsc. Microanal. , 6 156 –160 (2000). Google Scholar

27. 

J. B. Fishkin , P. T. C. So , A. E. Cerussi , S. Fantini , M. A. Franceschini , and E. Gratton , “Frequency-domain method for measuring spectral properties in multiple-scattering media: methemoglobin absorption spectrum in a tissuelike phantom,” Appl. Opt. , 34 (7), 1143 –1155 (1995). Google Scholar

28. 

C. Y. Dong , P. T. C. So , C. Buehler , and E. Gratton , “Spatial resolution in scanning pump-probe fluorescence microscopy,” Optik (Jena) , 106 (1), 7 –14 (1997). Google Scholar

29. 

H. Jacobsen , P. E. Ha¨nninen , and S. W. Hell , “Refractive-index-induced aberrations in two-photon confocal fluroescence microscopy,” J. Microsc. , 176 (3), 226 –230 (1994). Google Scholar
©(2003) Society of Photo-Optical Instrumentation Engineers (SPIE)
Chen-Yuan Dong, Karsten Koenig, and Peter T. C. So "Characterizing point spread functions of two-photon fluorescence microscopy in turbid medium," Journal of Biomedical Optics 8(3), (1 July 2003). https://doi.org/10.1117/1.1578644
Published: 1 July 2003
Lens.org Logo
CITATIONS
Cited by 83 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Point spread functions

Scattering

Objectives

Luminescence

Two photon excitation microscopy

Water

Data modeling

Back to Top